Skip to content

Trinias 503-79869-45Recalls & reported problems

Shimadzu Corporation · Class 2

All models of Trinias (12)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K203535); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Sep 30, 2025Class IIOpen, Classified

    The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.

    Z-0347-2027 · Root cause: Process control · Recall details · FDA record ↗

FDA clearance and classification

Cleared through 510(k) K203535 on Apr 28, 2021 (decision: Substantially Equivalent). Submitted by Shimadzu Corporation Medical Systems as “Trinias”. The FDA review took 146 days. FDA record ↗

The FDA regulates product code OWB (interventional fluoroscopic x-ray system) under 21 CFR 892.1650 as a Class II device. 21 CFR 892.1650 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K203535. One submission often covers many models, so check the product and lot details in each record.

  • Z-2294-2026 · Apr 20, 2026 · Open, Classified
    Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST alexa:UD150G-40 / D1

Common questions about the Trinias 503-79869-45

Has the Trinias 503-79869-45 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0347-2027 (Class II), began on Sep 30, 2025: The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the Trinias 503-79869-45 FDA approved?

It was cleared through 510(k) K203535 on Apr 28, 2021. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Trinias 503-79869-45, and is it still sold?

The FDA's device database lists Shimadzu Corporation as the company responsible for it (brand Trinias, model 503-79869-45). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Interventional Fluoroscopic X-Ray System devices

All Interventional Fluoroscopic X-Ray System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page