Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04540778162334), received between Aug 2, 2018 and Apr 8, 2019 — classified as 2 injuries.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Separation2
- Device Handling Problem1
- Use of Device Problem1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 8, 2019Material Separation; Material Separation
IN-STENT RESTENOSIS LESION IN LEFT SFA. THE CATHETER WAS APPROACHED TO THE LESION USING DESTINATION SHEATH FROM THE RIGHT FEMORAL ARTERY. WHEN THE BALLOON WAS DILATED AT 12 ATM AT THE LESION, IT WAS RUPTURED AND COULD NOT BE REMOVED. THE OPERATOR TRIED TO PULL THE BALLOON INTO THE SHEATH AFTER BRINGING THE DISTAL TIP OF THE SHEATH AS CLOSE TO THE BALLOON AS POSSIBLE, BUT THE OPERATOR COULDN'T DO IT. AFTER THAT, THE 6 FR SHEATH WAS REPLACED WITH 8FR AND IT WAS ATTEMPTED TO WITHDRAW THE BALLOON WITH SNARE, BUT IT COULD NOT BE DONE. FINALLY, WHILE THE CATHETER WAS CONTINUOUSLY PULLED OUT, THE BALLOON WAS BROKEN NEAR THE PUNCTURED PART.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8494202 ↗InjuryAug 2, 2018Material Separation; Use of Device Problem; Device Handling Problem; Material Separation; Use of Device Problem; Device Handling Problem
CROSSOVER ACCESS 7F DESTINATION, OCCLUSION DISTAL AFS TO SEGMENT P1, RECANALIZATION WITH GWA 18, VERY HEAVILY CALCIFIED VASCULAR SEGMENT, SUBSEQUENT DILATATION OF THE RECANALIZED AREA WITH THE BALLOON, THEREBY BURSTING OF THE BALLOON. UPON WITHDRAWAL, IT IS NOTED THAT THERE IS A DETACH OF THE 20 CM LONG BALLOON (ABOUT 5 CM DISTAL) AND REMAIN IN THE PATIENT. SUBSEQUENTLY, SUCCESSFUL RECANALIZATION OF THE REMAINING BALLOON SEGMENT AND OVERSTENTING WITH A SUPERA.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7742799 ↗
Other Catheter, Angioplasty, Peripheral, Transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page