Hitachi SCENARIA ViewRecalls & reported problems
Hitachi, Ltd. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04580292766694), received between Jul 26, 2022 and Jul 26, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Computer Software Problem” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jul 12, 2022Open, Classified
Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation
Z-1748-2022 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Computer Software Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 26, 2022Computer Software Problem; Computer Software Problem
ON MARCH 25, 2022, FUJIFILM HEALTHCARE AMERICAS CORPORATION RECEIVED A COMPLAINT INVOLVING THE SCENARIA VIEW CT. IT WAS REPORTED THAT WHILE PERFORMING A SCANOGRAM, AN ERROR OCCURRED DURING A PATIENT STUDY. THE ERROR WAS DUE TO SPECIFIC SEARCH STRINGS CONTAINING LOWERCASE LETTERS ENTERED IN THE PATIENT INFORMATION FIELDS. THE IMAGES WERE NOT RECONSTRUCTED BECAUSE OF THE ERROR AND THE RAWDATA WAS NOT DISPLAYED IN THE WORKLIST. THE PATIENT WAS SUCCESSFULLY SCANNED USING ANOTHER DEVICE. THIS ISSUE WAS INITIALLY IDENTIFIED AS NON-REPORTABLE BECAUSE THERE WAS NO DEATH, PATIENT INJURY, OR CONCERN OF OVEREXPOSURE ASSOCIATED WITH THE EVENT. HOWEVER, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED THAT THE MANUFACTURER INITIATED A RECALL ON JULY 12, 2022 TO CORRECT THIS ISSUE. AS SUCH, IT WAS DETERMINED THAT THIS COMPLAINT SHOULD BE REPORTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15102705 ↗MalfunctionJul 26, 2022Computer Software Problem; Computer Software Problem
ON (B)(6) 2022, FUJIFILM HEALTHCARE AMERICAS CORPORATION RECEIVED A COMPLAINT INVOLVING THE SCENARIA VIEW CT. IT WAS REPORTED THAT WHILE PERFORMING A DOUBLE SCANOGRAM, AN ERROR OCCURRED DURING A PATIENT STUDY. THE ERROR WAS DUE TO SPECIFIC SEARCH STRINGS CONTAINING LOWERCASE LETTERS ENTERED IN THE PATIENT INFORMATION FIELDS. AS A RESULT, THE IMAGES WERE NOT RECONSTRUCTED AND THE RAWDATA WAS NOT DISPLAYED IN THE WORKLIST. THE PATIENT WAS SUCCESSFULLY SCANNED USING ANOTHER DEVICE. THIS ISSUE WAS INITIALLY IDENTIFIED AS NON-REPORTABLE BECAUSE THERE WAS NO DEATH, PATIENT INJURY, OR CONCERN OF OVEREXPOSURE ASSOCIATED WITH THE EVENT. HOWEVER, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED THAT THE MANUFACTURER INITIATED A RECALL ON (B)(6) 2022 TO CORRECT THIS ISSUE. AS SUCH, IT WAS DETERMINED THAT THIS COMPLAINT SHOULD BE REPORTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15102698 ↗MalfunctionJul 26, 2022Computer Software Problem; Computer Software Problem
ON (B)(6), 2022, FUJIFILM HEALTHCARE AMERICAS CORPORATION RECEIVED A COMPLAINT INVOLVING THE SCENARIA VIEW CT. IT WAS REPORTED THAT WHILE PERFORMING A HEAD SCAN, AN ERROR OCCURRED DURING A PATIENT STUDY AND THE DEVICE STOPPED. THE ERROR WAS DUE TO SPECIFIC SEARCH STRINGS CONTAINING LOWERCASE LETTERS ENTERED IN THE PATIENT INFORMATION FIELDS. THE IMAGES WERE NOT RECONSTRUCTED BECAUSE OF THE ERROR AND THE RAWDATA WAS NOT DISPLAYED IN THE WORKLIST. THE PATIENT WAS SUCCESSFULLY SCANNED USING ANOTHER DEVICE. THIS ISSUE WAS INITIALLY IDENTIFIED AS NON-REPORTABLE BECAUSE THERE WAS NO DEATH, PATIENT INJURY, OR CONCERN OF OVEREXPOSURE ASSOCIATED WITH THE EVENT. HOWEVER, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED THAT THE MANUFACTURER INITIATED A RECALL ON (B)(6) 2022 TO CORRECT THIS ISSUE. AS SUCH, IT WAS DETERMINED THAT THIS COMPLAINT SHOULD BE REPORTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15102694 ↗
Other System, X-Ray, Tomography, Computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls6 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Brilliance CT Big BorePhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed7 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page