Single Use 3-Lumen Sphincterotome V KD-V411M-0725 (UDI 04953170184031)Recalls & reported problems

Olympus Medical Systems Corp. · Single Use 3-Lumen Sphincterotome V · Class 2

All models of Single Use 3-Lumen Sphincterotome V (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus Sphincterotome V is a single-use cutting tool used in endoscopic bile duct procedures. Of the 160 reports, 138 are malfunctions, and many describe the cutting wire breaking, with some describing melting or sparking. The reports include no deaths and 22 injuries, and reports fell to 10 in the last 12 months from 35 before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
160
Deaths reported
0
Injuries reported
22
Recalls naming it
0

Summary

The FDA's MAUDE database contains 160 adverse event reports that list this exact device (DI 04953170184031), received between Dec 24, 2019 and Sep 22, 2026 — classified as 22 injuries, 138 malfunctions.

The most frequently reported specific device problem is “Break” (129 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 3 reportsJan 2024: 4 reportsFeb 2024: 5 reportsMar 2024: 4 reportsApr 2024: 2 reportsMay 2024: 3 reportsJun 2024: 4 reportsJul 2024: 4 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 4 reportsNov 2024: 12 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 5 reportsApr 2025: 3 reportsMay 2025: 3 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 5 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break129
  • Detachment of Device or Device Component13
  • Adverse Event Without Identified Device or Use Problem10
  • Melted6
  • Sparking4
  • Peeled/Delaminated3
  • Material Split, Cut or Torn2
  • Thermal Decomposition of Device2

Patient problems

  • No Clinical Signs, Symptoms or Conditions142
  • No Consequences Or Impact To Patient6
  • Pancreatitis6
  • Hemorrhage/Blood Loss/Bleeding4
  • Unspecified Hepatic or Biliary Problem3
  • Perforation2
  • Abscess1
  • Cholangitis1

Recent adverse event reports

The 25 most recent of 160 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50862694 ↗
  • MalfunctionApr 20, 2026Break

    IT WAS REPORTED THE SINGLE USE 2-LUMEN SPHINCTEROTOME'S WIRE IS FRACTURED. THE ISSUE OCCURRED DURING AN UNKNOWN PROCEDURE AND WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24926856 ↗
  • MalfunctionApr 14, 2026Break

    IT WAS REPORTED THE SINGLE USE 2-LUMEN SPHINCTEROTOME KNIFE WIRE WAS BROKEN. THE ISSUE WAS FOUND WHEN REMOVING THE DEVICE FROM THE BOX. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24874062 ↗
  • MalfunctionJan 27, 2026Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24190645 ↗
  • MalfunctionJan 27, 2026Break

    IT WAS REPORTED THE SINGLE-USE SPHINCTEROTOME'S KNIFE TIP WAS FRACTURED. THIS OCCURRED DURING SET-UP FOR AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24189720 ↗
  • MalfunctionJan 21, 2026Break

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24135696 ↗
  • MalfunctionDec 26, 2025Break; Break

    IT WAS REPORTED THE SINGLE USE 3-LUMEN SPHINCTEROTOME V EXHIBITED KNIFE WIRE IS FRACTURED. THE ISSUE OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23899017 ↗
  • MalfunctionDec 11, 2025Break; Break

    ADDITIONAL INFORMATION PROVIDED: IT WAS REPORTED THAT THE KNIFE WIRE WAS FRACTURED UPON INITIAL OPENING OF THE BOX (OOB) AND HAD NOT BEEN USED BY THE USER. UPON RETURN OF THE DEVICE, THE VISUAL INSPECTION CONCLUDED THAT THE DEVICE HAD BEEN UTILIZED AND ENERGIZED DURING AN UNSPECIFIED PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23777602 ↗
  • MalfunctionNov 13, 2025Break; Fracture; Break; Fracture

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23542885 ↗
  • MalfunctionNov 10, 2025Break; Break

    IT WAS REPORTED THAT THE KNIFE WIRE ON THE SPHINCTEROTOME FRACTURED. THE EVENT OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ECRP) AND THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23510167 ↗
  • MalfunctionSep 16, 2025Break; Break

    IT WAS REPORTED THE SUBJECT DEVICE KNIFE BOW WAS FRACTURED. THE ISSUE OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE WITH PROCEDURAL DELAY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23067043 ↗
  • MalfunctionSep 16, 2025Break; Break

    IT WAS REPORTED THE SUBJECT DEVICE KNIFE WIRE WAS FRACTURED. THIS ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23066611 ↗
  • MalfunctionSep 16, 2025Break; Break

    IT WAS REPORTED THE SUBJECT DEVICE KNIFE WIRE WAS FRACTURED. THIS ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23066503 ↗
  • MalfunctionAug 26, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22896923 ↗
  • MalfunctionAug 21, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22861949 ↗
  • MalfunctionAug 18, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22830745 ↗
  • MalfunctionAug 15, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22817244 ↗
  • MalfunctionAug 1, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22686276 ↗
  • MalfunctionJul 8, 2025Break; Break

    THE CUSTOMER REPORTED THAT AN OLYMPUS SINGLE-USE SPHINCTEROTOME KNIFE WIRE FRACTURED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE. ACCORDING TO THE INITIAL REPORTER, THIS FAILURE OCCURRED WHILE THE DEVICE WAS INSIDE THE PATIENT AND THE PATIENT WAS UNDER SEDATION. REPORTEDLY, THE PROCEDURE WAS COMPLETED USING A BACK-UP OLYMPUS DEVICE WITH NO PATIENT INJURIES OR ADVERSE EFFECTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22435807 ↗
  • MalfunctionJun 19, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SINGLE USE 3-LUMEN SPHINCTEROTOME EXHIBITED MUCUS ADHERING TO THE SURFACE OF THE DISTAL END OF THE KNIFE WIRE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22293893 ↗
  • MalfunctionJun 6, 2025Break; Break

    IT WAS REPORTED THE SINGLE USE 3-LUMEN SPHINCTEROTOME HAD A BLADE FRACTURE. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY AND THE PROCEDURE WAS COMPLETED WITH A BACKUP OLYMPUS DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22161577 ↗
  • MalfunctionMay 28, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22098643 ↗
  • MalfunctionMay 15, 2025Break; Break

    IT WAS REPORTED THAT DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE. CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE KNIFE WIRE HAD BROKEN. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22026897 ↗
  • MalfunctionMay 5, 2025Break; Break

    IT WAS REPORTED THE SUBJECT DEVICE HAD A FRACTURE OF THE KNIFE BLADE DURING SURGERY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21960774 ↗
  • MalfunctionApr 29, 2025Break; Break

    IT WAS REPORTED THE SUBJECT DEVICE KNIFE WIRE BROKE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21926490 ↗

Other Unit, electrosurgical, endoscopic (with or without accessories) devices