SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCRecalls & reported problems
Olympus Medical Systems Corp. · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus SONICBEAT ultrasonic surgical instrument cuts and seals tissue during endoscopic surgery. Its 378 reports mostly describe tissue pads peeling or wearing away and probe tips breaking off. A few describe a part falling off inside a patient. The reports include 31 injuries and no deaths. Reports fell from 13 in the previous 12 months to 6 in the last 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 379
- Deaths reported
- 0
- Injuries reported
- 31
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 379 adverse event reports that list this exact device (DI 04953170370540), received between Oct 10, 2017 and Sep 30, 2026 — classified as 31 injuries, 348 malfunctions.
The most frequently reported specific device problem is “Material Separation” (125 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation125
- Peeled/Delaminated79
- Break70
- Detachment of Device or Device Component65
- Degraded31
- Melted15
- Fracture13
- Material Split, Cut or Torn7
Patient problems
- No Clinical Signs, Symptoms or Conditions352
- No Consequences Or Impact To Patient13
- Foreign Body In Patient3
- No Known Impact Or Consequence To Patient3
- Hematoma2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Bacterial Infection1
- Bowel Burn1
Recent adverse event reports
The 25 most recent of 379 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Break
The customer returned the instrument to the service center for a separate issue. During the device analysis, the device evaluator indicated that the probe was broken off. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951670 ↗MalfunctionJun 24, 2026Peeled/Delaminated
THE CUSTOMER RETURNED THE SONICBEAT SURGICAL INSTRUMENT TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING OLYMPUS¿ DEVICE ANALYSIS IT WAS INDICATED THAT THE SONICBEAT'S TISSUE GRASPING INSULATION PAD WAS PEELING OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25611943 ↗MalfunctionMay 29, 2026Degraded
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED THE TISSUE PAD IS SEVERELY WORN AWAY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333995 ↗MalfunctionFeb 9, 2026Break
IT WAS REPORTED THAT DURING A THERAPEUTIC HYSTERECTOMY PROCEDURE, THE TIP OF THE DEVICE PROBE BROKE OFF BUT DID NOT FALL INTO THE PATIENT. THE PROCEDURE WAS CONVERTED FROM VIDEOLAPAROSCOPY TO OPEN SURGERY; HOWEVER, THIS CHANGE WAS UNRELATED TO THE DEVICE FAILURE. THE CONVERSION WAS NECESSITATED BY THE PATIENT'S CONDITION¿THE UTERUS WAS TOO LARGE AND OCCUPIED THE ENTIRE ABDOMINAL CAVITY. THE PROCEDURE WAS PROLONGED BY APPROXIMATELY 30 MINUTES TO REPLACE THE DAMAGED PROBE WITH ANOTHER DEVICE TO FINISH THE PROCEDURE. NO PATIENT HARM WAS REPORTED IN ASSOCIATION WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24297516 ↗MalfunctionFeb 9, 2026Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE FRONT-ACTUATED GRIP EXHIBITED THE TISSUE PAD HAD PEELED OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24293966 ↗MalfunctionFeb 3, 2026Fracture
IT WAS REPORTED THE SONICBEAT'S PROBE TIP BROKE DURING A LAPAROSCOPIC ADRENALECTOMY. IT WAS NOT SPECIFIED HOW THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24250922 ↗MalfunctionNov 6, 2025Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC SURGICAL INSTRUMENT EXHIBITED THE PROBE WAS BROKEN OFF AT 14.11 MILLIMETERS FROM ITS DISTAL END, BROKEN PROBE TIP WAS RETURNED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23486765 ↗InjuryMay 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED DURING SEDATION OF A THERAPEUTIC TRANSANAL ENDOSCOPIC OPERATION THE GENERATOR REPORTED AN ERROR A SELF TEST WAS COMPLETED WITH NO EFFECT AND IT WAS SEEN THAT THE JAW OF THE INSTRUMENT HAD FALLEN OFF INSIDE OF THE PATIENT. THERE PROCEDURE WAS NOT PROLONGED AND THERE WERE NO REPORTS OF PATIENT INJURY.
Patient problems: Foreign Body In Patient
MAUDE report 21994063 ↗MalfunctionApr 25, 2025Degraded; Degraded
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE INSTRUMENT EXHIBITED A WORN, LIFTED UP TEFLON PAD WHICH EXPOSED THE JAW METAL PART DUE TO SEVERE WEAR. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21909155 ↗MalfunctionMar 21, 2025Material Fragmentation; Material Fragmentation
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21662331 ↗MalfunctionMar 7, 2025Peeled/Delaminated; Peeled/Delaminated
NO ADDITIONAL INFO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21543799 ↗MalfunctionMar 7, 2025Degraded; Degraded
IT WAS REPORTED THE SONICBEAT SURGICAL INSTRUMENT'S SCISSOR PAD STARTED TO WEAR DURING BODY INSERTION WHILE PERFORMING A HEPATO-BILIARY-PANCREATIC PROCEDURE. THE EVENT RESULTED IN A DELAY OF 30 MINUTES OR LESS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AFTER BEING REPLACED WITH A SECOND DEVICE OF THE SAME MODEL. THERE WAS NO REPORT OF DEVICE COMPONENTS FALLING WITHIN PATIENT ANATOMY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21543709 ↗MalfunctionJan 15, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A TISSUE PAD THAT PEELED OFF WHEN ENERGIZED FOR NORMAL USE. THE ISSUE OCCURRED DURING A THERAPEUTIC LIVER RESECTION PROCEDURE THAT WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21164492 ↗MalfunctionJan 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21068396 ↗MalfunctionDec 10, 2024Peeled/Delaminated; Peeled/Delaminated
NO NEW INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20897250 ↗MalfunctionNov 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THERAPEUTIC RAFAELO PROCEDURE THE TISSUE PAD OF THE SUBJECT DEVICE CAME OFF. THE ISSUE WAS FOUND DURING SEDATION AND THE DEVICE WAS INSIDE THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER MODEL SONOSURGE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20759092 ↗InjuryNov 18, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A THERAPEUTIC LAPAROSCOPIC SURGERY, THE PROBE FELL OFF INTO THE PATIENT'S ABDOMINAL CAVITY. IT TOOK LESS THAN 15 MINUTES TO RETRIEVE THE FALLEN PROBE WITH FORCEPS. THE PROCEDURE WAS FINISHED WITH A SIMILAR DEVICE. THERE WAS NO HARM OR INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20716374 ↗InjuryNov 18, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A THERAPEUTIC LAPAROSCOPIC SURGERY, THE PROBE FELL OFF INTO THE PATIENT'S ABDOMINAL CAVITY. IT TOOK LESS THAN 15 MINUTES TO RETRIEVE THE FALLEN PROBE WITH FORCEPS. THE PROCEDURE WAS FINISHED WITH A SIMILAR DEVICE. THERE WAS NO HARM OR INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20714806 ↗InjuryNov 15, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A THERAPEUTIC MYOMECTOMY, THE TEFLON PAD WAS PARTICULARLY DAMAGED. UPON INSPECTION OF THE INSTRUMENT, A PART OF THE TEFLON PAD WAS MISSING, AND ANOTHER PART WAS SEALED IN THE GROOVE ON THE UPPER SIDE OF THE JAWS. INITIALLY, THE STAFF COULD NOT DETERMINE EXACTLY WHERE THE REMAINING PART OF THE TEFLON WAS LOCATED. THE OPERATING DOCTOR STATED THAT IT WAS PROBABLY SUCTIONED OUT, AND IF IT WAS A PIECE OF TEFLON, IT SHOULD NOT BE AN ISSUE SINCE CLIPS ARE MADE FROM SIMILAR MATERIAL AND REMAIN IN THE BODY. THE VIDEO RECORDING WAS REVIEWED, AND ACCORDING TO THE FOOTAGE, THE INSTRUMENT WAS DAMAGED APPROXIMATELY ON THE THIRD ACTIVATION OF THE DEVICE. APPARENTLY, TWO SMALL PIECES THE SIZE OF PINHEADS FELL OFF, WHICH INITIALLY LOOKED LIKE STEAM DROPLETS. ACCORDING TO THE DOCTOR, THEY WERE LIKELY SUCTIONED OUT DURING THE ABDOMINAL WASH. THE REMAINING DAMAGED PART WAS VISIBLY REMOVED FROM THE ABDOMINAL CAVITY AS PART OF THE INSTRUMENT AND LIKELY FELL OFF DURING THE SUBSEQUENT CLEANING OF THE TOOL, ACCORDING TO THE VIDEO RECORDING. THIS WAS PROBABLY WHY THE TEFLON PAD WAS NOT FOUND AFTERWARD, CAUSING CONCERN. THE PROCEDURE WAS COMPLETED WITHOUT ANY REPORTS OF PATIENT HA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20696529 ↗InjuryNov 8, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE PROBE TIP BROKE AND FELL INTO THE PATIENT'S ABDOMINAL CAVITY AND WAS REMOVED IMMEDIATELY BY THE MEDICAL STAFF USING FORCEPS. THIS CREATED A 2 MINUTE DELAY. THE DEVICE WAS INSPECTED PRIOR TO USE AND THERE WAS NO ABNORMALITY NOTED. THE PROCEDURE WAS COMPLETED WITH A BACK UP (SAME) DEVICE WITHOUT DELAY OR PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20636765 ↗MalfunctionOct 29, 2024Material Separation; Material Separation
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE ULTRASONIC SURGICAL INSTRUMENT EXHIBITED TISSUE PAD PARTIAL SEPARATION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20556259 ↗MalfunctionOct 28, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED, THE ULTRASONIC SURGICAL INSTRUMENT TIP OF THE PROBE FELL OFF WHEN IT WAS REMOVED FROM THE PATIENT. IT WAS NOT INSIDE THE BODY, BUT BROKE AND FELL OFF DURING CLEANING. THE ISSUE OCCURRED DURING A THERAPEUTIC LAPAROSCOPIC MYOMA REMOVAL. THERE WERE NO REPORTS OF PATIENT HARM. THE PROCEDURE WAS COMPLETED AFTER CHANGING TO A DIFFERENT DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20553473 ↗MalfunctionOct 25, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC SURGICAL DEVICE EXHIBITED TISSUE PAD PEELING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20533997 ↗MalfunctionOct 23, 2024Break; Break
IT WAS REPORTED THAT THE PROBE OF THE SONICBEAT BROKE WHEN IT WAS OUTPUTTED 20-30 MINUTES AFTER THE START. THE ISSUE OCCURRED DURING THE THERAPEUTIC CHOLECYSTECTOMY. THE DEVICE DID NOT FALL OFF. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20519068 ↗MalfunctionOct 23, 2024Break; Break
IT WAS REPORTED THE SONIBEAT TIP BROKE DURING THE PROBE CHECK. THE ISSUE OCCURRED DURING A THERAPEUTIC THYROIDECTOMY SURGERY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20514317 ↗
FDA clearance and classification
The FDA regulates product code KNS (unit, electrosurgical, endoscopic (with or without accessories)) under 21 CFR 876.4300 as a Class II device. 21 CFR 876.4300 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FC
Has the SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FC been recalled?
No FDA recall record in our data names this exact device identifier (DI 04953170370540), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FC?
The FDA's MAUDE database lists 379 adverse event reports for this device, received between Oct 10, 2017 and Sep 30, 2026. In those reports, the most common specific device problems are “Material Separation” (125), “Peeled/Delaminated” (79) and “Break” (70); the most common patient problems are “Foreign Body In Patient” (3), “Hematoma” (2) and “Bacterial Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FC?
None of the reports are classified as deaths, 31 as injuries and 348 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FC, and is it still sold?
The FDA's device database lists Olympus Medical Systems Corp. as the company responsible for it (brand SONICBEAT 5 mm, 35 cm, Front-actuated Grip, model SB-0535FC). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall261 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall131 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
- Hydratome RX 44 M00583040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall98 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page