TRUEtome 44 M00584170Recalls & reported problems
Boston Scientific Corporation · Cannulating Sphincterotome · Class 2
Quick summary
This Boston Scientific TRUEtome 44 is an endoscopic cutting tool used in bile duct procedures (ERCP). Its 242 reports mostly describe the cutting wire breaking, bending or getting stuck and failing to retract, and some procedures were finished with another device. The reports include no deaths and 1 injury. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 237
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 237 adverse event reports that list this exact device (DI 08714729827634), received between Mar 3, 2017 and Sep 16, 2026 — classified as 1 injury, 236 malfunctions.
The most frequently reported specific device problem is “Break” (171 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break171
- Adverse Event Without Identified Device or Use Problem35
- Positioning Problem23
- Material Deformation21
- Device Contaminated During Manufacture or Shipping18
- Deformation Due to Compressive Stress11
- Device Dislodged or Dislocated10
- Use of Device Problem8
Patient problems
- No Clinical Signs, Symptoms or Conditions174
- No Consequences Or Impact To Patient63
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 25 most recent of 237 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Break; Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50793776 ↗MalfunctionSep 15, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50779521 ↗MalfunctionSep 10, 2026Deformation Due to Compressive Stress
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50720825 ↗MalfunctionAug 31, 2026Break
It was reported to Boston Scientific Corporation that a TRUEtome 44 was attempted to be used in the bile duct during an Endoscopic Retrograde Cholangiopancreatography (ERCP) procedure for a Duodenal papillotomy performed on (b)(6) 2026. During the procedure the cutting wire was fractured, and it could not be used. It was reported that no part of the cutting wire detached and fell into the patient. The procedure was completed with another of the same device. There was no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50533777 ↗MalfunctionAug 13, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50244682 ↗MalfunctionAug 13, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50243521 ↗MalfunctionJul 10, 2026Break; Failure to Conduct; Deformation Due to Compressive Stress
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT CHOLANGITIS PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE PHYSICIAN INSTRUCTED HER ASSISTANT TO TIGHTEN THE DEVICE TO FACILITATE FURTHER INCISION. AFTER PRESSING ON THE CAUTERY, NO CURRENT WAS GENERATED. PRIOR TO CONTINUING THE PROCEDURE, THE PHYSICIAN INSPECTED THE DEVICE AND OBSERVED THAT THE METAL TIP OF THE SPHINCTEROTOME WAS FRACTURED. A PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED AND SHOWED THAT THE CUTTING WIRE WAS BROKEN AND KINKED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE OKAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25788844 ↗MalfunctionJun 25, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TURETOME 44 WAS ATTEMPTED TO BE USED IN THE DUODENAL PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE WAS FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25622813 ↗MalfunctionJun 10, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS CHECKED IN THE STOCKROOM. THE EXACT EVENT DATE WAS UNKNOWN. DURING THE INVENTORY, A FOREIGN OBJECT WAS FOUND IN THE PACKAGING. SPECIFICALLY, A HAIR WAS OBSERVED WITHIN THE STERILE POUCH, VISIBLE ON THE WHITE SECTION OF THE PACKAGING ON THE RIGHT SIDE. NO PATIENT AND PROCEDURE WERE INVOLVED IN THIS EVENT. NOTE: PHOTOS OF THE COMPLAINT DEVICE INSIDE THE UNOPENED PACKAGING WERE PROVIDED BY THE CUSTOMER AND SHOWS A HAIR-LIKE PARTICLE INSIDE THE STERILE DEVICE PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25452705 ↗MalfunctionMay 25, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED IN THE DUODENUM DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE WAS FRACTURED AND COULD NOT BE USED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25271742 ↗MalfunctionApr 16, 2026Deformation Due to Compressive Stress; Difficult to Advance; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS PREPARED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT CHOLANGIOCARCINOMA PERFORMED ON (B)(6) 2026. DURING THE PREPARATION OUTSIDE THE PATIENT, IT WAS REPORTED THAT THE GUIDEWIRE CAME OUT AT THE MIDDLE OF THE TRUETOME AND NOT THOUGH THE DISTAL TIP. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF DEVICE CUTTING WIRE KINKED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24899604 ↗MalfunctionApr 15, 2026Deformation Due to Compressive Stress
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE WIRE ANCHOR HAS FRACTURED FROM THE CATHETER, BUT THE CUT WIRE WAS STILL INTACT AND COULD NOT BE USED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24889549 ↗MalfunctionApr 11, 2026Deformation Due to Compressive Stress
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE TO TREAT COMMON BILE DUCT STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE WAS BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24849476 ↗MalfunctionApr 3, 2026Break; Use of Device Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONES PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE OF THE TRUETOME 44 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. ADDITIONALLY, IT WAS ALSO REPORTED THAT THE EXPOSED CUTTING WIRE DID NOT FULLY EXIT THE ENDOSCOPE WHEN THE DEVICE WAS ENERGIZED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THE COMPLAINANT REPORTED THAT THE EXPOSED CUTTING WIRE DID NOT FULLY EXIT THE ENDOSCOPE WHEN THE DEVICE WAS ENERGIZED.; HOWEVER, ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), WARNING: VERIFY THAT THE CUTTING WIRE HAS EXITED THE ENDOSCOPE BY VISUALIZING IT ON THE ENDOSCOPE MONITOR. FAILURE TO DO SO MAY RESULT IN CONTACT BETWEEN THE CUTTING WIRE AND THE ENDOSCOPE WHILE ELECTRICAL CURRENT IS APPLIED. THIS MAY CAUSE GROUNDING, WHICH CAN RESULT IN PATIENT INJURY, OPERATOR INJURY,
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24778328 ↗MalfunctionApr 2, 2026Break; Use of Device Problem; Deformation Due to Compressive Stress
IT WAS REPORTED THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT CHOLEDOCHOLITHIASIS PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE SPHINCTEROTOMY TOME WAS FOUND TO BE FRACTURED AT THE TIP AFTER INSERTING INTO THE DUODENOSCOPE. A PHOTO WAS PROVIDED BY THE CUSTOMER SHOWING THAT THE CUTTING WIRE WAS BROKEN AND KINKED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. ADDITIONAL INFORMATION WAS RECEIVED ON MARCH 31, 2026. IT WAS REPORTED THAT THE DEVICE WAS ENERGIZED WHEN NOT IN CONTACT WITH TISSUE. NOTE: THE COMPLAINANT REPORTED THAT THE DEVICE WAS ENERGIZED WHEN NOT IN CONTACT WITH TISSUE; HOWEVER, PER THE INSTRUCTIONS FOR USE (IFU), PRECAUTION: IT IS RECOMMENDED TO MAINTAIN DIRECT AND CONSTANT CONTACT WITH TISSUE WHEN APPLYING ELECTROCAUTERY CURRENT. FAILURE TO DO SO MAY RESULT IN BROKEN CUT WIRE, DAMAGE TO THE ENDOSCOPE AND/OR PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24765917 ↗MalfunctionApr 1, 2026Break; Leak/Splash; Deformation Due to Compressive Stress
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS PREPARED TO BE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PREPARATION, IT WAS REPORTED THAT A BROKEN CUT WIRE WAS NOTICED STRAIGHT OUT OF PACKAGING. ADDITIONALLY, A MALFORMED LUER LOCK WAS ALSO NOTED. A PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED AND SHOWED THE CUTTING WIRE WAS BROKEN AND KINKED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24759977 ↗MalfunctionMar 20, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUTOME 44 WAS USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE OF THE TRUTOME 44 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24654006 ↗MalfunctionMar 20, 2026Deformation Due to Compressive Stress
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONE REMOVAL PERFORMED ON (B)(6), 2026. DURING THE PROCEDURE, THE CUTTING WIRE WAS BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24652436 ↗MalfunctionMar 20, 2026Break; Use of Device Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, IT WAS REPORTED THAT THE CUTTING WIRE WAS CUT OFF DURING USE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THE REPORTED EVENT MAY SUGGEST THAT THE CUTTING WIRE BROKE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24651518 ↗MalfunctionMar 20, 2026Break; Use of Device Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE OF THE TRUETOME 44 BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT THE EXPOSED CUTTING WIRE WAS NOT FULLY EXITED THE ENDOSCOPE WHEN THE DEVICE WAS ENERGIZED; HOWEVER, ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), VERIFY THAT THE CUTTING WIRE HAS EXITED THE ENDOSCOPE BY VISUALIZING IT ON THE ENDOSCOPE MONITOR. FAILURE TO DO SO MAY RESULT IN CONTACT BETWEEN THE CUTTING WIRE AND THE ENDOSCOPE WHILE ELECTRICAL CURRENT IS APPLIED. THIS MAY CAUSE GROUNDING, WHICH CAN RESULT IN PATIENT INJURY, OPERATOR INJURY, A BROKEN CUTTING WIRE, AND/OR DAMAGE TO THE ENDOSCOPE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24650572 ↗MalfunctionMar 16, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE WAS FRACTURED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24613088 ↗MalfunctionMar 13, 2026Break; Use of Device Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24596155 ↗MalfunctionMar 2, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA OF VATER (AOV) DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24478881 ↗MalfunctionFeb 18, 2026Break; Deformation Due to Compressive Stress
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, WHEN PULLING THE HANDLE, THERE IS A PHENOMENON WHERE THE WIRE TIP PART WAS BENT MUCH MORE THAN THE NORMAL RANGE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED REPORTABLE BASED ON THE INVESTIGATION RESULTS: WIRE BROKEN. PLEASE SEE BLOCK H10 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24382279 ↗MalfunctionFeb 16, 2026Break; Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem
NOTE: THIS REPORT PERTAINS TO ONE OF THE TWO TRUETOME 44 USED DURING THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT BILE DUCT STONES PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, BLADE PORTION OF THE TRUETOME BROKE OFF AND THE SHAFT MAY HAVE BEEN SIGNIFICANTLY TWISTED. A SECOND TRUETOME 44 WAS USED; HOWEVER, THE SAME PROBLEM OCCURRED. IT WAS REPORTED THAT THE PROCEDURE WAS COMPLETED USING THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24345454 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
- Hydratome RX 44 M00583040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall98 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.