TRUEtome 44 M00584170Recalls & reported problems

Boston Scientific Corporation · Cannulating Sphincterotome · Class 2

All models of TRUEtome 44 (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific TRUEtome 44 is an endoscopic cutting tool used in bile duct procedures (ERCP). Its 242 reports mostly describe the cutting wire breaking, bending or getting stuck and failing to retract, and some procedures were finished with another device. The reports include no deaths and 1 injury. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
237
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 237 adverse event reports that list this exact device (DI 08714729827634), received between Mar 3, 2017 and Sep 16, 2026 — classified as 1 injury, 236 malfunctions.

The most frequently reported specific device problem is “Break” (170 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 3 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 7 reportsDec 2024: 14 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 4 reportsApr 2025: 3 reportsMay 2025: 5 reportsJun 2025: 0 reportsJul 2025: 4 reportsAug 2025: 3 reportsSep 2025: 3 reportsOct 2025: 2 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 3 reportsFeb 2026: 4 reportsMar 2026: 7 reportsApr 2026: 6 reportsMay 2026: 2 reportsJun 2026: 2 reportsJul 2026: 1 reportsAug 2026: 2 reportsSep 2026: 3 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break170
  • Adverse Event Without Identified Device or Use Problem36
  • Positioning Problem23
  • Material Deformation21
  • Device Contaminated During Manufacture or Shipping18
  • Deformation Due to Compressive Stress11
  • Device Dislodged or Dislocated10
  • Use of Device Problem8

Patient problems

  • No Clinical Signs, Symptoms or Conditions174
  • No Consequences Or Impact To Patient63
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 25 most recent of 237 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Break; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50793776 ↗
  • MalfunctionSep 15, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50779521 ↗
  • MalfunctionSep 10, 2026Deformation Due to Compressive Stress

    It was reported to Boston Scientific Corporation that a TRUEtome 44 was attempted to be used in the Common bile duct stone during an Endoscopic Retrograde Cholangiopancreatography (ERCP) procedure for a Duodenal papillotomy performed on (b)(6) 2026.\\During the procedure, the cutting wire fractured and could not be used. It was reported that the cutting wire was intact but the wire anchor dislodged from the catheter. No part of the cutting wire detached and fell into the patient. The procedure was completed with another of the same device.\\There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be stable.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50720825 ↗
  • MalfunctionAug 31, 2026Break

    It was reported to Boston Scientific Corporation that a TRUEtome 44 was attempted to be used in the bile duct during an Endoscopic Retrograde Cholangiopancreatography (ERCP) procedure for a Duodenal papillotomy performed on (b)(6) 2026. During the procedure the cutting wire was fractured, and it could not be used. It was reported that no part of the cutting wire detached and fell into the patient. The procedure was completed with another of the same device. There was no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be stable.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50533777 ↗
  • MalfunctionAug 27, 2026Failure to Align; Material Twisted/Bent

    It was reported to Boston Scientific Corporation that a TRUEtome 44 was attempted to be used in the Biliary tract during an Endoscopic Retrograde Cholangiopancreatography (ERCP) procedure to treat Bile duct stone performed on (b)(6)2026.During the procedure, the cutting wire at the tip of the tome was stuck and could not retract inside the extrusion after being bowed. Photos of the complaint device were provided and showed that the cutting wire was twisted around the working length. The procedure was completed with another of the same device.There was no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be stable.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50482434 ↗
  • MalfunctionJul 10, 2026Break; Failure to Conduct; Deformation Due to Compressive Stress

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT CHOLANGITIS PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE PHYSICIAN INSTRUCTED HER ASSISTANT TO TIGHTEN THE DEVICE TO FACILITATE FURTHER INCISION. AFTER PRESSING ON THE CAUTERY, NO CURRENT WAS GENERATED. PRIOR TO CONTINUING THE PROCEDURE, THE PHYSICIAN INSPECTED THE DEVICE AND OBSERVED THAT THE METAL TIP OF THE SPHINCTEROTOME WAS FRACTURED. A PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED AND SHOWED THAT THE CUTTING WIRE WAS BROKEN AND KINKED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE OKAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25788844 ↗
  • MalfunctionJun 25, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TURETOME 44 WAS ATTEMPTED TO BE USED IN THE DUODENAL PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE WAS FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25622813 ↗
  • MalfunctionJun 10, 2026Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS CHECKED IN THE STOCKROOM. THE EXACT EVENT DATE WAS UNKNOWN. DURING THE INVENTORY, A FOREIGN OBJECT WAS FOUND IN THE PACKAGING. SPECIFICALLY, A HAIR WAS OBSERVED WITHIN THE STERILE POUCH, VISIBLE ON THE WHITE SECTION OF THE PACKAGING ON THE RIGHT SIDE. NO PATIENT AND PROCEDURE WERE INVOLVED IN THIS EVENT. NOTE: PHOTOS OF THE COMPLAINT DEVICE INSIDE THE UNOPENED PACKAGING WERE PROVIDED BY THE CUSTOMER AND SHOWS A HAIR-LIKE PARTICLE INSIDE THE STERILE DEVICE PACKAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25452705 ↗
  • MalfunctionMay 25, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED IN THE DUODENUM DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE WAS FRACTURED AND COULD NOT BE USED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25271742 ↗
  • MalfunctionMay 14, 2026Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT GALLSTONES PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE OF THE PAPILLOTOME BROKE SPONTANEOUSLY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25167091 ↗
  • MalfunctionApr 16, 2026Deformation Due to Compressive Stress; Difficult to Advance; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS PREPARED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT CHOLANGIOCARCINOMA PERFORMED ON (B)(6) 2026. DURING THE PREPARATION OUTSIDE THE PATIENT, IT WAS REPORTED THAT THE GUIDEWIRE CAME OUT AT THE MIDDLE OF THE TRUETOME AND NOT THOUGH THE DISTAL TIP. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF DEVICE CUTTING WIRE KINKED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24899604 ↗
  • MalfunctionApr 15, 2026Deformation Due to Compressive Stress

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE WIRE ANCHOR HAS FRACTURED FROM THE CATHETER, BUT THE CUT WIRE WAS STILL INTACT AND COULD NOT BE USED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24889549 ↗
  • MalfunctionApr 11, 2026Deformation Due to Compressive Stress

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE TO TREAT COMMON BILE DUCT STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE WAS BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24849476 ↗
  • MalfunctionApr 3, 2026Break; Use of Device Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONES PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE OF THE TRUETOME 44 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. ADDITIONALLY, IT WAS ALSO REPORTED THAT THE EXPOSED CUTTING WIRE DID NOT FULLY EXIT THE ENDOSCOPE WHEN THE DEVICE WAS ENERGIZED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THE COMPLAINANT REPORTED THAT THE EXPOSED CUTTING WIRE DID NOT FULLY EXIT THE ENDOSCOPE WHEN THE DEVICE WAS ENERGIZED.; HOWEVER, ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), WARNING: VERIFY THAT THE CUTTING WIRE HAS EXITED THE ENDOSCOPE BY VISUALIZING IT ON THE ENDOSCOPE MONITOR. FAILURE TO DO SO MAY RESULT IN CONTACT BETWEEN THE CUTTING WIRE AND THE ENDOSCOPE WHILE ELECTRICAL CURRENT IS APPLIED. THIS MAY CAUSE GROUNDING, WHICH CAN RESULT IN PATIENT INJURY, OPERATOR INJURY,

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24778328 ↗
  • MalfunctionApr 2, 2026Break; Use of Device Problem; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT CHOLEDOCHOLITHIASIS PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE SPHINCTEROTOMY TOME WAS FOUND TO BE FRACTURED AT THE TIP AFTER INSERTING INTO THE DUODENOSCOPE. A PHOTO WAS PROVIDED BY THE CUSTOMER SHOWING THAT THE CUTTING WIRE WAS BROKEN AND KINKED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. ADDITIONAL INFORMATION WAS RECEIVED ON MARCH 31, 2026. IT WAS REPORTED THAT THE DEVICE WAS ENERGIZED WHEN NOT IN CONTACT WITH TISSUE. NOTE: THE COMPLAINANT REPORTED THAT THE DEVICE WAS ENERGIZED WHEN NOT IN CONTACT WITH TISSUE; HOWEVER, PER THE INSTRUCTIONS FOR USE (IFU), PRECAUTION: IT IS RECOMMENDED TO MAINTAIN DIRECT AND CONSTANT CONTACT WITH TISSUE WHEN APPLYING ELECTROCAUTERY CURRENT. FAILURE TO DO SO MAY RESULT IN BROKEN CUT WIRE, DAMAGE TO THE ENDOSCOPE AND/OR PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24765917 ↗
  • MalfunctionApr 1, 2026Break; Leak/Splash; Deformation Due to Compressive Stress

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS PREPARED TO BE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PREPARATION, IT WAS REPORTED THAT A BROKEN CUT WIRE WAS NOTICED STRAIGHT OUT OF PACKAGING. ADDITIONALLY, A MALFORMED LUER LOCK WAS ALSO NOTED. A PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED AND SHOWED THE CUTTING WIRE WAS BROKEN AND KINKED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24759977 ↗
  • MalfunctionMar 20, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUTOME 44 WAS USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE OF THE TRUTOME 44 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24654006 ↗
  • MalfunctionMar 20, 2026Deformation Due to Compressive Stress

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONE REMOVAL PERFORMED ON (B)(6), 2026. DURING THE PROCEDURE, THE CUTTING WIRE WAS BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24652436 ↗
  • MalfunctionMar 20, 2026Break; Use of Device Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, IT WAS REPORTED THAT THE CUTTING WIRE WAS CUT OFF DURING USE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THE REPORTED EVENT MAY SUGGEST THAT THE CUTTING WIRE BROKE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24651518 ↗
  • MalfunctionMar 20, 2026Break; Use of Device Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE OF THE TRUETOME 44 BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT THE EXPOSED CUTTING WIRE WAS NOT FULLY EXITED THE ENDOSCOPE WHEN THE DEVICE WAS ENERGIZED; HOWEVER, ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), VERIFY THAT THE CUTTING WIRE HAS EXITED THE ENDOSCOPE BY VISUALIZING IT ON THE ENDOSCOPE MONITOR. FAILURE TO DO SO MAY RESULT IN CONTACT BETWEEN THE CUTTING WIRE AND THE ENDOSCOPE WHILE ELECTRICAL CURRENT IS APPLIED. THIS MAY CAUSE GROUNDING, WHICH CAN RESULT IN PATIENT INJURY, OPERATOR INJURY, A BROKEN CUTTING WIRE, AND/OR DAMAGE TO THE ENDOSCOPE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24650572 ↗
  • MalfunctionMar 16, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE WAS FRACTURED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24613088 ↗
  • MalfunctionMar 13, 2026Break; Use of Device Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24596155 ↗
  • MalfunctionMar 2, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA OF VATER (AOV) DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24478881 ↗
  • MalfunctionFeb 18, 2026Break; Deformation Due to Compressive Stress

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, WHEN PULLING THE HANDLE, THERE IS A PHENOMENON WHERE THE WIRE TIP PART WAS BENT MUCH MORE THAN THE NORMAL RANGE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED REPORTABLE BASED ON THE INVESTIGATION RESULTS: WIRE BROKEN. PLEASE SEE BLOCK H10 FOR FULL INVESTIGATION DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24382279 ↗
  • MalfunctionFeb 16, 2026Break; Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem

    NOTE: THIS REPORT PERTAINS TO ONE OF THE TWO TRUETOME 44 USED DURING THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 44 WAS USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT BILE DUCT STONES PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, BLADE PORTION OF THE TRUETOME BROKE OFF AND THE SHAFT MAY HAVE BEEN SIGNIFICANTLY TWISTED. A SECOND TRUETOME 44 WAS USED; HOWEVER, THE SAME PROBLEM OCCURRED. IT WAS REPORTED THAT THE PROCEDURE WAS COMPLETED USING THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24345454 ↗

FDA clearance and classification

The FDA regulates product code KNS (unit, electrosurgical, endoscopic (with or without accessories)) under 21 CFR 876.4300 as a Class II device. 21 CFR 876.4300 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the TRUEtome 44 M00584170

Has the TRUEtome 44 M00584170 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08714729827634), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the TRUEtome 44 M00584170?

The FDA's MAUDE database lists 237 adverse event reports for this device, received between Mar 3, 2017 and Sep 16, 2026. In those reports, the most common specific device problems are “Break” (170), “Positioning Problem” (23) and “Material Deformation” (21); the most common patient problem is “Hemorrhage/Blood Loss/Bleeding” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TRUEtome 44 M00584170?

None of the reports are classified as deaths, 1 as an injury and 236 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the TRUEtome 44 M00584170, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand TRUEtome 44, model M00584170). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Unit, electrosurgical, endoscopic (with or without accessories) devices

All Unit, electrosurgical, endoscopic (with or without accessories) devices →

Sources and updates