Autotome™ RX 44 M00545170Recalls & reported problems
Boston Scientific Corporation · Cannulating Sphincterotome · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Boston Scientific Autotome sphincterotome is an electrosurgical cutting tool used through an endoscope. Most reports describe the cutting wire or tool breaking, and some describe it not bending, not cutting or moving out of place. There are 258 reports, with no deaths and 7 injuries. 1 recall is open, begun in June 2024, over foreign material on the cutting wire that can stop the tool from bending and make procedures longer.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 262
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 262 adverse event reports that list this exact device (DI 08714729444749), received between Jan 11, 2018 and Sep 28, 2026 — classified as 7 injuries, 255 malfunctions.
The most frequently reported specific device problem is “Break” (171 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jun 12, 2024Open, Classified
Foreign material may be present on the enclosed section of the cutting wire which may lead to inability to optimally bow the sphincterotome. This issue may lead to prolongation of procedure.
Z-2350-2024 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Break171
- Adverse Event Without Identified Device or Use Problem43
- Positioning Problem37
- Material Deformation29
- Device Dislodged or Dislocated21
- Deformation Due to Compressive Stress18
- Device Contaminated During Manufacture or Shipping13
- Failure to Fold13
Patient problems
- No Clinical Signs, Symptoms or Conditions182
- No Consequences Or Impact To Patient74
- Pain3
- Hemorrhage/Blood Loss/Bleeding2
- No Known Impact Or Consequence To Patient2
- Perforation2
- Hematoma1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 262 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923789 ↗MalfunctionSep 23, 2026Break; Use of Device Problem; Difficult to Advance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873537 ↗MalfunctionSep 18, 2026Failure to Fold; Deformation Due to Compressive Stress; Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50831300 ↗MalfunctionSep 17, 2026Break; Material Twisted/Bent
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50809545 ↗MalfunctionAug 22, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50389934 ↗MalfunctionAug 21, 2026Break; Deformation Due to Compressive Stress
It was reported to Boston Scientific Corporation that an Autotome RX 44 was attempted to be used in the Duodenal Papilla during an Endoscopic Retrograde Cholangiopancreatography (ERCP) procedure to treat Bile duct stones performed on (b)(6) 2026.During the procedure, the cutting wire was fractured when the direction of the bow was adjusted for the first time. A photo of the complaint device was provided and showed that the cutting wire was broken and kinked. The procedure was completed with another of the same device.There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be Stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50360700 ↗MalfunctionJul 13, 2026Failure to Align; Deformation Due to Compressive Stress
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS ATTEMPTED TO BE USED IN THE INTESTINAL TRACT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STONE REMOVAL PROCEDURE TO TREAT TRANSDUODENAL ODDI'S SPHINCTEROPLASTY PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, WHEN BOWING, THE PLASTIC SHEATH AT THE AUTOTOME TIP WAS KINKED. AS A RESULT, THE CUTTING WIRE COULD NOT BE FULLY STRAIGHTENED. A PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED AND SHOWED THAT THE WORKING LENGTH WAS KINKED AND THE DEVICE IN A BOWED POSITION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25807891 ↗MalfunctionJul 3, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, WHEN THE PHYSICIAN PERFORMED THE SPHINCTEROTOMY USING THE DEVICE, IT WAS NOTED THAT THE CUTTING WIRE OF THE AUTOTOME BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25708379 ↗MalfunctionJun 19, 2026Break; Failure to Conduct
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT BILE DUCT STONES PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE DEVICE FUNCTIONED NORMALLY ON ITS FIRST USE. HOWEVER, THERE WAS NO RESPONSE WHEN IT WAS ACTIVATED A SECOND TIME, AND REPEATED ATTEMPTS PRODUCED THE SAME RESULT. THE AUTOTOME WAS THEN WITHDRAWN, AND EXAMINATION REVEALED THAT ITS TIP HAD FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THE REPORTED EVENT MAY SUGGEST THAT THE CUTTING WIRE BROKE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25566219 ↗MalfunctionMay 25, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS USED IN THE PAPILLA DURING A PAPILLO SPHINCTEROTOMY PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE CUTTING WIRE BREAK UPON SPHINCTEROTOMY. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25272206 ↗MalfunctionMay 25, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS PREPARED TO USE IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PREPARATION OUTSIDE THE PATIENT, THE PRODUCT WAS UNPACKED AND IT WAS FOUND THAT THE STEEL WIRE INSIDE THE HANDLE WAS FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25271748 ↗MalfunctionMay 23, 2026Deformation Due to Compressive Stress; Difficult to Advance
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS USED DURING A SPHINCTEROTOMY PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE DEVICE DOES NOT ALLOW THE GUIDEWIRE TO SLIDE. ALSO, THE DISTAL TIP DOES NOT BEND CORRECTLY AND LOSES ITS SHAPE MEMORY. DURING SPHINCTEROTOMY, THE CUTTING WIRE DETACHED FROM THE CATHETER. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25263555 ↗MalfunctionMay 5, 2026Break; Deformation Due to Compressive Stress
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS ATTEMPTED TO BE USED ON THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) AND BILIARY STENT IMPLANTATION PROCEDURE TO TREAT BILE DUCT STRICTURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE STEEL WIRE WAS FRACTURED WHEN PULLING OUT THE AUTOTOME. THE CUSTOMER PROVIDED PHOTOS SHOWING THAT THE CUTTING WIRE WAS BROKEN AND KINKED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25084758 ↗MalfunctionMay 4, 2026Failure to Conduct; Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS ATTEMPTED TO BE USED IN THE BILE DUCT DURING A TRANSDUODENOSCOPIC LITHOTOMY PROCEDURE TO TREAT BILE DUCT STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE DEVICE COULD NOT CUT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS: THE CUTTING WIRE KINKED. PLEASE REFER TO BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25074996 ↗MalfunctionApr 28, 2026Deformation Due to Compressive Stress
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE WAS BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25024065 ↗MalfunctionApr 13, 2026Break; Deformation Due to Compressive Stress; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A AUTOTOME RX 44 WAS USED ATTEMPTED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE TIP OF TOME WAS DEVIATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF CUTTING WIRE BROKEN AND KINKED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24866745 ↗MalfunctionApr 13, 2026Break; Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS USED IN THE DUODENAL PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, IT WAS REPORTED THAT THERE WAS A DEVIATION WITH THE TOMES TIP. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDINGS OF DEVICE CUTTING WIRE BROKEN AND KINKED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24865666 ↗MalfunctionApr 3, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS ATTEMPTED TO BE USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE CUTTING WIRE OF THE AUTOTOME RX 44 BROKE WHEN THE PHYSICIAN WAS TRYING TO MAKE A CUT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24777625 ↗MalfunctionMar 30, 2026Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6), 2026. DURING THE PREPARATION OUTSIDE THE PATIENT, IT WAS REPORTED THAT THE TOME WAS BENT AND THE INSERTION DIRECTION WAS INCORRECT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS: THE CUTTING WIRE KINKED. PLEASE REFER TO BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24731830 ↗MalfunctionMar 25, 2026Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE REMOVAL PERFORMED ON (B)(6), 2025. DURING THE PROCEDURE, THE TOME REACHED THE OPENING OF THE DUODENAL PAPILLA THROUGH THE DUODENOSCOPE, AND THE TIP DEVIATED AND COULD NOT FIND THE CORRECT ANGLE TO INSERT INTO THE DUODENAL PAPILLA. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF CUTTING WIRE KINKED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24683880 ↗MalfunctionMar 17, 2026Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE REMOVAL PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE DEVICE REACHED THE PAPILLA OPENING THROUGH THE DUODENOSCOPE, BUT THE TIP DEVIATED AND IT COULD NOT FIND THE CORRECT ANGLE TO INSERT INTO THE DUODENAL PAPILLA. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDINGS OF WIRE/ANCHOR DISLODGED/DISLOCATED AND CUTTING WIRE KINKED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24621459 ↗MalfunctionMar 16, 2026Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE REMOVAL PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE TOME WAS INSERTED THROUGH THE DUODENOSCOPE TO THE PAPILLA, BUT THE TIP OF THE TOME DEVIATED, AND IT WAS IMPOSSIBLE TO FIND THE CORRECT ANGLE TO INSERT INTO THE DUODENAL PAPILLA. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF CUTTING WIRE KINKED. PLEASE SEE BLOCK H11. FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24613072 ↗MalfunctionMar 16, 2026Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONES PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE TOME WAS ADVANCED TO THE PAPILLARY OPENING THROUGH THE DUODENOSCOPE. HOWEVER, THE TIP BECAME MISALIGNED AND COULD NOT ACHIEVE THE CORRECT ANGLE FOR CANNULATION OF THE DUODENAL PAPILLA. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS: THE CUTTING WIRE KINKED. PLEASE REFER TO BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24612472 ↗MalfunctionMar 13, 2026Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT COMMON BILE DUCT STONE REMOVAL PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE TOME WAS INSERTED THROUGH THE DUODENOSCOPE TO THE PAPILLA, BUT THE TIP OF THE TOME DEVIATED, AND IT WAS IMPOSSIBLE TO FIND THE CORRECT ANGLE TO INSERT INTO THE DUODENAL PAPILLA. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF CUTTING WIRE KINKED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24594239 ↗MalfunctionFeb 24, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 44 WAS USED IN A PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE OF THE AUTOTOME RX 44 WAS DAMAGED AND SEPARATED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24434154 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
- Hydratome RX 44 M00583040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall98 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.