Skip to content

Single Use Biliary Stent V PBD-V621R-1005Recalls & reported problems

Olympus Medical Systems Corp. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04953170185175), received between Oct 28, 2020 and Oct 28, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 28, 2020Break; Break

    WE RECEIVED THE FOLLOWING REPORT. *DURING AN ENDOSCOPIC BILIARY DRAINAGE, THE SUBJECT DEVICE WAS USED. WHEN THE USER TRIED TO EXCHANGE THE DEVICE FOR A NEW ONE, THE USER FOUND THAT THE FLAP ON THE BILE DUCT SIDE WAS BROKEN OFF AND THE DEVICE MIGRATED INSIDE THE BILE DUCT. THE DEVICE WAS RETRIEVED FROM THE BILE DUCT TO THE DUODENUM WITH A BALLOON. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10745642 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page