Single Use Rotatable Clip Fixing Device HX-201UR-135 (UDI 04953170193040)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2 · Implant
All models of Single Use Rotatable Clip Fixing Device (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170193040), received between Apr 26, 2022 and Apr 26, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break3
- Failure to Fire3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 26, 2022Break; Failure to Fire; Break; Failure to Fire
A USER FACILITY REPORTED TO OLYMPUS THAT FOLLOWING POLYP REMOVAL IN THE RECTUM, DURING A COLONOSCOPY PROCEDURE, AN ATTEMPT WAS MADE TO APPLY A CLIP TO A MUCOSAL BREACH, HOWEVER, THE CLIP DID NOT RELEASE FROM THE APPLICATOR AFTER CLOSURE AND REMAINED ATTACHED TO THE CONTROL WIRE, WHICH BROKE AT THE HANDLE. THE ENDOSCOPE WAS REMOVED, LEAVING THE CLIP WITH THE CONTROL WIRE IN PLACE. THE SCOPE WAS REINSERTED AND THE CLIP FORCIBLY REMOVED FROM THE PATIENT. THE INTENDED PROCEDURE WAS COMPLETED USING CLIPS FROM A DIFFERENT MANUFACTURER. THE PROCEDURE WAS COMPLETED AFTER A DELAY, CHARACTERIZED AS A FEW MINUTES. THERE WAS NO ADVERSE HEALTH EFFECT TO THE PATIENT THAT OCCURRED, ATTRIBUTED TO THE DELAY. THERE WAS NO PATIENT INJURY, ASSOCIATED WITH THE EVENT, REPORTED TO OLYMPUS. THIS IS REPORT #3 OF 3 DUE TO MULTIPLE EVENTS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14210665 ↗MalfunctionApr 26, 2022Break; Failure to Fire; Break; Failure to Fire
A USER FACILITY REPORTED TO OLYMPUS THAT FOLLOWING POLYP REMOVAL IN THE RECTUM, DURING A COLONOSCOPY PROCEDURE, AN ATTEMPT WAS MADE TO APPLY THE CLIP TO A MUCOSAL BREACH, HOWEVER, THE CLIP DID NOT RELEASE FROM THE APPLICATOR AFTER CLOSURE AND REMAINED ATTACHED TO THE CONTROL WIRE, WHICH BROKE AT THE HANDLE. THE ENDOSCOPE WAS REMOVED, LEAVING THE CLIP WITH THE CONTROL WIRE IN PLACE. THE SCOPE WAS REINSERTED AND THE CLIP FORCIBLY REMOVED FROM THE PATIENT. THE INTENDED PROCEDURE WAS COMPLETED USING CLIPS FROM A DIFFERENT MANUFACTURER. THE PROCEDURE WAS COMPLETED AFTER A DELAY, CHARACTERIZED AS A FEW MINUTES. THERE WAS NO ADVERSE HEALTH EFFECT TO THE PATIENT THAT OCCURRED, ATTRIBUTED TO THE DELAY. THERE WAS NO PATIENT INJURY, ASSOCIATED WITH THE EVENT, REPORTED TO OLYMPUS. THIS IS REPORT #1 OF 3 DUE TO MULTIPLE EVENTS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14210445 ↗MalfunctionApr 26, 2022Break; Failure to Fire; Break; Failure to Fire
A USER FACILITY REPORTED TO OLYMPUS THAT FOLLOWING POLYP REMOVAL IN THE RECTUM, DURING A COLONOSCOPY PROCEDURE, AN ATTEMPT WAS MADE TO APPLY A CLIP TO A MUCOSAL BREACH, HOWEVER, THE CLIP DID NOT RELEASE FROM THE APPLICATOR AFTER CLOSURE AND REMAINED ATTACHED TO THE CONTROL WIRE, WHICH BROKE AT THE HANDLE. THE ENDOSCOPE WAS REMOVED, LEAVING THE CLIP WITH THE CONTROL WIRE IN PLACE. THE SCOPE WAS REINSERTED AND THE CLIP FORCIBLY REMOVED FROM THE PATIENT. THE INTENDED PROCEDURE WAS COMPLETED USING CLIPS FROM A DIFFERENT MANUFACTURER. THE PROCEDURE WAS COMPLETED WITH A DELAY, CHARACTERIZED AS A FEW MINUTES. THERE WAS NO ADVERSE HEALTH EFFECT TO THE PATIENT THAT OCCURRED, ATTRIBUTED TO THE DELAY. THERE WAS NO PATIENT INJURY, ASSOCIATED WITH THE EVENT, REPORTED TO OLYMPUS. THIS IS REPORT #2 OF 3 DUE TO MULTIPLE EVENTS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14210201 ↗
Other Ligator, esophageal devices
- Speedband Superview Super 7 M00542250Boston Scientific Corporation · Ligator, esophageal6 deaths4,728 reports
- Saeed G22692Cook Incorporated · Ligator, esophageal295 reports
- Saeed G22553Cook Incorporated · Ligator, esophageal289 reports
- SINGLE USE LIGATING DEVICE HX-400U-30Olympus Medical Systems Corp. · Ligator, Esophageal1 recall239 reports
- Saeed G31917Cook Incorporated · Ligator, esophageal123 reports
- SINGLE USE LIGATING DEVICE HX-400U-30Olympus Medical Systems Corp. · Ligator, Esophageal97 reports
- Single Use Rotatable Clip Fixing Device HX-201UR-135Olympus Medical Systems Corp. · Ligator, esophageal1 recall1 death10 reports
- Single Use Rotatable Clip Fixing Device HX-201UR-135LOlympus Medical Systems Corp. · Ligator, esophageal1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page