EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE Q180AL (UDI 04953170202315)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA II COLONOVIDEOSCOPE (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EVIS EXERA II colonoscope is a flexible scope used to examine the colon. Many of the 82 reports describe scopes not being cleaned adequately, reprocessing problems, and images not displaying. The reports include 1 injury and no deaths. There is 1 open recall, from December 2022, after a scope was used in a veterinary procedure by mistake, which could allow microbial contamination.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
83
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 83 adverse event reports that list this exact device (DI 04953170202315), received between Sep 25, 2018 and Sep 3, 2026 — classified as 1 injury, 82 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (35 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 3 reportsJan 2024: 3 reportsFeb 2024: 4 reportsMar 2024: 3 reportsApr 2024: 4 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 3 reportsMay 2026: 4 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 21, 2022Class IIOpen, Classified

    A single CF-Q180AL colonovideoscope was utilized in a veterinary endoscopy procedure in advance of being assigned to a medical facility as a service loaner in error, potential for microbial contamination

    Z-1132-2023 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Clean Adequately35
  • Device Reprocessing Problem14
  • Microbial Contamination of Device10
  • No Display/Image4
  • Poor Quality Image4
  • Contamination /Decontamination Problem3
  • Communication or Transmission Problem2
  • Corroded2

Patient problems

  • No Clinical Signs, Symptoms or Conditions76
  • No Known Impact Or Consequence To Patient4
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1
  • Perforation1

Recent adverse event reports

The 25 most recent of 83 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 3, 2026Display or Visual Feedback Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50586995 ↗
  • MalfunctionJul 3, 2026No Display/Image; Failure to Clean Adequately

    THE CUSTOMER RETURNED THE COLONOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE EVALUATION, NO IMAGE WAS DISPLAYED, AND FOREIGN OBJECTS WERE OBSERVED ON THE CONNECTING TUBE, GRIP, RIGHT LEFT KNOB, AND FORCEPS ELEVATOR LEVER. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF THE COMPLAINT COULD NOT BE ESTABLISHED ALSO FOR THE NO IMAGE WAS DISPLAYED, CAUSE OF THE COMPLAINT WAS TRACED TO COMPONENT FAILURE. THERE WAS NO PATIENT INVOLVEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25703839 ↗
  • MalfunctionJun 26, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING TESTING, THE SERVICE TECHNICIAN IDENTIFIED THE DEVICE HAD EMERGENCE OF FOREIGN OBJECT FROM DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25641518 ↗
  • MalfunctionMay 20, 2026Device Reprocessing Problem; Corroded

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE COLONOVIDEOSCOPE HAD A CORRODED SHEET HOLDER UNIT AND IDENTIFICATION COVER, AS WELL AS A DIRTY SHIELD PLATE AND CHARGE-COUPLED DEVICE FLEXIBLE PRINTED CIRCUIT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25237034 ↗
  • MalfunctionMay 19, 2026Failure to Clean Adequately

    IT WAS OBSERVED DURING DEVICE EVALUATION THAT THE COLONOVIDEOSCOPE HAD A FOREIGN BODY (WHITE FOREIGN MATERIAL) IN THE WATER/AIR CHANNEL (TUBE) AND A FOREIGN OBJECT (WHITE FOREIGN MATERIAL) IN THE AIR/WATER VALVE SOCKET (TUBE). THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25222829 ↗
  • MalfunctionMay 15, 2026Erratic or Intermittent Display

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE COLONOVIDEOSCOPE EXHIBITED A FLICKERING IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25191032 ↗
  • MalfunctionMay 3, 2026Erratic or Intermittent Display

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25063069 ↗
  • MalfunctionApr 29, 2026No Display/Image

    IT WAS REPORTED THE COLONOVIDEOSCOPE HAD APPEARED NO IMAGE ON THE PROCESSOR. THE ISSUE WAS IDENTIFIED DURING DIAGNOSTIC COLONOSCOPY PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25032003 ↗
  • MalfunctionApr 8, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED DIRT ON OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821548 ↗
  • MalfunctionApr 8, 2026Obstruction of Flow

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD AN OBSTRUCTION (DEBRIS) IN THE DIFFUSER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24816188 ↗
  • MalfunctionMar 19, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD DIRT ON THE LIGHT GUIDE (LG) ROD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24640167 ↗
  • MalfunctionMar 16, 2026Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED CORROSION IN THE SUCTION COMPARTMENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24611634 ↗
  • MalfunctionJan 27, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL ON THE LIGHT GUIDE ROD LENS AND A DIRTY CHARGED COUPLED DEVICE COVER LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24186466 ↗
  • MalfunctionJan 26, 2026No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24175909 ↗
  • MalfunctionJan 7, 2026Device Reprocessing Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE HAD IMAGE FREEZE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23991919 ↗
  • MalfunctionDec 17, 2025Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED A BURN ON THE COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23830102 ↗
  • MalfunctionDec 16, 2025No Display/Image; No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE'S IMAGE WAS NOT DISPLAYED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23820912 ↗
  • MalfunctionNov 14, 2025Poor Quality Image; Failure to Clean Adequately; Poor Quality Image; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED A DISTURBED IMAGE WHEN CONNECTOR IS PUSHED. ALSO THE SCOPE CONNECTOR'S PLATE WAS DIRTY DUE TO WATER LEAKAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23560612 ↗
  • MalfunctionNov 13, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS ON AIR/WATER TUBE AND CYLINDER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23543583 ↗
  • MalfunctionOct 29, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED MALFUNCTION DUE TO DAMAGE TO THE CHARGED COUPLED DEVICE (CCD) UNIT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23413184 ↗
  • MalfunctionSep 4, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE EVALUATION, THE SUBJECT DEVICE'S LIGHT GUIDE ROD LENS HAD FOREIGN MATERIAL PRESENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22964821 ↗
  • MalfunctionAug 12, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE DISTAL END (LIGHT GUIDE ROD LENS) OF THE SCOPE HAD FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22787203 ↗
  • MalfunctionJul 15, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE NOZZLE AND CONNECTING TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22505586 ↗
  • MalfunctionJul 7, 2025Temperature Problem; Temperature Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SCOPE EXHIBITED THE PLASTIC PROBE TIP BURNED/BURNT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22412413 ↗
  • MalfunctionMay 27, 2025Poor Quality Image; Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED A BLURRY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22088896 ↗

Other Endoscope, accessories, narrow band spectrum devices