Skip to content

VISERA Pro VIDEO SYSTEM CENTER OLYMPUS OTV-S7ProRecalls & reported problems

Olympus Medical Systems Corp. · "OTV-S7Pro VISERA Pro VIDEO SYSTEM CENTER" · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus VISERA Pro video system center is an endoscope video processor. Of its 321 reports, all malfunctions, most describe a blank or unstable picture, loose or intermittent connections, and units that fail to power up. No deaths and no injuries are reported. No recalls name this device. Reports fell to 4 in the last 12 months from 13 in the previous 12.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
320
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 320 adverse event reports that list this exact device (DI 04953170228896), received between Jul 12, 2020 and Aug 31, 2026 — classified as 320 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (177 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 10 reportsDec 2023: 10 reportsJan 2024: 7 reportsFeb 2024: 8 reportsMar 2024: 5 reportsApr 2024: 6 reportsMay 2024: 5 reportsJun 2024: 4 reportsJul 2024: 4 reportsAug 2024: 6 reportsSep 2024: 13 reportsOct 2024: 11 reportsNov 2024: 5 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 4 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image177
  • Erratic or Intermittent Display32
  • Poor Quality Image31
  • Loose or Intermittent Connection21
  • Connection Problem19
  • Display or Visual Feedback Problem13
  • Failure to Power Up13
  • Electrical /Electronic Property Problem10

Patient problems

  • No Clinical Signs, Symptoms or Conditions309
  • No Consequences Or Impact To Patient8
  • No Patient Involvement4

Recent adverse event reports

The 25 most recent of 320 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 31, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50526527 ↗
  • MalfunctionJun 29, 2026No Display/Image; Loose or Intermittent Connection

    THE CUSTOMER REPORTED LOOSE CABLE WITH NO IMAGE INVOLVING AN OLYMPUS VIDEO SYSTEM CENTER. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25651737 ↗
  • MalfunctionApr 7, 2026Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM CENTER PRESENTED STREAKS IN THE IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24800946 ↗
  • MalfunctionAug 27, 2025Activation Failure; Activation Failure

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22907397 ↗
  • MalfunctionMar 28, 2025No Display/Image; Poor Quality Image; No Display/Image; Poor Quality Image

    IT WAS REPORTED THE VIDEO PROCESSOR EXPERIENCED COLOR BARS ON SLAVE MONITOR AND NO VIDEO IMAGE. THE EVENT HAPPENED DURING SET UP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21724506 ↗
  • MalfunctionMar 19, 2025Defective Component; Defective Component

    IT WAS REPORTED THAT THE VIDEO PROCESSOR HAD THE POWER BUTTON HAS TO BE HELD DOWN TO KEEP THE UNIT ON, OTHERWISE, IT TURNS OFF. THE ISSUE OCCURRED DURING DIAGNOSTIC CYSTOSCOPY PROCEDURE AND BEFORE SEDATION WHICH WAS COMPLETED USING A SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21640683 ↗
  • MalfunctionMar 19, 2025No Display/Image; Loose or Intermittent Connection; No Display/Image; Loose or Intermittent Connection

    IT WAS REPORTED THE SUBJECT DEVICE HAD OCCASIONAL POOR CONTACT BETWEEN THE ENDOSCOPE CABLE AND THE CONSOLE CONNECTOR AND NO IMAGE. THE ISSUE OCCURRED DURING THE PREPARATION OF THE DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21639906 ↗
  • MalfunctionMar 11, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21580263 ↗
  • MalfunctionJan 23, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THAT THE SUBJECT DEVICE IMAGE WOULD DISAPPEAR FOR NO REASON. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21218091 ↗
  • MalfunctionJan 8, 2025Complete Loss of Power; Complete Loss of Power

    IT WAS REPORTED THAT THE SUBJECT DEVICE DID NOT SWITCH ON. THE ISSUE OCCURRED DURING INSPECTION FOR USE FOR AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21101157 ↗
  • MalfunctionDec 20, 2024No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20995338 ↗
  • MalfunctionNov 19, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEO SYSTEM CENTER EXHIBITED NO IMAGE DUE TO A DAMAGED VIDEO CONNECTOR SOCKET. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20727276 ↗
  • MalfunctionNov 14, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED, DURING THE DEVICE INSPECTION. THAT THE VIDEO SYSTEM CENTER EXHIBITED NO IMAGE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20685349 ↗
  • MalfunctionNov 13, 2024No Display/Image; No Display/Image

    IT WAS REPORTED, THE VIDEO SYSTEM CENTER HAD BROKEN WIRES NEAR THE CAMERA. AND WHEN THE CABLE WAS MOVED, THE VIDEO WAS LOST. IT IS UNKNOWN, WHEN THE ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20674974 ↗
  • MalfunctionNov 6, 2024Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED, THE VIDEO SYSTEM CENTER VIDEO CONNECTOR WAS LOOSE AND COULD NOT LOCK THE CAMERA TIGHTLY. THE ISSUE OCCURRED DURING PREPARATION FOR USE IN AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM. THE FACILITY FINISHED THE PROCEDURE WITH THE REPLACEMENT DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20623449 ↗
  • MalfunctionNov 1, 2024No Display/Image; Loose or Intermittent Connection; No Display/Image; Loose or Intermittent Connection

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM CENTER EXHIBITED NO IMAGE WHEN SHAKING THE CONNECTOR DUE TO LOOSE LOCK OF VIDEO CONNECTOR SOCKET. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20589015 ↗
  • MalfunctionOct 30, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED, THE VIDEO SYSTEM CENTER IMAGE OFTEN FLICKERS. THE CUSTOMER REPLACED THE SCOPE, BUT THE ISSUE PERSISTS. THE ISSUE OCCURRED, DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20569512 ↗
  • MalfunctionOct 29, 2024No Display/Image; No Display/Image

    IT WAS REPORTED, DURING A THERAPEUTIC LAPAROSCOPIC SURGERY, THE VIDEO SYSTEM CENTER HAD NO IMAGE AND SCREEN TURNED WHITE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20559866 ↗
  • MalfunctionOct 29, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED DURING MAINTENANCE, THAT THE VIDEO PROCESSOR EXHIBITED NO IMAGE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20554183 ↗
  • MalfunctionOct 25, 2024No Display/Image; No Display/Image

    IT WAS REPORTED, THE VIDEO SYSTEM INTERMITTENTLY HAS NO IMAGE. THE ISSUE OCCURRED, DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20532803 ↗
  • MalfunctionOct 22, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THE VIDEO SYSTEM CENTER IMAGE WAS TURNING GREEN EVERY FEW SECONDS. IT IS UNKNOWN WHEN THE ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20509500 ↗
  • MalfunctionOct 22, 2024Complete Loss of Power; Complete Loss of Power

    IT WAS REPORTED THE VIDEO SYSTEM CENTER WAS NOT POWERING UP. UPON TROUBLESHOOTING, A FUSE WAS FOUND TO BE BLOWN, BUT AFTER REPLACEMENT AND TURNING ON THE POWER, THE FUSE BLEW AGAIN. IT IS UNKNOWN WHEN THE ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20509043 ↗
  • MalfunctionOct 16, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THE VIDEO SYSTEM CENTER DISPLAYED NO IMAGE WHEN SWITCHED ON. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20467745 ↗
  • MalfunctionOct 15, 2024Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THE VIDEO SYSTEM CENTER HAD POOR CONTACT OF THE VIDEO CABLE SOCKET. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20456209 ↗
  • MalfunctionOct 14, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEO SYSTEM CENTER EXHIBITED NO IMAGE DUE TO THE LOOSE VIDEO CONNECTOR SOCKET. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20441414 ↗

Other Endoscope, accessories, narrow band spectrum devices

All Endoscope, accessories, narrow band spectrum devices →

Sources and updates