VISERA Pro XENON LIGHT SOURCE OLYMPUS CLV-S40Pro (UDI 04953170228933)Recalls & reported problems
Olympus Medical Systems Corp. · CLV-S40PRO CLV-S40 PRO 300W VISERA PRO X · Class 2
All models of VISERA Pro XENON LIGHT SOURCE (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus VISERA Pro xenon light source provides light for endoscopes. Many reports describe the unit failing to power up or give output, and some describe flashing lights or a blank display. The 157 reports include no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 158
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 158 adverse event reports that list this exact device (DI 04953170228933), received between Nov 20, 2020 and Sep 23, 2026 — classified as 158 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (42 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem42
- No Device Output17
- Failure to Power Up16
- Display or Visual Feedback Problem11
- No Display/Image9
- Optical Problem8
- Power Problem8
- Erratic or Intermittent Display6
Patient problems
- No Clinical Signs, Symptoms or Conditions154
- No Consequences Or Impact To Patient5
Recent adverse event reports
The 25 most recent of 158 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Activation Failure
The customer reported Main lamp does not ignite during inspection for use involving an Olympus xenon light source. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50872468 ↗MalfunctionAug 2, 2025Optical Problem; Optical Problem
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22693782 ↗MalfunctionJun 6, 2025Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING MAINTENANCE, THE SUBJECT DEVICE EXHIBITED THAT THE MAIN LAMP WAS BLINKING AND MADE TISSUE INDISTINGUISHABLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22158336 ↗MalfunctionDec 11, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20905902 ↗MalfunctionDec 9, 2024Power Problem; Power Problem
IT WAS REPORTED THE SUBJECT DEVICE EXHIBITED POWER SUPPLY FAILURE. THE ISSUE OCCURRED DURING SET UP, BEFORE PATIENT IN ROOM. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20878758 ↗MalfunctionNov 26, 2024Flare or Flash; Activation Failure; Flare or Flash; Activation Failure
IT WAS REPORTED, THAT THE SUBJECT DEVICE HAD PANEL FLASHES. THE ISSUE WAS FOUND, DURING A SET UP/INSPECTION FOR USE, BEFORE PATIENT IN ROOM. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20789768 ↗MalfunctionNov 24, 2024No Device Output; No Device Output
IT WAS REPORTED, ALL LEDS OF THE SUBJECT DEVICE FLASHED AT IRREGULAR INTERVALS, DURING AN UNSPECIFIED PROCEDURE. THE PROCEDURE WAS PROLONGED LESS THAN 30 MINUTES, AS A RESULT OF THIS EVENT. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20767941 ↗MalfunctionSep 26, 2024No Visual Prompts/Feedback; No Visual Prompts/Feedback
IT WAS REPORTED THE XENON LIGHT SOURCE TURNED ON, BUT THE LAMP DID NOT TURN ON. THE ISSUE OCCURRED BEFORE AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20309167 ↗MalfunctionApr 24, 2024No Device Output; No Device Output
IT WAS REPORTED THE LIGHT SOURCE, FRONT PANEL WAS BLINKING/FLASHING AND CONTINUING TURNING O. HERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19171766 ↗MalfunctionApr 12, 2024Poor Quality Image; Noise, Audible; Poor Quality Image; Noise, Audible
IT WAS REPORTED, THE LIGHT SOURCE HAD IMAGE NOISE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19098487 ↗MalfunctionApr 11, 2024No Device Output; No Device Output
IT WAS REPORTED THE XENON LIGHT SOURCE FRONT PANEL WAS FLICKERING. IT WAS NOT CONFIRMED WHEN THIS EVENT TOOK PLACE. ADDITIONAL DETAILS RELATING TO THE EVENT HAVE BEEN REQUESTED, BUT NO RESPONSE HAS BEEN RECEIVED AT THIS TIME. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19088353 ↗MalfunctionMar 29, 2024Burst Container or Vessel; Burst Container or Vessel
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE LIGHT SOURCE HAD A BURNT INLET FUSE BOX. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19008663 ↗MalfunctionMar 11, 2024Output Problem; Output Problem
IT WAS REPORTED, THE XENON LIGHT SOURCE DOES NOT WORK. NO ADDITIONAL INFORMATION WAS PROVIDED. IT IS UNKNOWN WHEN THE ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18882868 ↗MalfunctionFeb 23, 2024No Display/Image; No Display/Image
IT WAS REPORTED, THE XENON LIGHT SOURCE HAD AN IMAGE ISSUE ON THE TOWER. POWERED ON UNIT AND SAW THAT THE "AUTO" LIGHT WAS NOT ON. THERE WAS NO REPORT OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18773399 ↗MalfunctionFeb 14, 2024Break; Break
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE LIGHT SOURCE EXHIBITED THE INLET FUSE BOX BEING BURNT OUT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18707422 ↗MalfunctionJan 10, 2024No Device Output; No Device Output
THE CUSTOMER REPORTED TO OLYMPUS, THE PRO XENON LIGHT SOURCE EMERGENCY LIGHTS ARE TURNED ON FREQUENTLY. IT IS UNKNOWN WHEN THE ISSUE WAS FOUND DURING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18486342 ↗MalfunctionJan 3, 2024No Display/Image; No Display/Image
THE CUSTOMER REPORTED TO OLYMPUS, THE XENON LIGHT SOURCE HAD A BROKEN RECORDING WITH A DEFECTIVE LIGHT CABLE. IT WAS UNSPECIFIED WHEN THE ISSUE WAS FOUND. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18440927 ↗MalfunctionDec 22, 2023Overheating of Device; No Visual Prompts/Feedback; Overheating of Device; No Visual Prompts/Feedback
THE CUSTOMER REPORTED TO OLYMPUS, THAT THE LIGHT SOURCE OVERHEATED AND TURNED OFF. THE EMERGENCY LIGHT DID NOT COME ON. THE ISSUE WAS FOUND DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18387135 ↗MalfunctionDec 21, 2023No Device Output; No Device Output
THE CUSTOMER REPORTED TO OLYMPUS, THE LIGHT SOURCE PANELS MAY LIGHT UP WHEN THE POWER IS TURNED ON. THE ISSUE WAS FOUND DURING INSPECTION FOR USE, FOR AN UNSPECIFIED DIAGNOSTIC PROCEDURE. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18377785 ↗MalfunctionDec 18, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
THE CUSTOMER REPORTED TO OLYMPUS THE VISERA PRO XENON LIGHT SOURCE, WHEN TURNING ON THE POWER TO THE DEVICE, WHEN THE LIGHT IS TURNED ON, THE LAMP IS ABOUT 300 HOURS, BUT THE EMERGENCY LIGHT IS ON. IT WAS RESTORED BY RESTARTING. THE ISSUE OCCURRED DURING THE PREPARATION FOR USE. THE PROCEDURE WAS UNKNOWN AND IT WAS COMPLETED USING THE SAME SET OF EQUIPMENT. THERE ARE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18344432 ↗MalfunctionNov 15, 2023Output Problem; Output Problem
THE CUSTOMER REPORTED TO OLYMPUS THAT THE VISERA XENON LIGHT SOURCE MAIN LAMP AND EMERGENCY LAMP DID NOT LIGHT. THE ISSUE WAS OBSERVED DURING PREPARATION FOR USE ON AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18139028 ↗MalfunctionNov 7, 2023No Display/Image; Defective Component; No Display/Image; Defective Component
THE CUSTOMER REPORTED TO OLYMPUS, THERE WAS NO IMAGE WHEN THE ENDOSCOPE WAS CONNECTED AND THE LIGHT SOURCE WAS TURNED ON. IT WAS ALSO REPORTED THA THE FRONT PANEL COULD NOT BE OPERATED. THE ISSUE WAS FOUND DURING PREPARATION FOR USE OF AN UNSPECIFIED PROCEDURE. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE WITH NO DELAYS OR PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18089241 ↗MalfunctionOct 12, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER REPORTED TO OLYMPUS THAT THE VISERA PRO XENON LIGHT SOURCE EMERGENCY LIGHTS TURNED ON. THERE WAS NO REPORT OF PATIENT HARM OR USER INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17918199 ↗MalfunctionOct 2, 2023No Display/Image; No Display/Image
THE CUSTOMER REPORTED TO OLYMPUS, THE VISERA PRO XENON LIGHT SOURCE AND VISERA PRO VIDEO SYSTEM CENTER GENERATED SMOKE JUST PRIOR THE END OF THE EXAMINATION, AND IT WAS FURTHER NOTED THAT THE ENDOSCOPIC IMAGE HAD NO OUTPUT, RESULTING IN A BLACKOUT. THE ISSUE WAS FOUND DURING THE EXAMINATION AND WAS COMPLETED USING THE SAME SET OF EQUIPMENT. THERE WAS NO DELAY IN TREATMENT AND THERE WERE NO REPORTS OF PATIENT HARM. RELATED PATIENT IDENTIFIER: (B)(6) VISERA PRO VIDEO SYSTEM CENTER FOR ADDITIONAL DEVICE REPORTED FOR SAME EVENT (B)(3).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17858267 ↗MalfunctionSep 25, 2023Intermittent Loss of Power; Intermittent Loss of Power
THE CUSTOMER REPORTED TO OLYMPUS THAT THE VISERA PRO XENON LIGHT SOURCE HAD RESTARTED DURING SURGERY. THE ISSUE OCCURRED DURING THE PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM. RELATED PATIENT IDENTIFIERS ARE AS FOLLOWS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17809257 ↗
Other Endoscope, accessories, narrow band spectrum devices
- EVIS EXERA III VIDEO SYSTEM CENTER OLYMPUS CV-190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall7,688 reports
- HD CAMERA HEAD OLYMPUS CH-S190-08-LBOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum3,793 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190LOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum2 recalls2,828 reports
- EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180VOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum6 deaths2,362 reports
- VISERA ELITE VIDEO SYSTEM CENTER OLYMPUS OTV-S190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1,563 reports
- HD AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S190-XZ-EOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1,424 reports
- VIDEO SYSTEM CENTER Type A OLYMPUS CV-170Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum841 reports
- EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180VOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall3 deaths653 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page