GASTROINTESTINAL VIDEOSCOPE GIF-Q150 (UDI 04953170229435)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of GASTROINTESTINAL VIDEOSCOPE (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus gastrointestinal videoscope is an endoscope used to look inside the digestive tract. The 285 reports often describe scopes that were not cleaned adequately during reprocessing, poor image quality, and scopes that showed no image. The reports include no deaths and 1 injury. No recalls name this device, and reports rose to 60 in the last year from 18 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
283
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 283 adverse event reports that list this exact device (DI 04953170229435), received between Aug 8, 2019 and Sep 10, 2026 — classified as 1 injury, 282 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (130 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 7 reportsDec 2023: 2 reportsJan 2024: 4 reportsFeb 2024: 6 reportsMar 2024: 4 reportsApr 2024: 5 reportsMay 2024: 8 reportsJun 2024: 9 reportsJul 2024: 9 reportsAug 2024: 7 reportsSep 2024: 5 reportsOct 2024: 12 reportsNov 2024: 3 reportsDec 2024: 6 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 3 reportsDec 2025: 2 reportsJan 2026: 11 reportsFeb 2026: 7 reportsMar 2026: 13 reportsApr 2026: 14 reportsMay 2026: 2 reportsJun 2026: 2 reportsJul 2026: 1 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately130
  • Poor Quality Image85
  • Device Reprocessing Problem24
  • Contamination15
  • No Display/Image11
  • Contamination /Decontamination Problem6
  • Corroded6
  • Device Contamination with Chemical or Other Material5

Patient problems

  • No Clinical Signs, Symptoms or Conditions278
  • Insufficient Information2
  • Injury1
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 283 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2026Thermal Decomposition of Device

    The customer returned the Gastrointestinal Videoscope for a separate issue. During the initial evaluation of the device it was observed, a burn on the terminal cover. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50711746 ↗
  • MalfunctionAug 26, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456196 ↗
  • MalfunctionAug 12, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50223968 ↗
  • MalfunctionJul 10, 2026Poor Quality Image

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES A BLURRY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25783473 ↗
  • MalfunctionJun 26, 2026Material Discolored

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE, WHICH IS AN OLYMPUS ASSET, WITH NO REPORTED COMPLAINT. HOWEVER, DURING THE EVALUATION OF THE DEVICE, A PROTECTOR OF THE UNIVERSAL CORD ON THE SCOPE CONNECTOR (S-CONNECTOR) SIDE WAS STICKY. THE SERVICE REPAIR TECHNICIAN ASSESSED THE CONDITION BUT COULD NOT DETERMINE THE CAUSE OF THE FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25638306 ↗
  • MalfunctionJun 3, 2026Poor Quality Image

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE CENTER IDENTIFIED THAT THE DEVICE HAD A BLURRY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25376618 ↗
  • MalfunctionMay 31, 2026Overheating of Device

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING EVALUATION OF THE DEVICE THE DEVICE WAS FOUND TO HAVE BURNT C-COVER. THE SERVICE REPAIR TECHNICIAN INDICATED IT IS POSSIBLE THAT HIGH-ENERGY INSTRUMENTS (SUCH AS HIGH-FREQUENCY DEVICES) WERE USED WITHOUT MAINTAINING SUFFICIENT DISTANCE FROM THE TIP CAUSING BURNT C-COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25342410 ↗
  • MalfunctionMay 31, 2026Poor Quality Image

    THE CUSTOMER REPORTED ARTIFACTS IN THE IMAGE DURING REPROCESSING INVOLVING AN OLYMPUS GASTROINTESTINAL VIDEOSCOPE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25341129 ↗
  • MalfunctionApr 22, 2026Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN THE LIGHT GUIDE COVER LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24956265 ↗
  • MalfunctionApr 15, 2026Material Discolored; No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE'S INSIDE OF THE LIGHT GUIDE LENS HAD DISCOLORATION AND THE IMAGE WAS NOT DISPLAYED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24885832 ↗
  • MalfunctionApr 14, 2026Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE GASTROINTESTINAL VIDEOSCOPE HAD EXCESSIVE DIRT ON THE DISTAL TIP COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24874281 ↗
  • MalfunctionApr 13, 2026No Display/Image; Melted

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A BURN MARK ON THE TERMINAL COVER AND A MISSING ENDOSCOPIC IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24855895 ↗
  • MalfunctionApr 11, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE'S AIR WATER-CYLINDER AND NOZZLE HAD CORROSION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24852268 ↗
  • MalfunctionApr 10, 2026Corroded

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE OBJECTIVE LENS HAD A CORROSION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24840364 ↗
  • MalfunctionApr 9, 2026Communication or Transmission Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A SCOPE COMMUNICATION B30 ERROR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24833496 ↗
  • MalfunctionApr 8, 2026Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL CAME OUT OF THE SUCTION CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821077 ↗
  • MalfunctionApr 8, 2026Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE PRESENTED BLURRY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24817540 ↗
  • MalfunctionApr 7, 2026Melted; Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD BLURRY IMAGE AND BURN ON LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24798416 ↗
  • MalfunctionApr 6, 2026No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION THE GASTROINTESTINAL VIDEOSCOPE HAD AN INSTANCE OF NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24795834 ↗
  • MalfunctionApr 6, 2026Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED CORROSION IN THE DIFFUSER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792842 ↗
  • MalfunctionApr 6, 2026Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD CORRODED SUCTION CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24788020 ↗
  • MalfunctionApr 3, 2026Material Protrusion/Extrusion

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED INSERTION TUBE HAD PROTRUSION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24782352 ↗
  • MalfunctionMar 31, 2026Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED NOZZLE HAD CORROSION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24748539 ↗
  • MalfunctionMar 31, 2026Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD CORROSION ON INSERTION TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742827 ↗
  • MalfunctionMar 31, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED STICKY GRIP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24739213 ↗

Other Endoscope, accessories, narrow band spectrum devices

All Endoscope, accessories, narrow band spectrum devices →