EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE 2TH180 (UDI 04953170275463)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04953170275463), received between Jun 6, 2018 and Jan 29, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Microbial Contamination of Device” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Microbial Contamination of Device4
- Failure to Clean Adequately3
- Device Reprocessing Problem1
- Inability to Irrigate1
- Melted1
- Thermal Decomposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Known Impact Or Consequence To Patient2
- Insufficient Information1
- No Patient Involvement1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 29, 2026Device Reprocessing Problem
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24209719 ↗MalfunctionJan 19, 2026Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED DIRT ON THE LIGHT GUIDE (LG) LENS, WHITE FOREIGN MATERIAL INSIDE THE AIR/WATER (A/W) CYLINDER, AND WHITE FOREIGN MATERIAL INSIDE THE AIR/WATER (A/W) CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24108459 ↗MalfunctionJul 31, 2025Melted; Melted
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A BURNED PLASTIC DISTAL END COVER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: Insufficient Information
MAUDE report 22683212 ↗MalfunctionSep 20, 2024Thermal Decomposition of Device; Failure to Clean Adequately; Thermal Decomposition of Device; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A BURN ON THE PLASTIC DISTAL END COVER AND WHITE FOREIGN OBJECTS ON THE AIR/WATER CYLINDER AND AIR/WATER TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20268045 ↗MalfunctionDec 22, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS, THE THREADED CONNECTION OF THE FLUSHING CHANNEL ON THE GASTROINTESTINAL VIDEOSCOPE HAS BROKEN OFF. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, FOREIGN MATERIAL WAS FOUND IN THE JET TUBE. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18388084 ↗MalfunctionAug 24, 2023Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS, THE EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE TESTED POSITIVE FOR MICROBIAL CONTAMINATION. THE SCOPE WAS SAMPLED ONCE. DURING THE SAMPLING, THE SCOPE TESTED POSITIVE FOR 1 COLONY FORMING UNITS (CFUS) OF STAPHYLOCOCCUS HOMINIS AND 1 CFUS OF CARYNEBACTERIUM MUCIFACIENS. THE ISSUE WAS FOUND AT REPROCESSING DURING A ROUTINE CULTURE OF THE SCOPE. THERE WAS NO PATIENT/USER HARM OR INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17616534 ↗MalfunctionApr 6, 2022Microbial Contamination of Device; Microbial Contamination of Device
AS REPORTED, THE ENDOSCOPE FAILED THE HYGIENE TEST DESPITE MULTIPLE CLEANINGS (MANUAL AND MECHANICAL). THE CUSTOMER THEN SENT THE DEVICE TO RRC (REGIONAL REPAIR CENTER) OLYMPUS HAMBURG FOR FURTHER INVESTIGATION. THE CUSTOMER DID NOT REPORT ANY CONTAMINATION OR ANY PATIENT INJURY OR PATIENT INFECTION TO WHICH THIS MEDICAL DEVICE COULD HAVE BEEN A CONTRIBUTORY CAUSE. NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14032022 ↗MalfunctionJun 4, 2020Inability to Irrigate; Inability to Irrigate
IT WAS REPORTED TO OLYMPUS THAT THERE WAS A SERVICE PORTAL CHANNEL MALFUNCTION AND THERE WAS A BIOPSY VALVE NEEDLE STUCK IN RED CHANNEL BIOPSY. IT WAS REPORTED THAT THE REPORTED EVENT OCCURRED DURING REPROCESSING AND NOT DURING A PROCEDURE IMMEDIATELY PRIOR TO REPROCESSING.
Patient problems: No Patient Involvement
MAUDE report 10121747 ↗MalfunctionJul 25, 2019Microbial Contamination of Device; Microbial Contamination of Device
OLYMPUS WAS INFORMED THAT AS A RESULT OF MICROBIOLOGICAL TESTING BY THE USER FACILITY, THE SUBJECT DEVICE TESTED POSITIVE FOR THE FOLLOWING BACTERIA. THE INSTRUMENT CHANNEL A: MICROCOCCUS SP. (2CFU/20ML). THE AUXILIARY WATER CHANNEL: PSEUDOMONAS SP. (5CFU/20ML). THE INSTRUMENT CHANNEL B: NO MICROBIAL GROWTH. THE DISTAL END: NO MICROBIAL GROWTH. THE SUBJECT DEVICE HAD BEEN DISINFECTED USING NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR (BELIMED). THERE WAS NO REPORT OF PATIENT INFECTION ASSOCIATED WITH THIS REPORT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8825483 ↗MalfunctionJun 6, 2018Microbial Contamination of Device; Microbial Contamination of Device
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT THE SUBJECT DEVICE TESTED POSITIVE FOR FOLLOWING BACTERIA DURING A SURVEILLANCE CULTURING TEST AT THE FACILITY ON (B)(6) 2018. BIOPSY CHANNEL 1: STAPHYLOCOCCUS AUREUS (990CFU/ML), BIOPSY CHANNEL 2: STAPHYLOCOCCUS AUREUS (1CFU/ML), AUXILIARY CHANNEL: STAPHYLOCOCCUS AUREUS (990CFU/ML). IT WAS ALSO REPORTED THAT THE SUBJECT DEVICE TESTED POSITIVE FOR FOLLOWING BACTERIA (18CFU/ENDOSCOPE) DURING ADDITIONAL MICROBIOLOGICAL CULTURING TEST AT THE FACILITY ON (B)(6) 2018. BIOPSY CHANNEL 1 :COAGULASE NEGATIVE STAPHYLOCOCCUS AUREUS (1CFU/ML), AIR/WATER CHANNEL: STAPHYLOCOCCUS AUREUS (990CFU/ML), AUXILIARY CHANNEL: MICROCOCCUS (1CFU/ML). OTHER DETAILED INFORMATION SUCH AS REPROCESSING METHOD WAS NOT PROVIDED FROM THE FACILITY BUT THERE WAS NO REPORT OF PATIENT INFECTION ASSOCIATED WITH THE EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7576653 ↗
Other Endoscope, accessories, narrow band spectrum devices
- EVIS EXERA III VIDEO SYSTEM CENTER OLYMPUS CV-190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall7,688 reports
- HD CAMERA HEAD OLYMPUS CH-S190-08-LBOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum3,793 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190LOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum2 recalls2,828 reports
- EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180VOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum6 deaths2,362 reports
- VISERA ELITE VIDEO SYSTEM CENTER OLYMPUS OTV-S190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1,563 reports
- HD AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S190-XZ-EOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1,424 reports
- VIDEO SYSTEM CENTER Type A OLYMPUS CV-170Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum841 reports
- EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180VOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall3 deaths653 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page