VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2 (UDI 04953170287091)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of VISERA CYSTO-NEPHRO VIDEOSCOPE (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04953170287091), received between Feb 8, 2019 and Dec 13, 2024 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Device Reprocessing Problem” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 4, 2023Class IIOpen, Classified
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Z-0691-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Device Reprocessing Problem2
- Difficult to Remove1
- Mechanical Problem1
- Misassembled1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Hypersensitivity/Allergic reaction1
- Injury1
- Local Reaction1
- Skin Discoloration1
- Swelling1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 13, 2024Misassembled; Misassembled
IT WAS REPORTED THAT THE SUBJECT DEVICE HAD WIRE COMING OUT THE TIP. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20925608 ↗MalfunctionOct 25, 2024Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED THE DISTAL END OF THE CHANNEL ON THE VIDEOSCOPE WAS STRIPPED DOWN TO METAL, AND IT IS SUSPECTED THAT THE ETHYLENE OXIDE (GAS) VALVE CAP WAS NOT PLACED ON THE UNIT DURING REPROCESSING. THE ISSUE OCCURRED DURING REPROCESSING. NO PATIENT INFECTIONS OR INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20536204 ↗MalfunctionJun 27, 2024Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED THE CYSTO-NEPHRO VIDEOSCOPE EXPERIENCED THE SCOPE WAS BLOWN DUE TO PRESSURE CAP NOT BEING ATTACHED. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19623776 ↗InjuryDec 20, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE SERVICE CENTER WAS INFORMED BY THE USER FACILITY¿S PERFORMING DOCTOR THAT THE PATIENT EXPERIENCES AN ALLERGIC REACTION AFTER UNDERGOING REPEAT CYSTOSCOPY PROCEDURES. DUE TO A PRE-EXISTING CONDITION, THE PATIENT MUST UNDERGO A CYSTOSCOPY EVERY THREE MONTHS. THE PATIENT EXPERIENCES SWELLING OF THE PENIS AND PURPLE DISCOLORATION. THE PATIENT HAS NO PRE-EXISTING ALLERGIES. THE PATIENT WAS GIVEN TOPICAL CREAM TO TREAT THE REACTION. THE DOCTOR REPORTED THAT THE PATIENT¿S CONDITION LAST APPROXIMATELY TWO OR THREE WEEKS. THE DOCTOR REPORTED THAT ONLY KY LUBRICANT IS USED. THERE ARE NO OTHER LUBRICANTS, CREAMS, OR SOAPS USED DURING THE PROCEDURE. THIS PATIENT IS AN ISOLATED CASE, NO OTHER PATIENTS HAVE BEEN NOTED WITH THIS REACTION. IN ADDITION, THE USER FACILITY REPORTED THAT THEY ARE NOT 100% SURE THAT THIS IS RELATED TO THE SCOPE.
Patient problems: Local Reaction
MAUDE report 9507500 ↗InjuryDec 20, 2019Adverse Event Without Identified Device or Use Problem
THE SERVICE CENTER WAS INFORMED BY THE USER FACILITY¿S PERFORMING DOCTOR THAT THE PATIENT EXPERIENCES AN ALLERGIC REACTION AFTER UNDERGOING REPEAT CYSTOSCOPY PROCEDURES. DUE TO A PRE-EXISTING CONDITION, THE PATIENT MUST UNDERGO A CYSTOSCOPY EVERY THREE MONTHS. THE PATIENT EXPERIENCES SWELLING OF THE PENIS AND PURPLE DISCOLORATION. THE PATIENT HAS NO PRE-EXISTING ALLERGIES. THE PATIENT WAS GIVEN TOPICAL CREAM TO TREAT THE REACTION. THE DOCTOR REPORTED THAT THE PATIENT¿S CONDITION LAST APPROXIMATELY TWO OR THREE WEEKS. THE DOCTOR REPORTED THAT ONLY KY LUBRICANT IS USED. THERE ARE NO OTHER LUBRICANTS, CREAMS, OR SOAPS USED DURING THE PROCEDURE. THIS PATIENT IS AN ISOLATED CASE, NO OTHER PATIENTS HAVE BEEN NOTED WITH THIS REACTION. THE DOCTOR DID NOT KNOW THE SPECIFIC SERIAL NUMBER FOR THE REFERENCED SCOPE; THEREFORE, IT IS UNKNOWN IF THE SCOPE WAS RETURNED TO OLYMPUS FOR EVALUATION/SERVICE AND A REVIEW OF THE INSTRUMENT¿S HISTORY COULD NOT BE PERFORMED. THE SERVICE CENTER DID FOLLOW UP WITH THE USER FACILITY TO OBTAIN THE FOLLOWING ADDITIONAL INFORMATION: THE DISINFECTANT/SOLUTION USED TO REPROCESS THE SCOPE AND THE PATIENT¿S PROCEDURE DATE; HOWEVER, THIS INFORMATION WAS NOT PROVIDED.
Patient problems: Hypersensitivity/Allergic reaction; Skin Discoloration; Swelling
MAUDE report 9507369 ↗InjuryFeb 8, 2019Mechanical Problem; Difficult to Remove; Mechanical Problem; Difficult to Remove
OLYMPUS WAS INFORMED THAT DURING A PROCEDURE, A SCOPE BECAME STUCK INSIDE THE PATIENT¿S URETHRA. THE SCOPE¿S INSERTION TUBE WAS IN A 360 DEGREE LOOP AND THE UP LEVER WORKED BUT THE DOWN LEVER DID NOT MOVE THE BENDING SECTION TIP OF THE SCOPE. THE PHYSICIAN HAD DECIDED TO CUT THE SCOPE¿S INSERTION TUBE AND HAD TO RUSH THE PATIENT INTO SURGERY.
Patient problems: Injury
MAUDE report 8324293 ↗
Other Endoscope, accessories, narrow band spectrum devices
- EVIS EXERA III VIDEO SYSTEM CENTER OLYMPUS CV-190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall7,688 reports
- HD CAMERA HEAD OLYMPUS CH-S190-08-LBOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum3,793 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190LOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum2 recalls2,828 reports
- EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180VOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum6 deaths2,362 reports
- VISERA ELITE VIDEO SYSTEM CENTER OLYMPUS OTV-S190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1,563 reports
- HD AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S190-XZ-EOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1,424 reports
- VIDEO SYSTEM CENTER Type A OLYMPUS CV-170Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum841 reports
- EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180VOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall3 deaths653 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page