EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-PH190I (REF N3802840)Recalls & reported problems

Olympus Medical Systems Corp. · "PCF-PH190I ULTRA SLIM EE III HD COLONOSC" · Class 2

All models of EVIS EXERA lll COLONOVIDEOSCOPE (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EVIS EXERA III colonoscope is a flexible camera used to examine the colon. Most of the 47 reports describe cleaning and reprocessing problems, such as microbial contamination (20), failure to clean adequately (15), and cracked or torn parts. There are no deaths and no injuries. Reports rose from 6 in the prior 12 months to 21 in the last 12 months. One open recall from September 2023 concerns an unvalidated air-drying process that left water inside repaired scopes, with a risk of contamination and infection.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
47
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 47 adverse event reports that list this exact device (DI 04953170305252), received between Aug 22, 2019 and Aug 18, 2026 — classified as 47 malfunctions.

The most frequently reported specific device problem is “Microbial Contamination of Device” (20 reports).

1 FDA recall record name this device's identifier, 1 still open.

7 other recalls cover devices cleared under the same FDA submission (K131780); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 8 reportsMay 2026: 4 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0341-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Microbial Contamination of Device20
  • Failure to Clean Adequately15
  • Device Reprocessing Problem9
  • Material Split, Cut or Torn2
  • Communication or Transmission Problem1
  • Contamination /Decontamination Problem1
  • Crack1
  • Dent in Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions43
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 25 most recent of 47 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 18, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50307897 ↗
  • MalfunctionAug 4, 2026Device Reprocessing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50134253 ↗
  • MalfunctionAug 3, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50121963 ↗
  • MalfunctionJun 27, 2026Communication or Transmission Problem

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR A FAILED LEAK TEST. DURING THE DEVICE ANALYSIS, THE SERVICE CENTER TECHNICIAN INDICATES THAT AN INCOMPATIBLE SCOPE ERROR(E315) WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25646448 ↗
  • MalfunctionMay 27, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE, WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, FOREIGN OBJECTS IN SCOPE CONNECTOR CASE UNIT, C-COVER AND LIGHT GUIDE LENS. THE REGIONAL REPAIR CENTER INDICATED THAT THE MOST LIKELY CAUSE OF THE FOREIGN OBJECTS IN SCOPE CONNECTOR CASE UNIT, C-COVER AND LIGHT GUIDE LENS WAS CAUSE NOT DETERMINED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25306079 ↗
  • MalfunctionMay 21, 2026Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE CONNECTING TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25241485 ↗
  • MalfunctionMay 8, 2026Device Reprocessing Problem

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE USED ON PATIENT WHILE WAITING FOR A POSITIVE CULTURE TEST RESULT. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25122136 ↗
  • MalfunctionMay 8, 2026Microbial Contamination of Device

    NO NEW ADDITIONAL INFORMATION WAS PROVIDED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25122083 ↗
  • MalfunctionApr 30, 2026Microbial Contamination of Device; Failure to Clean Adequately

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25049557 ↗
  • MalfunctionApr 30, 2026Microbial Contamination of Device; Failure to Clean Adequately

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25047381 ↗
  • MalfunctionApr 29, 2026Microbial Contamination of Device

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25038517 ↗
  • MalfunctionApr 22, 2026Microbial Contamination of Device

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24953974 ↗
  • MalfunctionApr 22, 2026Microbial Contamination of Device

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24953897 ↗
  • MalfunctionApr 22, 2026Microbial Contamination of Device; Failure to Clean Adequately

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24953754 ↗
  • MalfunctionApr 10, 2026Microbial Contamination of Device

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24838535 ↗
  • MalfunctionApr 7, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE'S IMAGE GUIDE PROTECTOR AND CONNECTING TUBE HAD FOREIGN OBJECTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24798757 ↗
  • MalfunctionFeb 10, 2026Optical Problem

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE HAD A MISSING LIGHT BEAM. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24309685 ↗
  • MalfunctionFeb 3, 2026Residue After Decontamination

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE SUCTION CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24248783 ↗
  • MalfunctionJan 21, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD STICKINESS AT THE CONNECTING TUBE AND FOREIGN MATERIAL PRESENT AT THE AIR/WATER (A/W) CYLINDER AND TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24137630 ↗
  • MalfunctionJan 20, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD DIRT ON LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24116024 ↗
  • MalfunctionNov 13, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN AIR/WATER TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23542781 ↗
  • MalfunctionAug 5, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX VIDEO SCOPE EXHIBITED A FOREIGN MATERIAL IN THE AIR WATER TUBE AND CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22715184 ↗
  • MalfunctionJul 7, 2025Device Reprocessing Problem; Crack; Poor Quality Image; Structural Problem; Detachment of Device or Device Component; Material Deformation; Mechanical Jam; Material Split, Cut or Torn; Physical Resistance/Sticking; Device Reprocessing Problem; Crack; Poor Quality Image; Structural Problem; Detachment of Device or Device Component; Material Deformation; Mechanical Jam; Material Split, Cut or Torn; Physical Resistance/Sticking

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS ON THE AIR/WATER TUBE AND CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22414206 ↗
  • MalfunctionApr 24, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED FOREIGN MATERIAL IN AIR/WATER TUBE AND CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21904899 ↗
  • MalfunctionMar 31, 2025Melted; Failure to Clean Adequately; Melted; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED A BURNT PLASTIC DISTAL END COVER AND THE AIR/WATER CHANNEL HAD WHITE FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21731753 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131780. One submission often covers many models, so check the product and lot details in each record.

  • Z-1210-2024 · Jan 11, 2024 · Open, Classified
    Colonoscope, Model Number PCF-H190DL.
  • Z-0418-2024 · Oct 16, 2023 · Open, Classified
    PCF-H190DL, EVIS EXERA III COLONOVIDEOSCOPE
  • Z-0417-2024 · Oct 16, 2023 · Open, Classified
    GIF-HQ190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
  • Z-0415-2024 · Oct 16, 2023 · Open, Classified
    CF-HQ190L EVIS EXERA III COLONOVIDEOSCOPE
  • Z-0419-2024 · Oct 16, 2023 · Open, Classified
    PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE
  • Z-0416-2024 · Oct 16, 2023 · Open, Classified
    GIF-H190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
  • Z-0056-2025 · Sep 14, 2023 · Open, Classified
    EVIS EXERA III Video System Center Model Number CV-190. Intended to be used with OLYMPUS camera heads, endoscopes, light sources, monitors, EndoTherapy accessories, and other ancil

Other Colonoscope and accessories, flexible/rigid devices

All Colonoscope and accessories, flexible/rigid devices →

Sources and updates