EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE N180 (UDI 04953170307508)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus GIF N180 video scope is a flexible colonoscope used to examine the digestive tract. Its 65 reports mostly describe contamination inside the scope found after reprocessing, cleaning failures, debris in the air/water cylinder, peeling or damaged tip parts, and poor or missing images. No deaths and no injuries are reported. One open recall from September 2023 covers scopes returned from repair with a wet channel because air drying was not validated, creating a contamination and infection risk.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
64
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 04953170307508), received between May 16, 2018 and Mar 10, 2026 — classified as 64 malfunctions.

The most frequently reported specific device problem is “Microbial Contamination of Device” (24 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 4 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0334-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Microbial Contamination of Device24
  • Failure to Clean Adequately15
  • Poor Quality Image8
  • Device Reprocessing Problem7
  • Contamination6
  • Image Display Error/Artifact2
  • Complete Blockage1
  • Defective Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions50
  • No Known Impact Or Consequence To Patient12
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 10, 2026Poor Quality Image

    IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED DAMAGE TO THE CHARGE-COUPLED DEVICE UNIT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24563112 ↗
  • MalfunctionJan 6, 2026Device Reprocessing Problem; No Display/Image; Defective Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED NO IMAGE, DAMAGED CHARGED-COUPLED UNIT AND FOREIGN OBJECTS IN THE C-COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23981847 ↗
  • MalfunctionAug 4, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED C-COVER HAS FOREIGN OBJECTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22699767 ↗
  • MalfunctionJul 25, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE INSERTION PART OF CHARGE COUPLED DEVICE (CCD) UNIT PEELED OFF. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22612879 ↗
  • MalfunctionJul 18, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED FOREIGN OBJECTS IN THE AIR/WATER CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22540690 ↗
  • MalfunctionJul 2, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SCOPE EXHIBITED FOREIGN MATERIAL IN THE AIR/WATER CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22383000 ↗
  • MalfunctionJun 18, 2025Poor Quality Image; Poor Quality Image

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE SUBJECT DEVICE EXHIBITED INSERTION PART ¿ CCD UNIT ¿ PEEL-OFF, WHICH RESULTED IN POOR IMAGE QUALITY, AND INSERTION PART ¿ DISTAL END ¿ BIOPSY CHANNEL ¿ LEAK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22285110 ↗
  • MalfunctionJun 17, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED FOREIGN OBJECTS IN THE BIOPSY CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22234529 ↗
  • MalfunctionMay 19, 2025Poor Quality Image; Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A GRAINY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22047525 ↗
  • MalfunctionApr 25, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE BIOPSY CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21913077 ↗
  • MalfunctionNov 6, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEOSCOPE EXHIBITED A CLOGGED NOZZLE BY CHEMICAL RESIDUES. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20622555 ↗
  • MalfunctionOct 14, 2024Degraded; Degraded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED THE BENDING SECTION WAS BROKEN AND SHOWING A SIGNIFICANT METAL EXPOSURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20444238 ↗
  • MalfunctionOct 7, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN OBJECTS INSIDE THE AIR/WATER TUBE AND CYLINDER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20384965 ↗
  • MalfunctionSep 30, 2024Complete Blockage; Complete Blockage

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED THE NOZZLE OF THE INSUFFLATION CHANNEL CLOGGED BY AN UNIDENTIFIED PLASTIC MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20339551 ↗
  • MalfunctionAug 12, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE SUBJECT DEVICE EXHIBITED THE AIR/WATER (INSUFFLATION) CHANNEL IS CLOGGED WITH FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19968940 ↗
  • MalfunctionJun 28, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE VIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE CHANNEL TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19638200 ↗
  • MalfunctionJun 17, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT THE GASTROINTESTINAL VIDEOSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19548679 ↗
  • MalfunctionJun 4, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT, DURING REPROCESSING, THE VIDEOSCOPE TESTED POSITIVE FOR 1 COLONY FORMING UNITS (CFUS) OF MICROORGANISMS THAT CAN BE REVIVED AT 30° C AND BACILLACEAE WAS DETECTED. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19461724 ↗
  • MalfunctionJun 3, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED WHITE MATERIAL IN THE AIR/WATER TUBE AND CYLINDER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19449756 ↗
  • MalfunctionMay 1, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED WHITE FOREIGN MATERIAL INSIDE THE NOZZLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19220901 ↗
  • MalfunctionApr 19, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT DURING REPROCESSING, THE GASTROINTESTINAL VIDEOSCOPE TESTED POSITIVE FOR 7 COLONY FORMING UNITS (CFUS)/100ML OF AEROBIC MICROORGANISMS. A SUBSEQUENT TEST WAS PERFORMED AND TESTED POSITIVE FOR 8 CFU'S/100ML OF AEROBIC MICROORGANISMS. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19149342 ↗
  • MalfunctionFeb 13, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED THE NOZZLE WAS CLOGGED BY A PIECE OF CLEAR AND YELLOW-COLORED PLASTIC MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18694931 ↗
  • MalfunctionFeb 5, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A DARK RUBBER MATERIAL THAT CLOGGED THE NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18640993 ↗
  • MalfunctionNov 14, 2023Microbial Contamination of Device; Microbial Contamination of Device

    THE CUSTOMER REPORTED TO OLYMPUS, DURING REPROCESSING, THE GASTROINTESTINAL VIDEOSCOPE TESTED POSITIVE OF ACINETOBACTER SP. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18136879 ↗
  • MalfunctionOct 10, 2023Microbial Contamination of Device; Microbial Contamination of Device

    THE CUSTOMER REPORTED TO OLYMPUS, THE EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE TESTED POSITIVE FOR UNEXPECTED CONTAMINATION. THE SCOPE WAS SAMPLED ONCE. DURING THE SAMPLING, THE SCOPE TESTED POSITIVE FOR > 80 COLONY FORMING UNITS (CFUS) OF ESCHERICHIA COLI. THE ISSUE WAS FOUND AT REPROCESSING DURING A ROUTINE CULTURE OF THE SCOPE. ALL CHANNELS WERE TESTED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17907404 ↗

Other Colonoscope and accessories, flexible/rigid devices

All Colonoscope and accessories, flexible/rigid devices →