Gastrointestinal videoscope-indicated for use within the… recallZ-0334-2024 · Aizu Olympus Co., Ltd. · initiated Sep 25, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Aizu Olympus Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
FDA root cause category: Under Investigation by firm
What the company is doing
Olympus America issued Urgent Medical Device Correction letter on Sept. 25, 2023 to Endoscopy, Respiratory, Pulmonary and Urology Departments, Risk Management Department. Letter states reason for recall, health risk and action to take: Our records indicate that Olympus has serviced your flexible endoscope at one of our Olympus repair facilities. Following repair activities, Olympus uses a water flush of endoscope channels as part of our final endoscope channel inspection, and we subsequently dry the endoscope channels with compressed filtered air. Olympus discovered that our channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Olympus is in the process of implementing a validated drying process at our repair centers. In the meantime, Olympus has established interim processes to perform an additional inspection of your scope for water prior to return. 1. Ensure your users are knowledgeable on the content of this letter. 2. Complete the enclosed response form and return it to our recall partner, Sedgwick, via e-mail at [email protected] or by fax at 844-294-7184. For any questions about the acknowledgement form, please call the Sedgwick team at 888-667-1504. 3. Olympus requests that you report complaints, including infections, and endoscopes returned wet from our repair centers to Olympus. Please report complaints to Technical Assistance Center (TAC) at 1-800-848-9024, option 1. Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you require additional information, please do not hesitate to contact me at (647) 999-3203 or [email protected].
Recall details
- Recall number
- Z-0334-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 25, 2023
- Posted by the FDA
- Nov 17, 2023
- Recalling firm
- Aizu Olympus Co., Ltd.
- Quantity
- 3627 units
- Product code
- FDF
- FDA submissions
- K100584, K954451, K051645, K111756, K112680, K011151
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q160Z, GIF-Q180, GIF-XP180N, GIF-XP190N, GIF-XTQ160. 510(k) Numbers: K112680, K954451, K051645, K100584, K111756, K011151, K123317.
Code information (models, lots, serial numbers)
Model Numbers: GIF-1TH190 (04953170343360), GIF-1TQ160 (Not recorded), GIF-2T160 (04953170307492), GIF-2TH180 (04953170339745), GIF-H170 (04953170334115), GIF-H180 (04953170339769), GIF-H180J (04953170307539), GIF-H190 (04953170305290), GIF-HQ190 (04953170305276), GIF-LV1 (04953170312373), GIF-N180 (04953170307508), GIF-Q150 (04953170307485), GIF-Q160Z (04953170066795), GIF-Q180 (04953170307522), GIF-XP180N (04953170307515), GIF-XP190N (04953170305313), GIF-XTQ160 (04953170339837).
Devices named in this recall
16 devices in the FDA device registry whose identifier appears in the recall's code information.
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,025 reports
- GASTROINTESTINAL VIDEOSCOPE GIF-H170Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall852 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall559 reports
- GASTROINTESTINAL VIDEOSCOPE GIF-Q150Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall330 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180JOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall202 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE 2TH180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall76 reports
- GASTROINTESTINAL VIDEOACOPE OLYMPUS GIF-LV1Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall67 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE N180Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall64 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE Q180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall32 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall30 reports
- EVIS EXERA GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE XTQ160Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall30 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE XP180NOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall12 reports
- EVIS EXERA GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE Q160ZOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall4 reports
- EVIS EXERA GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE 2T160Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall3 reports
Product lines
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPEOlympus Medical Systems Corp. · 20 models · 2,267 reports · 1 recall
- EVIS EXERA GASTROINTESTINAL VIDEOSCOPEOlympus Medical Systems Corp. · 5 models · 71 reports · 1 recall
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPEOlympus Medical Systems Corp. · 7 models · 13,448 reports · 7 recalls
- GASTROINTESTINAL VIDEOSCOPEOlympus Medical Systems Corp. · 8 models · 2,677 reports · 2 recalls
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPEOlympus Medical Systems Corp. · 2 models · 587 reports · 1 recall
Other products in the same recall event
The FDA groups related products into one recall event (event 93222).
- Z-0327-2024 · Open, ClassifiedBronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers: BF-1T150, BF-1T60, BF-1TH190, BF-1TQ170, BF-3C160, BF-3C40, BF-H190…
- Z-0328-2024 · Open, ClassifiedBronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, …
- Z-0329-2024 · Open, ClassifiedColonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). Model Numbers: CF-2T160I…
- Z-0330-2024 · Open, ClassifiedCholeodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60.
- Z-0331-2024 · Open, ClassifiedCystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
- Z-0332-2024 · Open, ClassifiedRhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea). Model Numbers: ENF-T3, ENF-VT2, ENF-V…
- Z-0333-2024 · Open, ClassifiedGastroscope-endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) and for endoscopic surgery within the upper digestive tract. Model Numbers: GF-UC140P-AL5, GF-UCT180, …
- Z-0335-2024 · Open, ClassifiedHysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.
- Z-0336-2024 · Open, ClassifiedHysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.
- Z-0337-2024 · Open, ClassifiedHysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP.
- Z-0338-2024 · Open, ClassifiedTracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: LF-DP, LF-GP, LF-…
- Z-0339-2024 · Open, ClassifiedTracheal videoscope- For airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Number: LF-V.
- Z-0340-2024 · Open, ClassifiedMobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-…
- Z-0341-2024 · Open, ClassifiedColonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). Model Numbers: PCF-H180AI, PCF-H180AL, PCF-…
- Z-0342-2024 · Open, ClassifiedSignmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum and sigmoid colon). Model Number: PCF-S.
- Z-0343-2024 · Open, ClassifiedSmall intestinal videoscope-For endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach, duodenum, and small intestine) by oral insertio…
- Z-0344-2024 · Open, ClassifiedTransnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model Number: PEF-V.
- Z-0345-2024 · Open, ClassifiedUltrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding…
- Z-0346-2024 · Open, ClassifiedDuodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V…
- Z-0347-2024 · Open, ClassifiedUreteralscope- for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct). Model Numbers: URF-P6, URF-P6R.
- Z-0348-2024 · Open, ClassifiedUretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page