Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04953170307720), received between Jan 14, 2021 and Dec 5, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (3 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jan 11, 2024Class IIOpen, Classified
Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Z-1214-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0341-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Clean Adequately3
- Partial Blockage2
- Device Reprocessing Problem1
- Material Twisted/Bent1
- Melted1
- No Display/Image1
- Poor Quality Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Patient Involvement1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 5, 2025No Display/Image; Poor Quality Image; Failure to Clean Adequately; No Display/Image; Poor Quality Image; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED IMAGE DISTORTION (A NOISE IMAGE), NO IMAGE, FOREIGN OBJECTS COME OUT OF AIR/WATER CYLINDER, IMAGE GUIDE PROTECTOR IS DIRTY, SCOPE-CONNECTOR IS DIRTY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23727764 ↗MalfunctionNov 25, 2025Melted; Melted
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLON VIDEOSCOPE EXHIBITED BURNED PLASTIC ON THE DISTAL END COVER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23648749 ↗MalfunctionOct 9, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS ON THE BENDING SECTION COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20407266 ↗MalfunctionMay 2, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE COLONOVIDEOSCOPE AIR CHANNEL WAS UNABLE TO WORK DUE TO A CLOGGED NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19235021 ↗MalfunctionMar 10, 2021Partial Blockage; Partial Blockage
IT IS REPORTED WHILE PRE-CLEANING OF AN EVIS EXERA II COLONOVIDEOSCOPE, A CLIP WAS DISLODGED FROM THE INSTRUMENT CHANNEL. THERE WAS NO PATIENT CONTACT/IMPACT RELATED TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11453815 ↗MalfunctionJan 14, 2021Partial Blockage; Device Reprocessing Problem; Material Twisted/Bent; Partial Blockage; Device Reprocessing Problem; Material Twisted/Bent
IT IS REPORTED WHILE PREPARING AN EVIS EXERA II COLONOVIDEOSCOPE FOR USE, BLACK SPECS CAME OUT OF THE DEVICE WHEN IT WAS HOOKED UP TO A DRYER. THERE WAS NO PATIENT CONTACT/IMPACT RELATED TO THIS OCCURRENCE.
Patient problems: No Patient Involvement
MAUDE report 11174013 ↗
Other Gastroscope and accessories, flexible/rigid devices
- PENTAX Medical EG29-i10cHoya Corporation · Gastroscope And Accessories, Flexible/Rigid1,623 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,025 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall559 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GIF-Q165Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid349 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 death302 reports
- PENTAX EG27-i10Hoya Corporation · Gastroscope and accessories, flexible/rigid261 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 deaths252 reports
- BALLOON MAJ-1351Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid174 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page