Skip to content

Adapter for UHI-2/3 MAJ-1873Recalls & reported problems

Olympus Medical Systems Corp. · Class U

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170308475), received between Jun 22, 2022 and Sep 10, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Display or Visual Feedback Problem1
  • Failure to Power Up1
  • Gas/Air Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THE HIGH FLOW INSUFFLATION UNIT HAD A GAS LEAK OCCURRED. THE ISSUE OCCURRED DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20181042 ↗
  • MalfunctionOct 20, 2023Failure to Power Up; Failure to Power Up

    THE CUSTOMER REPORTED THAT THE DEVICE WOULD NOT SWITCH ON. THE REPORTED ISSUE WAS OBSERVED WHILE PREPARING THE SUBJECT DEVICE, PRIOR TO AN UNSPECIFIED PROCEDURE. THERE WAS NO REPORT OF PATIENT OR USER INJURY DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17976516 ↗
  • MalfunctionJun 22, 2022Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THE DISPLAY DOES NOT LIGHT UP. THE ISSUE FOUND DURING PREPARATION FOR USE. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14767494 ↗

Other Instrument, ultrasonic surgical devices

All Instrument, ultrasonic surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page