CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VH (UDI 04953170310461)Recalls & reported problems

Olympus Medical Systems Corp. · CYF-VH HD Flex CystoNephro Videoscope · Class 2

All models of CYSTO-NEPHRO VIDEOSCOPE (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus CystoNephro videoscope has 425 reports, including 28 injuries and no deaths. Many reports describe parts coming loose, such as glue or covers detaching near the lens and bending section, and some describe trouble cleaning the device properly. Reports rose to 173 in the last year from 95 in the year before. 1 open recall, from December 2023, concerns updated cleaning instructions: reusable brushes were removed from manual cleaning.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
435
Deaths reported
0
Injuries reported
27
Recalls naming it
1

Summary

The FDA's MAUDE database contains 435 adverse event reports that list this exact device (DI 04953170310461), received between May 24, 2018 and Sep 30, 2026 — classified as 27 injuries, 408 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (217 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 7 reportsJan 2024: 7 reportsFeb 2024: 8 reportsMar 2024: 4 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 2 reportsMar 2025: 3 reportsApr 2025: 3 reportsMay 2025: 5 reportsJun 2025: 1 reportsJul 2025: 4 reportsAug 2025: 7 reportsSep 2025: 8 reportsOct 2025: 59 reportsNov 2025: 17 reportsDec 2025: 11 reportsJan 2026: 18 reportsFeb 2026: 3 reportsMar 2026: 13 reportsApr 2026: 31 reportsMay 2026: 33 reportsJun 2026: 22 reportsJul 2026: 11 reportsAug 2026: 12 reportsSep 2026: 12 reportsOct 2026: 0 reports20242025202659
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 4, 2023Class IIOpen, Classified

    Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.

    Z-0689-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component217
  • Failure to Clean Adequately40
  • Device Reprocessing Problem33
  • Corroded17
  • Material Deformation14
  • Material Discolored14
  • Material Separation14
  • Degraded13

Patient problems

  • No Clinical Signs, Symptoms or Conditions402
  • Unspecified Infection13
  • Bacterial Infection7
  • No Patient Involvement6
  • Fever3
  • Sepsis3
  • Urinary Frequency / Polyuria3
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 25 most recent of 435 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Detachment of Device or Device Component

    The customer returned the Cysto-Nephro Videoscope to the service center for evaluation related to a separate issue. During testing, the service repair technician identified that the device bending section was detached. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950985 ↗
  • MalfunctionSep 29, 2026Degraded

    The customer reported the Cysto-Nephro Videoscope had a separate issue. During the initial evaluation of the device it was observed, that there was corrosion around the objective lens. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931451 ↗
  • MalfunctionSep 25, 2026Detachment of Device or Device Component

    The customer returned the Cysto-Nephro Videoscope for a separate issue. During the device analysis, the bending section was detached. There was no patient involvement, and no harm was reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50893613 ↗
  • MalfunctionSep 24, 2026Material Fragmentation; Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50892356 ↗
  • MalfunctionSep 24, 2026Corroded; Crack; Detachment of Device or Device Component

    The customer returned the Cysto-Nephro Videoscope for a separate issue. During the device analysis, the bending section cover glue was cracked off/detached, and the Objective lens had corrosion. There was no patient involvement, and no harm was reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889286 ↗
  • MalfunctionSep 23, 2026Corroded

    The customer reported that the Cysto-Nephro Videoscope had a separate issue. During the initial evaluation of the device, it was observed that the objective lens had corrosion around the lens. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50871865 ↗
  • MalfunctionSep 23, 2026Detachment of Device or Device Component

    The customer returned the Cysto Nephro Videoscope for a separate issue. During the initial evaluation of returned device it was found the adhesive around the objective lens is missing. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870333 ↗
  • MalfunctionSep 22, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50859264 ↗
  • MalfunctionSep 4, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50618921 ↗
  • MalfunctionSep 3, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50589092 ↗
  • MalfunctionSep 3, 2026Detachment of Device or Device Component

    The customer returned the Cysto-Nephro Videoscope for a separate issue. During the evaluation it was observed that the distal end was detached. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50577716 ↗
  • MalfunctionSep 3, 2026Dent in Material; Material Split, Cut or Torn

    The customer returned the Cysto-Nephro Videoscope for a separate issue. During the evaluation it was observed that the insertion tube was damaged with exposed mesh. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50577684 ↗
  • MalfunctionAug 27, 2026Detachment of Device or Device Component

    The customer returned the Cysto-Nephro Videoscope for a separate issue. During the initial evaluation of the device it was observed that there was a missing glue on the objective lens. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50487379 ↗
  • MalfunctionAug 19, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50327346 ↗
  • MalfunctionAug 14, 2026Contamination /Decontamination Problem; Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50253941 ↗
  • MalfunctionAug 12, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50217924 ↗
  • MalfunctionAug 11, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50216751 ↗
  • MalfunctionAug 11, 2026Crack

    The customer returned the Cystonephrofiberscope to the service center for a separate issue. During the device analysis, the bending section adhesive was found detached. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50203791 ↗
  • MalfunctionAug 10, 2026Detachment of Device or Device Component

    The customer returned the Cysto Nephro Videoscope to the service center for a separate issue. During the device analysis, the service repair technician indicated distal end plastic cover detached from bending section. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50197697 ↗
  • MalfunctionAug 7, 2026Material Separation

    The cysto-nephro videoscope was returned to the service center for a separate issue. During evaluation of the device, the objective lens adhesive was found to be missing/detached, likely due to degradation, and was repaired. There was no patient involvement, and no harm was reported as a result of the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50181930 ↗
  • MalfunctionAug 7, 2026Detachment of Device or Device Component

    The customer returned the Cysto-Nephro Videoscope to the service center for a separate issue. During the device analysis, it was indicated that a detached C-Cover was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50179392 ↗
  • MalfunctionAug 7, 2026Loss of or Failure to Bond

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50168814 ↗
  • MalfunctionAug 6, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50168379 ↗
  • MalfunctionAug 6, 2026Stretched

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50159233 ↗
  • MalfunctionJul 31, 2026Detachment of Device or Device Component

    The customer returned the Cysto-Nephro Videoscope to the service center for a separate issue. During the device analysis, the repair technician indicates that a detached distal end was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50105849 ↗

Other Endoscope, accessories, narrow band spectrum devices

All Endoscope, accessories, narrow band spectrum devices →