GASTROINTESTINAL VIDEOACOPE OLYMPUS GIF-LV1Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus GIF-LV1 video scope is a flexible colonoscope used to look inside the digestive tract. Its 67 reports mostly describe cleaning and reprocessing problems, including residue or foreign objects found inside the air/water nozzle and tip, and poor image quality. No deaths and no injuries are reported. One open recall, from September 2023, covers scopes returned after repair whose channel was still wet, which can allow contamination and infection.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
67
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 67 adverse event reports that list this exact device (DI 04953170312373), received between Nov 24, 2021 and May 15, 2025 — classified as 67 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (34 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 3 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 4 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 5 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0334-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Clean Adequately34
  • Poor Quality Image15
  • Contamination9
  • Device Reprocessing Problem7
  • Contamination /Decontamination Problem4
  • Residue After Decontamination2
  • Degraded1
  • Device Contamination with Chemical or Other Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions66
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 67 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2025Residue After Decontamination; Residue After Decontamination

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22031666 ↗
  • MalfunctionMay 9, 2025Electrical /Electronic Property Problem; Flare or Flash; Electrical /Electronic Property Problem; Flare or Flash

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD A BURN ON PLASTIC DISTAL END PART AND COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21993427 ↗
  • MalfunctionApr 22, 2025Residue After Decontamination; Residue After Decontamination

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN THE CONNECTING TUBE AND NOZZLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 21886062 ↗
  • MalfunctionOct 2, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED THE BENDING SECTION COVER, NOZZLE AND ADHESIVE AROUND OBJECTIVE LENS HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20353466 ↗
  • MalfunctionSep 17, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN OBJECTS WITHIN THE NOZZLE AND DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20239644 ↗
  • MalfunctionSep 12, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIALS WAS EXITING FROM THE AIR/WATER NOZZLE. THE ISSUE OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM OR PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20200890 ↗
  • MalfunctionSep 9, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN OBJECTS ON THE NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20170870 ↗
  • MalfunctionSep 9, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED, DURING THE DEVICE INSPECTION. THAT THE NOZZLE OF THE VIDEOSCOPE HAD FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20168472 ↗
  • MalfunctionSep 4, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN THE NOZZLE, ADHESIVE ON THE BENDING SECTION COVER (A-RUBBER), AND THE DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20136933 ↗
  • MalfunctionAug 23, 2024Poor Quality Image; Failure to Clean Adequately; Poor Quality Image; Failure to Clean Adequately

    IT WAS REPORTED THE SUBJECT DEVICE HAD AN IMAGE ISSUE. ON SEPTEMBER 2, 2024 THE CUSTOMER CONFIRMED THAT THE IMAGE QUALITY WAS NOT GOOD AND THE IMAGE WAS BLURRED. THE ISSUE OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20050243 ↗
  • MalfunctionJul 8, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE NOZZLE AND BENDING SECTION COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19688525 ↗
  • MalfunctionJul 2, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS REPORTED THE GASTROINTESTINAL HAS FOREIGN MATTER IN THE CHANNEL TUBE. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR A DIAGNOSTIC GASTROSCOPE PROCEDURE. THERE WAS NO DELAY AND THE PROCEDURE WAS COMPLETED BY USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19660896 ↗
  • MalfunctionJun 22, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN THE NOZZLE AND ADHESIVE ON THE BENDING SECTION COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19590544 ↗
  • MalfunctionJun 12, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE GASTROINTESTINAL VIDEOSCOPE HAD FOREIGN MATERIALS IN NOZZLE, PLASTIC DISTAL END COVER (C-COVER), ADHESIVE AROUND OBJECTIVE LENS, BENDING SECTION COVER (A-RUBBER), UPWARD/DOWNWARD AND RIGHT/LEFT KNOBS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19522289 ↗
  • MalfunctionJun 10, 2024No Display/Image; Poor Quality Image; No Display/Image; Poor Quality Image

    IT WAS REPORTED THAT THE GASTROINTESTINAL VIDEOSCOPE HAD A BLURRED IMAGE. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19506688 ↗
  • MalfunctionJun 10, 2024Power Problem; Power Problem

    PATIENT WAS NOT UNDER ANESTHESIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19500067 ↗
  • MalfunctionMay 7, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A FOREIGN OBJECT ON THE NOZZLE, THE ADHESIVE AROUND OBJECTIVE LENS, THE BENDING SECTION COVER, UP/DOWN KNOB, RIGHT/LEFT KNOB AND THE CONNECTING TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19259853 ↗
  • MalfunctionApr 26, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN THE NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19198342 ↗
  • MalfunctionApr 3, 2024Device Reprocessing Problem; Device Reprocessing Problem

    DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE NOZZLE, PLASTIC DISTAL END COVER, BENDING SECTION COVER, ADHESIVE AROUND THE OBJECTIVE LENS, CONNECTING TUBE, AND THE UP/DOWN/RIGHT/LEFT KNOBS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19038358 ↗
  • MalfunctionApr 2, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED ADHESIVE AROUND OBJECTIVE LENS HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19029431 ↗
  • MalfunctionMar 20, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING EVALUATION THAT THE NOZZLE AND THE ADHESIVE AROUND OBJECTIVBEL LENS HAD FOREIGN OBJECTS ON THEM. THE BENDING SECTION COVER ALSO HAD FOREIGN OBJECTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18944958 ↗
  • MalfunctionMar 19, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOACOPE EXHIBITED ADHESIVE ON BENDING SECTION COVER HAD FOREIGN OBJECTS, CONNECTING TUBE HAD FOREIGN OBJECTS AND BENDING SECTION COVER HAD FOREIGN OBJECTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18935572 ↗
  • MalfunctionMar 5, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE NOZZLE AND IN THE ADHESIVE AROUND THE OBJECTIVE LENS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18837254 ↗
  • MalfunctionFeb 1, 2024Difficult to Insert; Failure to Clean Adequately; Difficult to Insert; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEOSCOPE EXHIBITED FOREIGN MATERIAL INSIDE THE CHANNEL BLOCKING BRUSH AND FORCEPS FROM BEING INSERTED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18627947 ↗
  • MalfunctionDec 20, 2023Failure to Clean Adequately; Failure to Clean Adequately

    THE CUSTOMER REPORTED TO OLYMPUS, THE GASTROINTESTINAL VIDEOACOPE HAD REDUCED ANGULATION. THE ISSUE WAS FOUND DURING GENERAL ROUTINE INSPECTION. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE BENDING SECTION COVER HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18365652 ↗

Other Colonoscope and accessories, flexible/rigid devices