COLONOVIDEOSCOPE OLYMPUS CF-LV1LRecalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of COLONOVIDEOSCOPE (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04953170312397), received between Jan 25, 2022 and Nov 12, 2024 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0329-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Clean Adequately7
- Contamination2
- No Display/Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 12, 2024No Display/Image; No Display/Image
IT WAS REPORTED, THE FLEX VIDEO SCOPE HAD IMAGE LOSS. IMAGE DISAPPEARED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20668223 ↗MalfunctionJun 17, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THERE WAS FOREIGN MATERIAL IN THE NOZZLE OF THE COLONOVIDEOSCOPE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19555099 ↗MalfunctionJun 10, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE NOZZLE AND DISTAL END COVER HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19500267 ↗MalfunctionDec 27, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS, THE COLONOVIDEOSCOPE HAD REDUCED ANGULATION/PLAY IN SCOPE. THE ISSUE WAS FOUND DURING GENERAL ROUTINE INSPECTION BY THE CUSTOMER. THE OLYMPUS FIELD SERVICE ENGINEER CONFIRMED THE DEVICE WAS REPROCESSED AT THE CUSTOMER SITE, BEFORE PICK UP. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE FOUND: NOZZLE, ADHESIVE AROUND OBJECTIVE LENS, AND THE BENDING SECTION COVER ALL HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18406790 ↗MalfunctionNov 9, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE OLYMPUS FIELD SERVICE ENGINEER REPORTED (ON BEHALF OF THE CUSTOMER) THAT THERE WAS NO UPWARD/DOWNWARD ANGULATION ON THE COLONOVIDEOSCOPE. THE ISSUE WAS FOUND DURING AN UNSPECIFIC PROCEDURE. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: THE BENDING SECTION COVER HAD FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18107151 ↗MalfunctionAug 8, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS MEDICAL THAT THE COLONOVIDEOSCOPE HAD A WRINKLED, DISCOLORED CONNECTOR TUBE, AN LED ISSUE AND LOW ANGULATION CONTROL. THE CUSTOMER REPORTED THE ISSUE WAS FOUND DURING MAINTENANCE. THE DEVICE WAS RETURNED TO OLYMPUS MEDICAL FOR EVALUATION. DURING EVALUATION, FOREIGN MATERIAL WAS FOUND IN THE AIR/WATER NOZZLE AND CONNECTOR TUBE AND THE ADHESIVE ON THE BENDING SECTION COVER WAS DIRTY. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTIONS FOUND DURING EVALUATION. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17482394 ↗MalfunctionMay 24, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE COLONOVIDEOSCOPE WAS RETURNED AS PART OF THE ASSET RETURN PROCESS. DURING ROUTINE INSPECTION OF THE DEVICE BY OLYMPUS, A YELLOWISH/BROWN FOREIGN OBJECT WAS FOUND IN THE NOZZLE. THERE WAS NO PROCEDURAL OR PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT. THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16995030 ↗MalfunctionMar 31, 2022Contamination; Contamination
THE USER FACILITY REPORTED TO OLYMPUS DURING PREPARATION FOR USE, SCOPE LED WAS NOT WORKING ON THE COLONOVIDEOSCOPE. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT. DURING THE EVALUATION OF THE RETURNED DEVICE, A FOREIGN MATERIAL WAS OBSERVED INSIDE THE AIR/WATER NOZZLE. THIS REPORT IS BEING SUBMITTED AS AN MDR BASED ON REPORTABLE MALFUNCTION OBSERVED DURING THE EVALUATION OF THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13959534 ↗MalfunctionMar 30, 2022Contamination; Contamination
CUSTOMER RETURNED THE DEVICE FOR INSPECTION. THERE IS NO REPORTED HARM TO ANY PATIENT. UPON EVALUATION OF THE RETURNED DEVICE, IT WAS OBSERVED THAT THE NOZZLE HAD FOREIGN MATERIAL. THIS MEDWATCH IS BEING SUBMITTED FOR THE ISSUE OF FOREIGN MATERIAL FOUND IN THE NOZZLE DURING DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13955568 ↗MalfunctionJan 25, 2022Failure to Clean Adequately; Failure to Clean Adequately
THE USER FACILITY REPORTED TO OLYMPUS THE COLONOVIDEOSCOPE HAD PLAY IN UPWARD/DOWNWARD ANGULATION CONTROL KNOB, LEAKAGE FROM THE BIOPSY CHANNEL, THE COAT WAS PEELING, BENDING MANIPULATION AND INSERTION TUBE DEFECTS. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS OBSERVED THAT THE NOZZLE HAD FOREIGN MATERIAL WHICH OCCURRED FROM INSUFFICIENT CLEANING. THIS REPORT IS BEING SUBMITTED FOR THE EVENT FOUND DURING ESTIMATION. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13347739 ↗
Other Colonoscope and accessories, flexible/rigid devices
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- PENTAX Medical EC38-i10cLHoya Corporation · Colonoscope and accessories, flexible/rigid1,618 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page