COLONOVIDEOSCOPE OLYMPUS CF-LV1IRecalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of COLONOVIDEOSCOPE (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04953170312410), received between Jul 6, 2022 and Mar 5, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0329-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Clean Adequately3
- Contamination2
- Contamination /Decontamination Problem1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 5, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS INSIDE THE NOZZLE AND BENDING SECTION COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18836876 ↗MalfunctionDec 19, 2022Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS, THE COLONOVIDEOSCOPE EXHIBITED AIR/WATER FEEDING PROBLEM. THE ISSUE WAS FOUND DURING AN ENTEROSCOPY PROCEDURE. THE PATIENT WAS NOT UNDER SEDATION. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THE DEVICE WAS RETURNED AND AN EVALUATION REVEALED, THE DISTAL END OF THE DEVICE WAS DIRTY. THIS REPORT IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16003193 ↗MalfunctionOct 14, 2022Contamination; Contamination
THE CUSTOMER REPORTED THE SUBJECT DEVICE EXHIBITED INSUFFICIENT AIR/WATER SUPPLY. THE REPORTED ISSUE OCCURRED DURING INSPECTION OF THE SUBJECT DEVICE, PRIOR TO AN UNSPECIFIED DIAGNOSTIC PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING ANOTHER DEVICE WITHOUT A DELAY. THERE WAS NO REPORT OF PATIENT OR USER INJURY DUE TO THE EVENT. THE SUBJECT DEVICE WAS RETURNED TO AN OLYMPUS SERVICE CENTER FOR EVALUATION. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS OBSERVED THE NOZZLE WAS CLOGGED WITH SOME RUBBER-LIKE FOREIGN MATERIAL. THIS REPORT IS BEING SUBMITTED FOR THE FOREIGN MATERIAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15610755 ↗MalfunctionAug 2, 2022Obstruction of Flow; Obstruction of Flow
A USER FACILITY REPORTED TO OLYMPUS THAT AIR/WATER WOULD NOT FLOW DUE TO INSUFFICIENT OR INCORRECT REPROCESSING OF THE SCOPE. THE PROBLEM OCCURRED DURING AN ENTEROSCOPY PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING ANOTHER SIMILAR DEVICE WITH NO DELAY IN PROCEDURE REPORTED. THERE WAS NO PATIENT INJURY, ASSOCIATED WITH THE PROBLEM, REPORTED TO OLYMPUS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15149728 ↗MalfunctionJul 19, 2022Contamination /Decontamination Problem; Failure to Clean Adequately; Contamination /Decontamination Problem; Failure to Clean Adequately
A USER FACILITY REPORTED TO OLYMPUS, THE COLONOVIDEOSCOPE HAD AIR/WATER FLOW PROBLEM. THE EVENT OCCURRED DURING PREPARATION FOR USE, PRIOR TO AN ENTEROSCOPY PROCEDURE. IT WAS REPORTED THAT THE PROCEDURE WAS COMPLETED USING SIMILAR DEVICE WITH NO DELAYED. UPON INSPECTION AND TESTING OF THE CUSTOMER RETURNED DEVICE, FIBROUS FOREIGN MATERIAL WAS FOUND CLOGGED IN THE SCOPE NOZZLE DUE TO INSUFFICIENT OR INCORRECT REPROCESSING. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. THERE WAS NO REPORT OF PATIENT HARM OR USER INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15053481 ↗MalfunctionJul 6, 2022Contamination; Contamination
THE CUSTOMER REPORTED THE SUBJECT DEVICE EXHIBITED REDUCED ANGULATION AND ISSUES WITH THE AIR/WATER FLOW. THE REPORTED ISSUE OCCURRED WHILE PREPARING THE SUBJECT DEVICE FOR USE, PRIOR TO A DIAGNOSTIC ENTEROSCOPY. THE INTENDED PROCEDURE WAS COMPLETED USING THE SUBJECT DEVICE. THERE WAS NO PROCEDURE DELAY. THERE WAS NO PATIENT OR USER INJURY REPORTED DUE TO THE EVENT. THE DEVICE WAS RETURNED TO AN OLYMPUS SERVICE CENTER FOR EVALUATION. DURING INSPECTION AND TESTING, FOREIGN MATERIAL WAS OBSERVED ON/AROUND THE OBJECTIVE LENS AND THE AIR/WATER NOZZLE. THIS REPORT IS BEING SUBMITTED FOR THE MALFUNCTION FOUND DURING EVALUATION OF THE DEVICE (FOREIGN MATERIAL).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14935694 ↗
Other Colonoscope and accessories, flexible/rigid devices
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- PENTAX Medical EC38-i10cLHoya Corporation · Colonoscope and accessories, flexible/rigid1,618 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page