HIGH FLOW INSUFFLATION UNIT UHI-4 (REF N3829640)Recalls & reported problems

Olympus Medical Systems Corp. · "UHI-4 45L HIGHFLOW INSUFFLATION UNIT" · Class 2

All models of HIGH FLOW INSUFFLATION UNIT (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus UHI-4 is a carbon dioxide insufflator for endoscopy. Its 2,490 reports mention screens going blank, pressure problems and gas leaks, along with 6 deaths and 56 injuries. Three open recalls cover the front panel display switching off and air supply stopping, and patients suffering short cardiac arrests during surgery that may have come from over-inflating the abdomen. Reports fell from 234 in the prior 12 months to 67 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,484
Deaths reported
6
Injuries reported
56
Recalls naming it
3

Summary

The FDA's MAUDE database contains 2,484 adverse event reports that list this exact device (DI 04953170324147), received between Sep 6, 2019 and Aug 7, 2026 — classified as 6 deaths, 56 injuries, 2,422 malfunctions.

The most frequently reported specific device problem is “Pressure Problem” (292 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 77 reportsDec 2023: 76 reportsJan 2024: 83 reportsFeb 2024: 49 reportsMar 2024: 70 reportsApr 2024: 55 reportsMay 2024: 64 reportsJun 2024: 88 reportsJul 2024: 74 reportsAug 2024: 84 reportsSep 2024: 127 reportsOct 2024: 151 reportsNov 2024: 84 reportsDec 2024: 31 reportsJan 2025: 20 reportsFeb 2025: 17 reportsMar 2025: 8 reportsApr 2025: 5 reportsMay 2025: 2 reportsJun 2025: 5 reportsJul 2025: 5 reportsAug 2025: 20 reportsSep 2025: 9 reportsOct 2025: 28 reportsNov 2025: 24 reportsDec 2025: 3 reportsJan 2026: 4 reportsFeb 2026: 11 reportsMar 2026: 8 reportsApr 2026: 1 reportsMay 2026: 3 reportsJun 2026: 4 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports202420252026151
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 16, 2026Open, Classified

    Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.

    Z-2776-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. Sep 25, 2023Open, Classified

    Olympus has become aware of an increased trend of both repairs and customer complaints reporting "UHI-4 stopping air supply due to the front panel LED turning off".

    Z-1375-2024 · Root cause: Process control · Recall details · FDA record ↗

  3. Sep 22, 2023Open, Classified

    There have been reports of patients suffering arrhythmias, reported as short cardiac arrests, during surgical procedures where UHI-4s were used. These events may have been due to an over insufflation of the abdominal cavity resulting from use of the UHI-4 during the procedures.

    Z-0075-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pressure Problem292
  • No Display/Image247
  • Display or Visual Feedback Problem159
  • Failure to Power Up153
  • Gas/Air Leak140
  • No Visual Prompts/Feedback125
  • Electrical /Electronic Property Problem112
  • Loss of Power112

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,424
  • No Consequences Or Impact To Patient13
  • Cardiac Arrest12
  • Bradycardia6
  • Abdominal Distention4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
  • Insufficient Information4
  • No Patient Involvement4

Recent adverse event reports

The 25 most recent of 2,484 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2026No Display/Image

    It was reported that the high flow insufflation unit had the unit screen went black during a case. The issue was found during an unknown procedure. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50179803 ↗
  • MalfunctionJul 17, 2026Break; Gas/Air Leak

    THE CUSTOMER RETURNED THE [COMMON DEVICE] TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE ENGINEER INDICATED THAT THE CO2 GAS CONNECTOR/INPUT (K-CONNECTOR) WAS LEAKING AND THE PINCH VALVE HAD BROKEN PLASTICS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25861124 ↗
  • MalfunctionJul 15, 2026Key or Button Unresponsive/not Working

    THE CUSTOMER REPORTED THAT THE INSUFFLATOR WAS NOT WORKING, INVOLVING AN OLYMPUS HIGHT FLOW INSUFFLATION UNIT. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25833746 ↗
  • MalfunctionJun 24, 2026Gas/Air Leak

    THE CUSTOMER RETURNED THE HIGH FLOW INSUFFLATION UNIT TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE ENGINEER INDICATED THAT A FAILED SECONDARY PIPING LEAKAGE MEASUREMENT, AND A FAULTY ELECTROPNEUMATIC PROPORTIONAL VALVE. IT WAS DETERMINED THAT THE CIRCUIT BOARD NEEDED TO BE REPLACED.  THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25616852 ↗
  • MalfunctionJun 11, 2026Corroded

    IT WAS OBSERVED THAT DURING THE EVALUATION, THE HIGH FLOW INSUFFLATION UNIT EXHIBITED THAT THE AIR/WATER/SUCTION INSUFFLATION TUBE HAS RUST. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25480033 ↗
  • MalfunctionJun 8, 2026Break; Crack

    THE CUSTOMER RETURNED THE HIGH FLOW INSUFFLATION UNIT TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATED THE PLUG ON THE CABLE CONNECTING THE POWER SUPPLY AND THE CIRCUIT BOARD WAS FAULTY AND A CRACK WAS FOUND ON THE PINCH VALVE IN THE AIR/WATER/SUCTION SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25425455 ↗
  • MalfunctionJun 1, 2026Display or Visual Feedback Problem

    THE CUSTOMER RETURNED THE HIGH FLOW INSUFFLATION UNIT, WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, IT WAS OBSERVED THAT ONE LED DISPLAY SEGMENT WAS NOT WORKING ON THE PRESSURE SENSOR DISPLAY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25348022 ↗
  • MalfunctionMay 28, 2026Material Separation; Output Problem

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING EVALUATION OF THE DEVICE, IT WAS FOUND THAT THE POWER SWITCH HAD DISLOCATED FROM ORIGINAL POSITION AND IT HAD BECOME OBSOLETE AND LIGHTS ON DISPLAY APPEARED DARK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25320873 ↗
  • MalfunctionMay 20, 2026Erratic or Intermittent Display

    IT WAS OBSERVED DURING OLYMPUS' DEVICE INSPECTION THAT THE HIGH FLOW INSUFFLATOR FRONT PANEL WAS FAULTY. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25233425 ↗
  • MalfunctionMay 18, 2026Naturally Worn

    IT WAS REPORTED, THE INSUFFLATION UNIT EXHIBITED THE CO2 HOSE ASSEMBLY WAS WORN. THE EVENT OCCURRED DURING SET UP / INSPECTION FOR USE (DURING SET UP IN ROOM / BEFORE PATIENT IN ROOM). THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25202615 ↗
  • MalfunctionApr 30, 2026Computer Software Problem; No Visual Prompts/Feedback

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HIGH FLOW INSUFFLATION UNIT FRONT PANEL SEGMENT LIGHT EMITTING DIODE (LED) WAS BURNED OUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25048378 ↗
  • MalfunctionMar 26, 2026Electrical /Electronic Property Problem; Gas/Air Leak

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HIGH FLOW INSUFFLATION UNIT ELECTROPNEUMATIC PROPORTIONAL VALVE WAS FAULTY AND IT FAILED THE PRIMARY LEAKAGE TEST. ADDITIONALLY, THE PRINTED CIRCUIT BOARD WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700487 ↗
  • MalfunctionMar 23, 2026Power Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24662251 ↗
  • MalfunctionMar 20, 2026Power Problem

    IT WAS REPORTED THE HIGH FLOW INSUFFLATION UNIT WAS NOT WORKING. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649872 ↗
  • MalfunctionMar 16, 2026Electrical /Electronic Property Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HIGH FLOW INSUFFLATION UNIT EXHIBITED DIMMER LIGHT EMITTING DIODE INDICATORS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24606879 ↗
  • MalfunctionMar 9, 2026Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HIGH FLOW INSUFFLATION UNIT EXHIBITED A RUSTY INSUFFLATION TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24547622 ↗
  • MalfunctionMar 3, 2026Contamination; Gas Output Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HIGH FLOW INSUFFLATION UNIT EXHIBITED BLACK OILY RESIDUE INSIDE THE TUBING ALONG THE PNEUMATIC PIPELINE AND THE MANIFOLD SYSTEM, AND NO FLOW. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24497043 ↗
  • MalfunctionMar 3, 2026Gas/Air Leak

    IT WAS REPORTED THE HIGH FLOW INSUFFLATION UNIT CARBON DIOXIDE DRAINS VERY FAST AND THERE IS A LEAK. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24496451 ↗
  • MalfunctionMar 2, 2026Break; Circuit Failure; Crack

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HIGH FLOW INSUFFLATION UNIT FRONT PANEL LIGHT-EMITTING DIODES (LEDS) WERE DIMMER THEN INTENDED, THE FRONT PANEL NEEDED TO BE REPLACED, PRINTED CIRCUIT BOARD FAILURE NEEDED TO BE REPLACED, AND THE POWER SWITCH NEEDED TO BE REPLACED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24490780 ↗
  • MalfunctionFeb 27, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THE HIGH FLOW INSUFFLATION UNIT THE CAVITY MODE IS ILLUMINATED AND FLASHES BETWEEN NORMAL AND SMALL. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24469577 ↗
  • MalfunctionFeb 25, 2026Circuit Failure

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HIGH FLOW INSUFFLATION UNIT HAD A CIRCUIT BOARD FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442816 ↗
  • MalfunctionFeb 24, 2026Degraded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HIGH FLOW INSUFFLATION UNIT EXHIBITED A DAMAGED SCREEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24428534 ↗
  • MalfunctionFeb 23, 2026Increase in Pressure

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24419259 ↗
  • MalfunctionFeb 19, 2026Failure to Power Up

    IT WAS REPORTED THE HIGH FLOW INSUFFLATION UNIT WAS NOT TURNING ON. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24388339 ↗
  • MalfunctionFeb 18, 2026Material Integrity Problem; Optical Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HIGH FLOW INSUFFLATION UNIT DISPLAY HAS DIM LIGHT-EMITTING DIODE INDICATORS AND THE MECHANICAL SYSTEM HAS A FAULTY MANIFOLD UNIT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24377199 ↗

Other INSUFFLATOR, AUTOMATIC CARBON-DIOXIDE FOR ENDOSCOPE devices

All INSUFFLATOR, AUTOMATIC CARBON-DIOXIDE FOR ENDOSCOPE devices →