Skip to content

HIGH FLOW INSUFFLATION UNIT recallZ-2776-2026 · Olympus Corporation of the Americas · initiated Jun 16, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Olympus notified consignees on about 06/16/2026 via letter. Consignees were informed that they many continue to use their UHI-4 until the offset adjustment is performed. However, as noted in the Instructions for Use (IFU), prepare a spare UHI-4 unit as a backup to ensure uninterrupted operation can be completed without complication in the case of a malfunction. Actions to take included to carefully read the notification letter and ensure all personnel are completely knowledgeable on the content of the notification, an Olympus field service representative will contact you to schedule a mutually convenient time to perform the offset adjustment, acknowledge receipt of the notification through the Olympus recall web portal, and forward the notification to other users who may have the affected products if you have further distributed it.

Recall details

Recall number
Z-2776-2026
Status
Open, Classified
Initiated by the company
Jun 16, 2026
Posted by the FDA
Jul 22, 2026
Recalling firm
Olympus Corporation of the Americas
Quantity
3,087 units
Product code
HIF
FDA submissions
K243527
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, El Salvador, Germany, Honduras, Hong Kong, India, Ireland, Japan, Korea, Malaysia, Mexico, Myanmar, New Zealand, Paraguay, Peru, Russia, Singapore, Sri Lanka, Taiwan, Thailand, United Kingdom, Uruguay, Vietnam.

Product description

Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4

Code information (models, lots, serial numbers)

Model Number: UHI-4. UDI: 04953170324147, 04953170467219. All Serial Numbers.

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page