HIGH FLOW INSUFFLATION UNIT UHI-4 (REF N3829642)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04953170467219), received between Apr 21, 2025 and Aug 11, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 16, 2026Open, Classified

    Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.

    Z-2776-2026 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Insufficient Flow or Under Infusion2
  • Activation Failure1
  • Circuit Failure1
  • Decrease in Pressure1
  • Excess Flow or Over-Infusion1
  • Gas/Air Leak1
  • Material Split, Cut or Torn1
  • Pressure Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50211216 ↗
  • MalfunctionJun 8, 2026Decrease in Pressure

    THE CUSTOMER REPORTED THE HIGH FLOW INSUFFLATION UNIT WAS UNABLE TO MAINTAIN PRESSURE DURING A THERAPEUTIC LAPAROSCOPIC PROCEDURE. THE EVENT RESULTED IN A DELAY TO THE PROCEDURE OF LESS THAN 30 MINUTES. THE PROCEDURE WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25436369 ↗
  • MalfunctionMay 31, 2026Gas/Air Leak; Pressure Problem

    IT WAS OBSERVED THAT DURING THE EVALUATION, THE HIGH FLOW INSUFFLATION UNIT FAILS SECONDARY PIPING LEAKAGE TEST, BUT A SUBSTITUTE TEST EMERGENCY PRESSURE (EP) VALVE FIXED THE ISSUE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25341132 ↗
  • MalfunctionMay 27, 2026Insufficient Flow or Under Infusion

    THE CUSTOMER REPORTED THE HIGH FLOW INSUFFLATOR FLOW RATE WAS NOT JUMPING BACK UP TO 20 MMHG TO INSUFFLATE. THE ISSUE WAS FOUND DURING A THERAPEUTIC LAPAROSCOPIC POLYPECTOMY PROCEDURE AND RESULTED IN A PROCEDURE PROLONGATION OF GREATER THAN OR EQUAL TO 30 MINUTES. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25306450 ↗
  • MalfunctionJan 23, 2026Excess Flow or Over-Infusion

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24172984 ↗
  • MalfunctionJun 19, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE SUBJECT DEVICE'S GAS LINE EXPLODED/SPLIT OPEN WHEN CONNECTED TO E-TANK. THE ISSUE OCCURRED DURING THERAPEUTIC LAPAROSCOPIC SPAY PROCEDURE THAT WAS COMPLETED WITH A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22292940 ↗
  • MalfunctionMay 27, 2025Circuit Failure; Circuit Failure

    IT WAS REPORTED THAT THE SUBJECT DEVICE EXHIBITED A CIRCUIT BOARD FAILURE, THE ISSUE OCCURRED RUING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22090306 ↗
  • MalfunctionApr 30, 2025Suction Problem; Suction Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED DETERIORATION OF THE PINCH VALVE AND THE FADED TEXTS ON THE REAR PANEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21936847 ↗
  • MalfunctionApr 21, 2025Activation Failure; Activation Failure

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21873915 ↗

Other Insufflator, Automatic Carbon-Dioxide For Endoscope devices

All Insufflator, Automatic Carbon-Dioxide For Endoscope devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.