EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS TGF-UC180J (UDI 04953170328480)Recalls & reported problems
Olympus Medical Systems Corp. · "N4472740 OLA TGF-UC180J FORWARD VIEWING" · Class 2
All models of EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170328480), received between Jan 19, 2023 and Dec 28, 2023 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0345-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Insufficient Device Problem Information1
Patient problems
- Biliary Cirrhosis1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- Thrombocytopenia1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 28, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, THE ULTRASOUND GASTROVIDEOSCOPE HAD A DAMAGED DISTAL TIP AND THE PROBE ITSELF WAS CHIPPED. THE DAMAGE OF THE SCOPE WAS SEEN DURING REGULAR INSPECTION OF EXPIRED SHELF LIFE SCOPES. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18415957 ↗InjuryJan 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) RECEIVED A LITERATURE TITLED, "ENDOSCOPIC ULTRASOUND-GUIDED PORTO-SYSTEMIC PRESSURE GRADIENT MEASUREMENT CORRELATES WITH HISTOLOGICAL HEPATIC FIBROSIS". BACKGROUND AND AIMS: ENDOSCOPIC ULTRASOUND IS A NOVEL DIAGNOSTIC APPROACH TO CHRONIC LIVER DISEASES (CLDS), AND EUS-GUIDED PORTO-SYSTEMIC PRESSURE GRADIENT MEASUREMENT (EUS-PPG) IS AN IMPORTANT EXPANSION WITH A WELL-DEVELOPED TECHNIQUE. HOWEVER, THE CLINICAL VALUE AND APPLICABILITY OF EUS-PPG MEASUREMENT IN PREDICTING HISTOLOGICALLY ADVANCED HEPATIC FIBROSIS REMAIN UNKNOWN. METHODS: THIS WAS A SINGLE-CENTER RETROSPECTIVE STUDY ON PATIENTS WITH VARIOUS CLDS UNDERGOING EUS-PPG AND EUS-GUIDED LIVER BIOPSY (EUS-BX) TO ASSESS IF EUS-PPG MEASUREMENTS CORRELATE WITH HISTOLOGICAL FIBROSIS STAGE AND VARIOUS SURROGATE MARKERS FOR SEVERITY OF CLDS AND ITS SAFETY. CASES WITH EUS-PPG WERE IDENTIFIED AT THE UNIVERSITY OF CALIFORNIA IRVINE, A TERTIARY ENDOSCOPY CENTER, BETWEEN JANUARY 2014 AND MARCH 2020. RESULTS: IN 64 PATIENTS, THE MEAN AGE WAS 57.5; 40 (62.5%), MALES; MEAN CHILD-TURCOTTE-PUGH (CTP) AND MODEL FOR END-STAGE LIVER DISEASE (MELD) SCORES, 5.9 AND 10.4, RESPECTIVELY. THE PROCEDURE SUCCESS RATE WA
Patient problems: Biliary Cirrhosis; Thrombocytopenia
MAUDE report 16212254 ↗InjuryJan 19, 2023Insufficient Device Problem Information
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) RECEIVED A LITERATURE TITLED, "ENDOSCOPIC ULTRASOUND-GUIDED PORTO-SYSTEMIC PRESSURE GRADIENT MEASUREMENT CORRELATES WITH HISTOLOGICAL HEPATIC FIBROSIS." BACKGROUND AND AIMS: ENDOSCOPIC ULTRASOUND IS A NOVEL DIAGNOSTIC APPROACH TO CHRONIC LIVER DISEASES (CLDS), AND EUS-GUIDED PORTO-SYSTEMIC PRESSURE GRADIENT MEASUREMENT (EUS-PPG) IS AN IMPORTANT EXPANSION WITH A WELL-DEVELOPED TECHNIQUE. HOWEVER, THE CLINICAL VALUE AND APPLICABILITY OF EUS-PPG MEASUREMENT IN PREDICTING HISTOLOGICALLY ADVANCED HEPATIC FIBROSIS REMAIN UNKNOWN. METHODS: THIS WAS A SINGLE-CENTER RETROSPECTIVE STUDY ON PATIENTS WITH VARIOUS CLDS UNDERGOING EUS-PPG AND EUS-GUIDED LIVER BIOPSY (EUS-BX) TO ASSESS IF EUS-PPG MEASUREMENTS CORRELATE WITH HISTOLOGICAL FIBROSIS STAGE AND VARIOUS SURROGATE MARKERS FOR SEVERITY OF CLDS AND ITS SAFETY. CASES WITH EUS-PPG WERE IDENTIFIED AT (B)(6), A TERTIARY ENDOSCOPY CENTER, BETWEEN JANUARY 2014 AND MARCH 2020. RESULTS IN 64 PATIENTS, THE MEAN AGE WAS 57.5; 40 (62.5%), MALES; MEAN CHILD-TURCOTTE-PUGH (CTP) AND MODEL FOR END-STAGE LIVER DISEASE (MELD) SCORES, 5.9 AND 10.4, RESPECTIVELY. THE PROCEDURE SUCCESS RATE WAS 100%. TWENTY-NINE (45.3%) HA
Patient problems: Insufficient Information
MAUDE report 16212158 ↗
Other Endoscope, accessories, narrow band spectrum devices
- EVIS EXERA III VIDEO SYSTEM CENTER OLYMPUS CV-190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall7,688 reports
- HD CAMERA HEAD OLYMPUS CH-S190-08-LBOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum3,793 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190LOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum2 recalls2,828 reports
- EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180VOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum6 deaths2,362 reports
- VISERA ELITE VIDEO SYSTEM CENTER OLYMPUS OTV-S190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1,563 reports
- HD AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S190-XZ-EOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1,424 reports
- VIDEO SYSTEM CENTER Type A OLYMPUS CV-170Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum841 reports
- EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180VOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall3 deaths653 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page