GASTROINTESTINAL VIDEOSCOPE GIF-XP170NRecalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of GASTROINTESTINAL VIDEOSCOPE (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus gastrointestinal videoscope has 105 reports, with no deaths and no injuries. Many reports describe the scope failing to clean adequately during reprocessing, and some describe noisy or blank images, dirt on the lens, or corrosion found during evaluation. The reports note no patient involvement. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
105
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 105 adverse event reports that list this exact device (DI 04953170334146), received between Apr 1, 2022 and Feb 12, 2026 — classified as 105 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (78 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 9 reportsJan 2024: 3 reportsFeb 2024: 5 reportsMar 2024: 4 reportsApr 2024: 2 reportsMay 2024: 5 reportsJun 2024: 9 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 4 reportsOct 2024: 4 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 5 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately78
  • Contamination8
  • Poor Quality Image8
  • Device Reprocessing Problem5
  • Contamination /Decontamination Problem2
  • Corroded2
  • Break1
  • Complete Blockage1

Patient problems

  • No Clinical Signs, Symptoms or Conditions105

Recent adverse event reports

The 25 most recent of 105 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2026Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD NOISY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24322113 ↗
  • MalfunctionFeb 10, 2026No Display/Image; Display or Visual Feedback Problem; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED DIRT ON OBJECTIVE LENS AND SHOWED A WHITE SCREEN WITH NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24306132 ↗
  • MalfunctionFeb 4, 2026Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED IMAGE NOISE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24259496 ↗
  • MalfunctionJan 30, 2026Corroded; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD CORRODED LIGHT GUIDE COVER GLASS, DIRTY C-COVER, CHARGED COUPLE DEVICE COVER LENS, AND NOZZLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24222738 ↗
  • MalfunctionJan 30, 2026Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD FOREIGN OBJECTS IN NOZZLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24220834 ↗
  • MalfunctionJan 15, 2026Poor Quality Image

    IT WAS REPORTED THAT THE GASTROINTESTINAL VIDEOSCOPE HAD ANGLE NOISE IMAGE. THE EVENT WAS IDENTIFIED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24078966 ↗
  • MalfunctionJan 14, 2026Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD NOISE IN THE IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24075008 ↗
  • MalfunctionJan 14, 2026Corroded; Overheating of Device

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A BURNED A-RUBBER IN THE INSERTION PART AND CORROSION ON THE CONNECTING TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24071758 ↗
  • MalfunctionDec 15, 2025Poor Quality Image; Poor Quality Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23803101 ↗
  • MalfunctionMay 23, 2025Melted; Melted

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A BURN AT THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22080698 ↗
  • MalfunctionMay 12, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FORCEPS THAT COULDN'T BE INSERTED SMOOTHLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22006028 ↗
  • MalfunctionNov 14, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATTER ATTACHED TO THE SURFACE OF THE TIP, AND A FOREIGN OBJECT ADHERING TO THE AIR AND WATER SUPPLY NOZZLE PIPE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20684659 ↗
  • MalfunctionNov 8, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN OBJECTS INSIDE THE SUCTION CYLINDER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20645158 ↗
  • MalfunctionOct 23, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE AIR/WATER CYLINDER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20516876 ↗
  • MalfunctionOct 22, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE NOZZLE, DISTAL END, AND BENDING SECTION COVER OF THE GASTROINTESTINAL VIDEOSCOPE HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20504455 ↗
  • MalfunctionOct 7, 2024Thermal Decomposition of Device; Failure to Clean Adequately; Thermal Decomposition of Device; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A FOREIGN MATERIAL IN THE NOZZLE AND THE DISTAL END HAD A BURN. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20392742 ↗
  • MalfunctionOct 2, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL EXHIBITED FOREIGN OBJECTS COMING OUT OF THE NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20356369 ↗
  • MalfunctionSep 12, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATTER IN THE NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20205171 ↗
  • MalfunctionSep 10, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL INSIDE THE NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20186796 ↗
  • MalfunctionSep 10, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEOSCOPE EXHIBITED THE NOZZLE, AND THE BENDING SECTION COVER HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20182290 ↗
  • MalfunctionSep 6, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE NOZZLE AND DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20158693 ↗
  • MalfunctionAug 20, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX VIDEO SCOPE EXHIBITED TIP COVER HAD FOREIGN OBJECT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20001458 ↗
  • MalfunctionJul 12, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATTER CAME OUT OF THE AIR/WATER NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19730470 ↗
  • MalfunctionJun 26, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEOSCOPE EXHIBITED FOREIGN BODY IN THE NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19614812 ↗
  • MalfunctionJun 21, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN OBJECT IN NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19580076 ↗

Other Endoscope, accessories, narrow band spectrum devices

All Endoscope, accessories, narrow band spectrum devices →