EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS PCF TYPE Q180AI (UDI 04953170339172)Recalls & reported problems
Olympus Medical Systems Corp. · "PCF-Q180AI EVIS EXERA II COLONOSCOPE" · Class 2
All models of EVIS EXERA II COLONOVIDEOSCOPE (28)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04953170339172), received between Oct 28, 2021 and Oct 31, 2025 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Clean Adequately6
- Device Reprocessing Problem2
- Microbial Contamination of Device2
- Thermal Decomposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 31, 2025Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED THE ETHYLENE OXIDE (GAS) VALVE OF THE COLONOVIDEOSCOPE LEFT OFF. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23439609 ↗MalfunctionOct 15, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE SUBJECT DEVICE EXHIBITED A BURN ON THE DISTAL END COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20453668 ↗MalfunctionOct 14, 2024Microbial Contamination of Device; Microbial Contamination of Device
IT WAS REPORTED THAT THE COLONOVIDEOSCOPE TESTED 3 TIMES POSITIVE FOR PSEUDOMONAS. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20446347 ↗MalfunctionAug 28, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS INSIDE THE AIR/WATER CYLINDER, AIR/WATER TUBE AND JET TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20086320 ↗MalfunctionApr 22, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL FROM THE JET TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19154652 ↗MalfunctionFeb 27, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED THE LIGHT GUIDE AND OBJECTIVE LENSES ADHESIVES HAD FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18792997 ↗MalfunctionFeb 20, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE AIR/WATER TUBE, AIR/WATER CYLINDER, JET-TUBE AND CONNECTING TUBE. IN ADDITION, THE ADHESIVE ON THE BENDING SECTION COVER HAD FOREIGN MATERIAL ATTACHED IN THE EDGES. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18744409 ↗MalfunctionFeb 1, 2024Device Reprocessing Problem; Device Reprocessing Problem
DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE WAS FOUND TO HAVE FOREIGN MATERIAL ON OR IN THE AIR/WATER TUBE AND CYLINDER AND THE AUXILIARY JET CHANNEL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18627317 ↗MalfunctionJan 16, 2024Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS, THAT THE EVIS EXERA II COLONOVIDEOSCOPE HAD A ISSUE WITH THE TIE ROD. IT IS UNKNOWN WHEN THE ISSUE WAS FOUND AND NO PROCEDURE INFORMATION WAS PROVIDED. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE ADHESIVE AROUND THE OBJECTIVE LENS HAD FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18517633 ↗MalfunctionSep 19, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS THAT EVIS EXERA II COLONOVIDEOSCOPE THAT THE BOWDEN CABLE REQUIRED ADJUSTMENT. THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION. DURING EVALUATION, FOREIGN MATERIAL WAS FOUND IN THE AUXILIARY TUBE. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17778126 ↗MalfunctionOct 28, 2021Microbial Contamination of Device; Microbial Contamination of Device
IT WAS REPORTED BY THE CUSTOMER, THE INFECTED (UNWASHED) EVIS EXERA II COLONOVIDEOSCOPE PRESENTED A BLACK STRIP IN THE IMAGE. NO PATIENT HARM REPORTED. OLYMPUS SENT THE DEVICE TO AN INDEPENDENT LABORATORY FOR CULTURE TESTING. THE AUXILIARY WATER CHANNEL TESTED POSITIVE FOR ONE (1) COLONY FORMING UNIT (CFU) OF STAPHYLOCOCCUS WARNERI. THE SCOPE WAS SAMPLED AGAIN WHICH RESULTED IN A NEGATIVE TEST. THIS REPORT IS TO CAPTURE THE REPORTABLE MALFUNCTION OF THE POSITIVE CULTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12712463 ↗
Other Endoscope, accessories, narrow band spectrum devices
- EVIS EXERA III VIDEO SYSTEM CENTER OLYMPUS CV-190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall7,688 reports
- HD CAMERA HEAD OLYMPUS CH-S190-08-LBOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum3,793 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190LOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum2 recalls2,828 reports
- EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180VOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum6 deaths2,362 reports
- VISERA ELITE VIDEO SYSTEM CENTER OLYMPUS OTV-S190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1,563 reports
- HD AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S190-XZ-EOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1,424 reports
- VIDEO SYSTEM CENTER Type A OLYMPUS CV-170Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum841 reports
- EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180VOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall3 deaths653 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page