EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS PCF TYPE H180AI (UDI 04953170339875)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of EVIS EXERA II COLONOVIDEOSCOPE (28)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 04953170339875), received between Sep 26, 2019 and May 11, 2026 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0341-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Clean Adequately7
- Microbial Contamination of Device6
- Contamination /Decontamination Problem1
- Device Reprocessing Problem1
- Poor Quality Image1
- Residue After Decontamination1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2026Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE COLONOVIDEOSCOPE SCOPE COVER WAS DIRTY, THE AIR WATER CYLINDER CONTAINED FOREIGN OBJECTS, AND THE AIR WATER TUBE CONTAINED FOREIGN OBJECTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25135570 ↗MalfunctionMar 3, 2026Contamination /Decontamination Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL ON AIR WATER CYLINDER / TUBE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24501173 ↗MalfunctionDec 16, 2025Poor Quality Image; Poor Quality Image
IT WAS REPORTED THAT THE COLONOVIDEOSCOPE EXHIBITED LINES ON THE MONITOR. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23821113 ↗MalfunctionApr 23, 2025Residue After Decontamination; Residue After Decontamination
IT WAS REPORTED THAT DURING SET UP THE COLONOVIDEOSCOPE HAD FOREIGN MATERIAL (RESIDUES) IN THE CHANNEL. THERE WAS NO REPORT OF PATIENT HARM OR USER INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21891679 ↗MalfunctionMar 20, 2025Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED THE AIR/WATER-CYLINDER (TUBE) HAS FOREIGN OBJECTS, THE JET TUBE HAS FOREIGN OBJECTS, THE AIR/WATER TUBE HAS FOREIGN OBJECTS, THE PLASTIC DISTAL END (C-COVER) HAS FOREIGN OBJECTS, THE NOZZLE HAS FOREIGN OBJECTS, THE ADHESIVE AROUND THE OBJECTIVE LENS HAS FOREIGN OBJECTS, THE BENDING RUBBER (A-RUBBER) HAS FOREIGN OBJECTS, AND THE CONNECTING TUBE HAS FOREIGN OBJECTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21651722 ↗MalfunctionMar 13, 2025Device Reprocessing Problem; Device Reprocessing Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE AIR/WATER TUBE AND CYLINDER HAD FOREIGN OBJECTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21600211 ↗MalfunctionOct 15, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED WHITE FOREIGN MATERIAL IN THE AIR/WATER TUBE AND AIR/WATER CHANNEL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20456945 ↗MalfunctionApr 30, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE HAD WHITE FOREIGN MATERIAL IN THE AIR/WATER CYLINDER AND TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19213779 ↗MalfunctionApr 8, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE AW (AIR, WATER) CYLINDER AND TUBE HAD WHITE FOREIGN MATERIAL/FOREIGN OBJECTS. AND THE JET TUBE HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19068719 ↗MalfunctionMar 4, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN THE AIR/WATER CYLINDER AND AIR/WATER TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18832585 ↗MalfunctionAug 8, 2023Failure to Clean Adequately; Failure to Clean Adequately
IT WAS REPORTED THAT THE LOANER COLONOVIDEOSCOPE CONNECTION WAS DEFECTIVE DUE TO INCORRECT SCREWING. THERE WAS NO REPORT OF PATIENT HARM. THE DEVICE WAS RETURNED, EVALUATED, AND A WHITISH FOREIGN MATERIAL WAS FOUND IN THE AIR/WATER TUBE AND AIR/WATER CYLINDER. THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17485185 ↗MalfunctionOct 4, 2022Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS THAT DURING ROUTINE TESTING, THE ENDOSCOPE TESTED POSITIVE FOR MICROBIAL CONTAMINATION. ON (B)(6) 2022, ALL CHANNELS WERE SAMPLED AND TESTED POSITIVE FOR GREATER THAN 25 COLONY FORMING UNITS (CFU) OF PSEUDOMONAS AERUGINOSA. ON (B)(6) 2022, ALL CHANNELS WERE SAMPLED AND TESTED POSITIVE FOR GREATER THAN 25 COLONY FORMING UNITS (CFU) OF GERMS (RAOULTELLA PLANTICOLA: 8 CFU, CITROBACTER FREUNDII: 1 CFU, AND PSEUDOMONAS AERUGINOSA: 14 CFU). THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN THE STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15538053 ↗MalfunctionDec 13, 2021Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS FRANCE, DURING ROUTINE REPROCESSING, THE SUBJECT DEVICE WAS TESTED AND FOUND TO HAVE MICROBIAL CONTAMINATION. THE TEST RESULTS INCLUDED: SUCTION CHANNEL WITH 3 COLONY FORMING UNITS OF PSEUDOMONAS CHLORORAPHIS; AIR/WATER CHANNEL WITH 2 COLONY FORMING UNITS OF PSEUDOMONAS CHLORORAPHIS. THE CUSTOMER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THIS MEDICAL DEVICE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12984666 ↗MalfunctionOct 23, 2020Microbial Contamination of Device; Microbial Contamination of Device
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MULTIPLE MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. FIRST TIME; SEPT. 21ST, 2020; AIR/WATER CHANNEL: UNSPECIFIED MICROBES (15 CFU/ENDOSCOPE), INSTRUMENT CHANNEL: UNSPECIFIED MICROBES (>25 CFU/ENDOSCOPE), SECOND TIME; SEPT. 21ST, 2020; AIR/WATER CHANNEL: UNSPECIFIED MICROBES (7 CFU/ENDOSCOPE), INSTRUMENT CHANNEL: UNSPECIFIED MICROBES (>25 CFU/ENDOSCOPE). THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, POKA YOKE (GETINGE), USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10724813 ↗MalfunctionMay 28, 2020Microbial Contamination of Device; Microbial Contamination of Device
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MULTIPLE MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. [FIRST TIME; (B)(6) 2020] STENOTROPHOMONAS MALTOPHILIA (103 CFU/150ML) [SECOND TIME; (B)(6) 2020] POLYMORPHIC FLORA (COUNTLESS BACTERIA CFU/150ML) THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, SOLUSCOPE SERIE4, USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10099541 ↗MalfunctionNov 17, 2019Microbial Contamination of Device; Microbial Contamination of Device
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF ROUTINE MICROBIOLOGICAL TESTING BY THE USER FACILITY ON THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE TESTED POSITIVE FOR ENTEROBACTERIACEAE AND PSEUDOMONAS AERUGINOSA (GREATER THAN 100 CFU). THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, SOLUSCOPE 4, USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9333691 ↗MalfunctionSep 26, 2019Microbial Contamination of Device; Microbial Contamination of Device
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MULTIPLE MICROBIOLOGICAL TESTING BY THE USER FACILITY, THE FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. -SUCTION CHANNEL AND INSTRUMENT CHANNEL: SERRATIA MARCESCENS (2 CFU/25ML). -INSTRUMENT CHANNEL: SERRATIA MARCESCENS (11 CFU/50ML) AND PSEUDOMONAS AERUGINOSA (2 CFU/50ML). THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, SOLUSCOPE 3 AND SOLUSCOPE 4, AND SOMETIMES MANUALLY REPROCESSED USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9126718 ↗
Other Gastroscope and accessories, flexible/rigid devices
- PENTAX Medical EG29-i10cHoya Corporation · Gastroscope And Accessories, Flexible/Rigid1,623 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,025 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall559 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GIF-Q165Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid349 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 death302 reports
- PENTAX EG27-i10Hoya Corporation · Gastroscope and accessories, flexible/rigid261 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 deaths252 reports
- BALLOON MAJ-1351Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid174 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page