Skip to content

Single Use Aspiration Needle NA-U401SX NA-U401SX-4022 (UDI 04953170388286)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of Single Use Aspiration Needle NA-U401SX (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus single-use aspiration needle is used to take samples with a biopsy needle. The 30 reports mostly describe the needle breaking, snapping off, or bending during use, along with sheath damage and punctured packaging. There are 3 injuries and no deaths. One recall, started in March 2023 and still open, concerns needles whose tip angle does not match manufacturing specifications, causing slight resistance when the needle is pushed in or pulled back.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 04953170388286), received between May 25, 2018 and Aug 19, 2026 — classified as 3 injuries, 26 malfunctions.

The most frequently reported specific device problem is “Break” (14 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K160098); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 31, 2023Class IIIOpen, Classified

    Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction.

    Z-1616-2023 · Root cause: Employee error · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break14
  • Detachment of Device or Device Component3
  • Packaging Problem3
  • Difficult to Advance2
  • Material Twisted/Bent2
  • Mechanical Problem2
  • Component Missing1
  • Contamination /Decontamination Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions20
  • No Consequences Or Impact To Patient4
  • No Known Impact Or Consequence To Patient2
  • Cough1
  • Exposure to Body Fluids1
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 19, 2026Device Damaged by Another Device; Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50324955 ↗
  • MalfunctionApr 28, 2026Material Twisted/Bent

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SINGLE USE ASPIRATION NEEDLE'S SHEATH WAS BUCKLED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25018964 ↗
  • MalfunctionApr 17, 2026Fracture

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SINGLE USE ASPIRATION NEEDLE'S INSERTION PORTION WAS SEVERED APPROXIMATELY 3 CM UNDER THE BOOT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24910743 ↗
  • MalfunctionFeb 24, 2026Packaging Problem

    NO ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24430419 ↗
  • MalfunctionFeb 24, 2026Packaging Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24430311 ↗
  • MalfunctionFeb 24, 2026Packaging Problem

    IT WAS REPORTED THE ASPIRATION NEEDLE HAD PUNCTURED THE PACKAGING SEAL. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24429992 ↗
  • MalfunctionJan 12, 2026Break

    IT WAS REPORTED THE SINGLE-USE ASPIRATION NEEDLE APPEARED TO HAVE "ERODED" AROUND THE DISTAL TIP OF THE GREEN CATHETER AFTER "A COUPLE OF PASSES". THIS WAS NOTED DURING AN ENDOBRONCHIAL ULTRASOUND PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A DEVICE OF A DIFFERENT MAKE/MANUFACTURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24037580 ↗
  • MalfunctionNov 19, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SINGLE USE ASPIRATION NEEDLE, SHEATH WAS BUCKLING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23598601 ↗
  • InjuryMay 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED DURING A THERAPEUTIC ENDOBRONCHIAL ULTRASOUND (EBUS) PROCEDURE, THE METAL SHEATH OF THE SINGLE USE ASPIRATION NEEDLE CAME LOOSE AND FELL INTO THE PATIENT. THE FALLEN PIECE WAS RETRIEVED FROM THE PATIENT WITHOUT ANY COMPLICATIONS. THERE WAS A PROCEDURAL DELAY OF LESS THAN 30 MINUTES, AND THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22105374 ↗
  • MalfunctionMay 20, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED THE SINGLE USE ASPIRATION NEEDLE LOST CONTROL OF ITS FORWARD AND BACKWARD MOVEMENT. THE ISSUE OCCURRED BEFORE MEDIASTINAL TUMOR BIOPSY/BRONCHOSCOPY WITH TRANSBRONCHIAL BIOPSY BEFORE THE PATIENT WAS UNDER SEDATION. THERE WAS A DELAY OF AN UNKNOWN TIME PERIOD. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: Insufficient Information

    MAUDE report 22056133 ↗
  • MalfunctionNov 6, 2024Failure to Advance; Failure to Advance

    IT WAS REPORTED THAT THE SINGLE USE ASPIRATION NEEDLE'S STYLET COULD NOT BE INSERTED. THE ISSUE OCCURRED DURING EBUS-TBNA (ENDOBRONCHIAL ULTRASOUND-GUIDED TRANSBRONCHIAL NEEDLE ASPIRATION) PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20622545 ↗
  • InjuryFeb 2, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ENDOBRONCHIAL ULTRASOUND BRONCHOSCOPY THE NEEDLE BROKE OFF INSIDE THE PATIENT. THE PHYSICIAN WAS ABLE TO RETRIEVE IT USING AN ADULT BRONCHOSCOPE AND BIOPSY FORCEPS, AVOIDING INJURY WHILE DOING SO. THE PROCEDURE WAS COMPLETED BUT NOT WITH THE SAME DEVICE. THE DEVICE ISSUE DID NOT IMPACT THE OUTCOME OF THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18636724 ↗
  • MalfunctionOct 16, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, THE SINGLE USE ASPIRATION NEEDLE¿S TIP WAS BROKEN OFF. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17941706 ↗
  • MalfunctionOct 13, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, THE SINGLE USE ASPIRATION NEEDLE¿S TIP WAS BROKEN OFF. THERE WERE NO REPORTS OF PATIENT HARM. ASSOCIATED PATIENT IDENTIFIERS: (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17936304 ↗
  • MalfunctionOct 13, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, THE SINGLE USE ASPIRATION NEEDLE¿S TIP WAS BROKEN OFF. THERE WERE NO REPORTS OF PATIENT HARM. ASSOCIATED PATIENT IDENTIFIERS: (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17936302 ↗
  • MalfunctionOct 5, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, THE SINGLE USE ASPIRATION NEEDLE TIP BROKE OFF. THERE WERE NO REPORTS OF PATIENT HARM. ASSOCIATED PATIENT IDENTIFIERS: (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17874327 ↗
  • MalfunctionSep 21, 2023Break; Break

    THE CUSTOMER REPORTED THE SINGLE USE ASPIRATION NEEDLE BROKE OFF IN THE HANDLE DURING THE PROCEDURE. THE NEEDLE COULD NOT BE REMOVED WITHOUT HARMING THE PATIENT OR SCOPE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17792308 ↗
  • MalfunctionAug 15, 2023Component Missing; Tear, Rip or Hole in Device Packaging; Component Missing; Tear, Rip or Hole in Device Packaging

    THE CUSTOMER REPORTED TO OLYMPUS, THE SINGLE USE ASPIRATION NEEDLE WAS MISSING FROM THE ITEM¿S PACKAGE DUE TO A HOLE, OR IRREGULARITY FOUND IN THE SEALED STERILE PACKAGE. THE ISSUE WAS FOUND DURING AN UNKNOWN PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17553632 ↗
  • MalfunctionFeb 7, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AN OLYMPUS EMPLOYEE REPORTED THAT, WHEN SETTING UP FOR A CASE, THE CUSTOMER PICKED UP HER DEMONSTRATION DEVICE AND USED IT DURING THE UNSPECIFIED DIAGNOSTIC PROCEDURE. THE CUSTOMER ASKED THE EMPLOYEE TO COME INTO THE FACILITY TO SUPPORT A PROCEDURE AS IT HAD BEEN A WHILE SINCE THEY PERFORMED THIS PROCEDURE. WHEN SHE ARRIVED, SHE BROUGHT IN HER DEMONSTRATION 22G NEEDLE THAT WAS OPENED AND HAD BEEN USED IN A DEMONSTRATION PREVIOUSLY AND TOUCHED BY MULTIPLE PEOPLE. THE SURGICAL ROOM WAS NOT PREPARED AND SHE WAS ASKED TO HELP SET UP THE INSTRUMENTS. SHE SET DOWN HER DEMONSTRATION NEEDLE IN AN AREA OUTSIDE THE STERILE AREA IN THE SURGICAL ROOM. THE CUSTOMER REQUESTED HER HELP TO TRACK DOWN INSTRUMENTS NEEDED IN THE PROCEDURE. WHEN SHE RETURNED SHE WAS TOLD THAT THE DOCTOR WANTED A 19G NEEDLE BUT ONE COULD NOT BE LOCATED. WHEN THE EMPLOYEE CAME BACK FROM LOOKING FOR A 19G NEEDLE, THE DOCTOR WAS USING A NEEDLE IN THE CASE. IT WAS NOT UNTIL LATER THAT SHE REALIZED HER DEMONSTRATION NEEDLE WAS NOT WHERE SHE LEFT IT AND HAD BEEN USED BY THE DOCTOR IN THE PROCEDURE. SHE INFORMED THE DOCTOR THAT THE NEEDLE WAS OPENED AND USED IN OTHER DEMONSTRATIONS AND THAT OTHER DOCTORS HAD TOUCHED THE DEVIC

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16312845 ↗
  • MalfunctionFeb 26, 2022Mechanical Problem; Mechanical Problem

    AS REPORTED, CUSTOMER REPORTED OF "DIFFICULTY IN FIXATION OF THE NEEDLE" (IT MOVES EASILY EVEN WITH THE MAXIMUM FIXATION). VERY DIFFICULT TO PENETRATE THE TRACHEAL WALL, LYMPH NODE STRUCTURE AND IN THE BACK AND FORTH MOVEMENTS. DEFORMATION OF THE NEEDLE WITH THE USE. VERY POOR ULTRASOUND VISUALIZATION. THE EVENT OCCURRED DURING A PROCEDURE (ULTRASOUND GUIDED PUNCTURE). NO PATIENT DAMAGED. NO EXTENSION OF THE PROCEDURE AS A RESULT OF THE EVENT. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13611190 ↗
  • InjuryJan 19, 2022Break; Difficult to Advance; Break; Difficult to Advance

    THE CUSTOMER REPORTED TO OLYMPUS, THE SUBJECT DEVICE BROKE DURING A DIAGNOSTIC ENDOBRONCHIAL ULTRASOUND TRANSBRONCHIAL NEEDLE ASPIRATION. THE PHYSICIAN HAD DIFFICULTY DEPLOYING THE NEEDLE INTO A LYMPH NODE. THEN, THE SHEATH WAS RETRACTED INTO THE BRONCHOSCOPE AND THE STAFF NOTICED THE TIP OF THE NEEDLE WAS MISSING. THE INTENDED PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. ANOTHER BRONCHOSCOPY AND X-DAY DID NOT REVEAL THE NEEDED TIP INSIDE OF THE PATIENT. THE PATIENT DID NOT HAVE AN EXTENDED HOSPITAL STAY PER THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13288861 ↗
  • MalfunctionJul 15, 2021Break; Break

    OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED BY A CLINICAL STAFF THAT DURING A PROCEDURE OF AN ENDOSCOPIC BRONCHIAL ULTRASOUND USING THE SUBJECT DEVICE AS FOLLOWS. ¿WE HAVE AN EBUS NEEDLE THAT DURING A PROCEDURE COULD NOT RETRACT AND WHEN IT WAS RETRACTED THE NEEDLE HAD BROKEN OFF. NEEDLE FAILED TO RETRACT DURING PROCEDURE, NEEDLE REMOVED FROM PATIENT AND SCOPE AND FOUND PIECE OF NEEDLE MISSING DURING A BRONCH PROCEDURE THE PATIENT GETS SPRAYED WITH NOVOCAINE SPRAY TO NUMB THE VOCAL CORDS AND USUALLY HAVE NO FEELING IS SURROUNDING AREA. COUGHING IS MOSTLY A REFLEX REACTION. THEY WERE LOOKING FOR THE NEEDLE IN THE LUNGS WITH A SCOPE, WHEN THEY RETRIEVED THE SCOPE SHE COUGHED AND THEN SWALLOWED. THEY ASSUME THAT¿S WHEN SHE SWALLOWED IT. SHE WAS THEN SENT TO CT AND A GASTROSCOPY, THEY DID NOT FIND IT IN THE STOMACH. THE PATIENT UNDERWENT A CT SCAN ON SATURDAY MORNING AT THAT STAGE THE NEEDLE TEST WAS IN THE SIGMOID.¿ ALSO, IT WAS INFORMED BY A NURSE AS FOLLOWS. ¿WEDNESDAY 23 JUNE - NEEDLE BROKE OFF IN PATIENT. THURSDAY 24 JUNE - OLYMPUS REPRESENTATIVE COLLECTED DEFECTIVE PRODUCT (WE WERE TOLD AT THE TIME THAT THE PATIENT WAS ADMITTED AND UNDER OBSERVATION IN HDU). GASTROSCOPY WAS

    Patient problems: Cough; Foreign Body In Patient

    MAUDE report 12178866 ↗
  • MalfunctionJan 27, 2021Mechanical Problem; Mechanical Problem

    IT IS REPORTED A SINGLE USE ASPIRATION NEEDLE WOULD NOT LOCK. THE SHEATH WOULD NOT STAY LOCKED AND THE NEEDLE CAME OUT PUNCTURING THE INSERTION TUBE OF THE SCOPE. THERE WAS NO PATIENT IMPACT RELATED TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11237410 ↗
  • MalfunctionJan 5, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE SERVICE CENT WAS INFORMED THAT DURING A PROCEDURE, WHILE EXPELLING THE SPECIMEN FROM THE TIP OF THE NEEDLE THE CUSTOMER EXPERIENCED DIFFICULTY AND THE PATHOLOGIST WAS SPRAYED TWICE. THE CUSTOMER FURTHER INVESTIGATED THE NEEDLE AND NOTICED THE SPRAY WAS COMING FROM THE GREEN COVERING THAT MEETS THE COIL SHEATH AT THE CONNECTOR. THE REPORTED THAT WHEN EXPELLING THE SPECIMEN THE NEEDLE WAS NOT EXTEND OUT OF THE SHEATH, WHICH MAY HAVE CAUSED THE PROBLEM DUE TO THE FACT THE SPECIMEN WAS VERY LARGE AND CLOGGED THE TIP. THE CUSTOMER BELIEVES THIS MAY HAVE BEEN THE REASON FOR SPECIMEN BACKING UP IN THE SHEATH. IT IS UNKNOWN IF THE INTENDED PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT/USER INJURY OR INFECTION REPORTED.

    Patient problems: Exposure to Body Fluids

    MAUDE report 11118010 ↗
  • MalfunctionDec 6, 2019Break; Break

    DURING AN ENDOBRONCHIAL ULTRASONOGRAPHY, THE SUBJECT DEVICE WAS USED. IN THE PROCEDURE, THE USER FOUND THAT THE TUBE SHEATH WAS BROKEN. AND THE USER COULD NOT RETRACT THE NEEDLE INTO THE TUBE SHEATH. THE USER COULD NOT TAKE OUT THE SUBJECT DEVICE FROM THE ENDOSCOPE. THE USER TOOK OUT THE ENDOSCOPE FROM THE PATIENT WHILE THE SUBJECT DEVICE WITH THE EXTENDED NEEDLE LEFT IN THE ENDOSCOPE. NO FURTHER INFORMATION WAS PROVIDED. THIS IS THE REPORT REGARDING THE INABILITY TO RETRACT THE NEEDLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9421324 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160098. One submission often covers many models, so check the product and lot details in each record.

  • Z-1615-2023 · Mar 31, 2023 · Open, Classified
    ViziShot 2 Single Use Aspiration Needle, 21G- To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the t

Other Biopsy needle devices

All Biopsy needle devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page