Skip to content

Interject M00518250Recalls & reported problems

Boston Scientific Corporation · Injection Therapy Needle Catheter · Class 2

Quick summary

This Boston Scientific Interject injection therapy needle catheter is used to inject fluid or medication through a scope during procedures such as colonoscopies and gastroscopies. Most of the 70 reports describe the needle's channel becoming blocked so fluid or medication could not be pushed through, and a few describe the needle detaching from its sheath. No deaths or injuries were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
70
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 08714729415091), received between Aug 2, 2017 and May 27, 2025 — classified as 70 malfunctions.

The most frequently reported specific device problem is “Occlusion Within Device” (69 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Occlusion Within Device69
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1

Patient problems

  • No Consequences Or Impact To Patient63
  • No Known Impact Or Consequence To Patient13
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2025Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT NEEDLE WAS USED IN THE POLYPS DURING A POLYPECTOMY PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE NEEDLE DETACHED FROM THE SHEATH. ANOTHER INTERJECT NEEDLE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22088046 ↗
  • MalfunctionJun 26, 2018Occlusion Within Device; Occlusion Within Device

    AN INTERJECT NEEDLE WAS RECEIVED FOR ANALYSIS. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS HAS BEEN MADE A REPORTABLE EVENT DUE TO INVESTIGATION RESULTS OF NEEDLE -BLOCKED/OCCLUDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7640160 ↗
  • MalfunctionJun 26, 2018Occlusion Within Device; Occlusion Within Device

    AN INTERJECT NEEDLE WAS RECEIVED FOR ANALYSIS. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS HAS BEEN MADE A REPORTABLE EVENT DUE TO INVESTIGATION RESULTS OF NEEDLE -BLOCKED/OCCLUDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7640144 ↗
  • MalfunctionJun 26, 2018Occlusion Within Device; Occlusion Within Device

    AN INTERJECT NEEDLE WAS RECEIVED FOR ANALYSIS. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS HAS BEEN MADE A REPORTABLE EVENT DUE TO INVESTIGATION RESULTS OF NEEDLE -BLOCKED/OCCLUDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7640136 ↗
  • MalfunctionJun 26, 2018Occlusion Within Device; Occlusion Within Device

    AN INTERJECT NEEDLE WAS RECEIVED FOR ANALYSIS. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS HAS BEEN MADE A REPORTABLE EVENT DUE TO INVESTIGATION RESULTS OF NEEDLE -BLOCKED/OCCLUDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7640107 ↗
  • MalfunctionJun 26, 2018Occlusion Within Device; Occlusion Within Device

    AN INTERJECT NEEDLE WAS RECEIVED FOR ANALYSIS. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS HAS BEEN MADE A REPORTABLE EVENT DUE TO INVESTIGATION RESULTS OF NEEDLE -BLOCKED/OCCLUDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7640044 ↗
  • MalfunctionJun 22, 2018Occlusion Within Device; Occlusion Within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ NEEDLE WAS USED IN THE COLON DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE NEEDLE WAS UNABLE TO FLUSH THE SODIUM CHLORIDE THROUGH THE WORKING CHANNEL. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7628809 ↗
  • MalfunctionJun 22, 2018Occlusion Within Device; Occlusion Within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ NEEDLE WAS USED IN THE ESOPHAGUS DURING A GASTROSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CHANNEL WAS BLOCKED AND NO ADRENALINE COULD BE PUSHED THROUGH. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7628782 ↗
  • MalfunctionJun 18, 2018Occlusion Within Device; Occlusion Within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ SCLEROTHERAPY NEEDLE WAS USED IN THE COLON DURING AN INJECTION PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE THE NEEDLE WAS OCCLUDED. REPORTEDLY THE SUBSTANCE THEY WERE TRYING TO INJECT WAS SODIUM CHLORIDE. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ SCLEROTHERAPY NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7609510 ↗
  • MalfunctionJun 18, 2018Occlusion Within Device; Occlusion Within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A INTERJECT SCLEROTHERAPY NEEDLE WAS USED IN THE COLON DURING AN INJECTION PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE IT WAS IMPOSSIBLE TO INJECT ANY LIQUID. REPORTEDLY THE SUBSTANCE THEY WERE TRYING TO INJECT WAS SODIUM CHLORIDE. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT SCLEROTHERAPY NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7609507 ↗
  • MalfunctionJun 15, 2018Occlusion Within Device; Occlusion Within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ SCLEROTHERAPY NEEDLE WAS USED IN THE COLON DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE THE NEEDLE WAS UNABLE TO FLUSH THE LIQUID OUT. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ SCLEROTHERAPY NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7608009 ↗
  • MalfunctionJun 15, 2018Occlusion Within Device; Occlusion Within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ SCLEROTHERAPY NEEDLE WAS USED IN THE COLON DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE IT WAS IMPOSSIBLE TO INJECT ANY FLUID. REPORTEDLY THE SUBSTANCE THEY WERE TRYING TO INJECT WAS SODIUM CHLORIDE. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ SCLEROTHERAPY NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7607979 ↗
  • MalfunctionJun 11, 2018Occlusion Within Device; Occlusion Within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ NEEDLE WAS USED DURING A GASTROSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, AT THE TIME OF THE INJECTION, THE LIQUID DID NOT LEAVE THE NEEDLE. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7588189 ↗
  • MalfunctionJun 7, 2018Occlusion Within Device; Occlusion Within Device

    NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT# 3005099803-2018-01827 FOR THE FIRST INTERJECT¿ SCLEROTHERAPY NEEDLE AND 3005099803-2018-01828 FOR THE SECOND INTERJECT¿ SCLEROTHERAPY NEEDLE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ NEEDLE WAS USED IN THE STOMACH DURING A GASTROSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, AT THE TIME OF INJECTION THE LIQUID DID NOT LEAVE THE NEEDLE. A SECOND INTERJECT ¿ NEEDLE WAS USED, AND HAD THE SAME ISSUE OF LIQUID NOT LEAVING THE NEEDLE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7578758 ↗
  • MalfunctionJun 7, 2018Occlusion Within Device; Occlusion Within Device

    NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT# 3005099803-2018-01827 FOR THE FIRST INTERJECT¿ SCLEROTHERAPY NEEDLE AND 3005099803-2018-01828 FOR THE SECOND INTERJECT¿ SCLEROTHERAPY NEEDLE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ NEEDLE WAS USED IN THE STOMACH DURING A GASTROSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, AT THE TIME OF INJECTION THE LIQUID DID NOT LEAVE THE NEEDLE. A SECOND INTERJECT ¿ NEEDLE WAS USED, AND HAD THE SAME ISSUE OF LIQUID NOT LEAVING THE NEEDLE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7578740 ↗
  • MalfunctionJun 6, 2018Occlusion Within Device; Occlusion Within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ NEEDLE WAS USED IN THE STOMACH DURING A GASTROSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, WHEN TRYING TO INJECT, THE LIQUID DID NOT LEAVE THE NEEDLE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7575939 ↗
  • MalfunctionJun 1, 2018Occlusion Within Device; Occlusion Within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ NEEDLE WAS USED IN THE COLON DURING AN ENDOSCOPIC MUCOSAL RESECTION (EMR) PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, IT WAS IMPOSSIBLE TO INJECT THROUGH THE NEEDLE. IT WAS REPORTED TO BE COMPLETELY BLOCKED. IT WAS REPORTED THAT EVEN WATER COULD NOT BE FLUSHED THROUGH THE NEEDLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE "OKAY".

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7563329 ↗
  • MalfunctionJun 1, 2018Occlusion Within Device; Occlusion Within Device

    NOTE: THIS REPORT PERTAINS TO ONE OF SIX DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT# 3005099803-2018-01696 FOR THE FIRST INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01697 FOR THE SECOND INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01698 FOR THE THIRD INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01699 FOR THE FOURTH INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01700 FOR THE FIFTH INTERJECT¿ SCLEROTHERAPY NEEDLE AND 3005099803-2018-01701 FOR THE SIXTH INTERJECT¿ SCLEROTHERAPY NEEDLE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ SCLEROTHERAPY NEEDLE WAS USED IN THE COLON DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE IT WAS IMPOSSIBLE TO INJECT THE PRODUCT INTO THE WORKING CHANNEL. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7563167 ↗
  • MalfunctionJun 1, 2018Occlusion Within Device; Occlusion Within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ NEEDLE WAS USED IN THE COLON DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS IMPOSSIBLE TO INJECT THE PRODUCT INTO THE WORKING CHANNEL. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7563160 ↗
  • MalfunctionJun 1, 2018Occlusion Within Device; Occlusion Within Device

    NOTE: THIS REPORT PERTAINS TO ONE OF SIX DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT# 3005099803-2018-01696 FOR THE FIRST INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01697 FOR THE SECOND INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01698 FOR THE THIRD INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01699 FOR THE FOURTH INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01700 FOR THE FIFTH INTERJECT¿ SCLEROTHERAPY NEEDLE AND 3005099803-2018-01701 FOR THE SIXTH INTERJECT¿ SCLEROTHERAPY NEEDLE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ SCLEROTHERAPY NEEDLE WAS USED IN THE COLON DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE IT WAS IMPOSSIBLE TO INJECT THE PRODUCT INTO THE WORKING CHANNEL. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7563153 ↗
  • MalfunctionJun 1, 2018Occlusion Within Device; Occlusion Within Device

    THIS REPORT PERTAINS TO ONE OF SIX DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT# 3005099803-2018-01696 FOR THE FIRST INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01697 FOR THE SECOND INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01698 FOR THE THIRD INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01699 FOR THE FOURTH INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01700 FOR THE FIFTH INTERJECT¿ SCLEROTHERAPY NEEDLE AND 3005099803-2018-01701 FOR THE SIXTH INTERJECT¿ SCLEROTHERAPY NEEDLE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ SCLEROTHERAPY NEEDLE WAS USED IN THE COLON DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE IT WAS IMPOSSIBLE TO INJECT THE PRODUCT INTO THE WORKING CHANNEL. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7563117 ↗
  • MalfunctionJun 1, 2018Occlusion Within Device; Occlusion Within Device

    NOTE: THIS REPORT PERTAINS TO ONE OF SIX DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT# 3005099803-2018-01696 FOR THE FIRST INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01697 FOR THE SECOND INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01698 FOR THE THIRD INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01699 FOR THE FOURTH INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01700 FOR THE FIFTH INTERJECT¿ SCLEROTHERAPY NEEDLE AND 3005099803-2018-01701 FOR THE SIXTH INTERJECT¿ SCLEROTHERAPY NEEDLE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ SCLEROTHERAPY NEEDLE WAS USED IN THE COLON DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE IT WAS IMPOSSIBLE TO INJECT THE PRODUCT INTO THE WORKING CHANNEL. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7563068 ↗
  • MalfunctionJun 1, 2018Occlusion Within Device; Occlusion Within Device

    NOTE: THIS REPORT PERTAINS TO ONE OF SIX DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT# 3005099803-2018-01696 FOR THE FIRST INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01697 FOR THE SECOND INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01698 FOR THE THIRD INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01699 FOR THE FOURTH INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01700 FOR THE FIFTH INTERJECT¿ SCLEROTHERAPY NEEDLE AND 3005099803-2018-01701 FOR THE SIXTH INTERJECT¿ SCLEROTHERAPY NEEDLE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ SCLEROTHERAPY NEEDLE WAS USED IN THE COLON DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE IT WAS IMPOSSIBLE TO INJECT THE PRODUCT INTO THE WORKING CHANNEL. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7563000 ↗
  • MalfunctionJun 1, 2018Occlusion Within Device; Occlusion Within Device

    NOTE: THIS REPORT PERTAINS TO ONE OF SIX DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT# 3005099803-2018-01696 FOR THE FIRST INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01697 FOR THE SECOND INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01698 FOR THE THIRD INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01699 FOR THE FOURTH INTERJECT¿ SCLEROTHERAPY NEEDLE, 3005099803-2018-01700 FOR THE FIFTH INTERJECT¿ SCLEROTHERAPY NEEDLE AND 3005099803-2018-01701 FOR THE SIXTH INTERJECT¿ SCLEROTHERAPY NEEDLE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ SCLEROTHERAPY NEEDLE WAS USED IN THE COLON DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE IT WAS IMPOSSIBLE TO INJECT THE PRODUCT INTO THE WORKING CHANNEL. THE PROCEDURE WAS COMPLETED WITH ANOTHER INTERJECT¿ NEEDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7562988 ↗
  • MalfunctionMay 25, 2018Occlusion Within Device; Occlusion Within Device

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INTERJECT¿ NEEDLE WAS TO BE USED IN A POLYP INFILTRATION AND ULCER INJECTION THERAPY PROCEDURE PERFORMED IN THE ESOPHAGUS AND STOMACH. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, THE NEEDLE WAS NOTICED TO BE OCCLUDED. REPORTEDLY THE SUBSTANCE THEY WERE TRYING TO INJECT WAS ADRENALINE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7545467 ↗

Other Biopsy needle devices

All Biopsy needle devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.