Quick summary

This Olympus THUNDERBEAT Open Fine Jaw Type X is an ultrasonic surgical instrument. The 26 reports mostly describe the tissue pad peeling, wearing away, or going missing, along with parts loosening or coming apart; one report mentions the handpiece disassembling during surgery. There were 2 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
26
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 04953170394584), received between Sep 26, 2022 and Aug 6, 2026 — classified as 2 injuries, 24 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Peeled/Delaminated11
  • Detachment of Device or Device Component8
  • Material Fragmentation2
  • Material Separation2
  • Naturally Worn2
  • Break1
  • Crack1
  • Degraded1

Patient problems

  • No Clinical Signs, Symptoms or Conditions26

Recent adverse event reports

The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50158801 ↗
  • MalfunctionJun 3, 2026Material Fragmentation

    THE CUSTOMER RETURNED THE THUNDERBEAT TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE OLYMPUS SERVICE TECHNICIAN INDICATED THAT THE TISSUE PAD WAS WORN OUT, AND A PORTION OF IT WAS MISSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25374752 ↗
  • MalfunctionMay 29, 2026Peeled/Delaminated

    THE CUSTOMER RETURNED THE THUNDERBEAT TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING OLYMPUS¿ DEVICE ANALYSIS IT WAS INDICATED THAT THE THUNDERBEAT SURGICAL INSTRUMENT'S TISSUE PAD PEELED OFF . THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVEN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25340775 ↗
  • MalfunctionApr 29, 2026Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE THUNDERBEAT HAD PEELING OF THE TISSUE PAD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25034053 ↗
  • MalfunctionJan 23, 2026Crack; Peeled/Delaminated; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE OPEN FINE JAW TYPE X EXHIBITED PEELING OF THE TISSUE PAD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24160232 ↗
  • MalfunctionJan 5, 2026Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23969443 ↗
  • MalfunctionDec 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE SUBJECT DEVICE HANDPIECE DISASSEMBLED WHILE IN USE DUE TO LOOSENING. THE ISSUE OCCURRED DURING A THERAPEUTIC THYROIDECTOMY PROCEDURE. THE PROCEDURE WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23877784 ↗
  • InjuryDec 17, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23829575 ↗
  • MalfunctionNov 17, 2025Naturally Worn; Naturally Worn

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE OPEN FINE JAW TYPE X EXHIBITED TISSUE PAD WAS WORN. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23567765 ↗
  • MalfunctionNov 14, 2025Naturally Worn; Naturally Worn

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE OPEN FINE JAW TYPE X EXHIBITED THE TISSUE PAD WAS UNEVENLY WORN. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23563422 ↗
  • MalfunctionOct 29, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE THUNDERBEAT EXHIBITED PEELING AWAY OF THE TISSUE PAD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23417515 ↗
  • MalfunctionAug 29, 2025Peeled/Delaminated; Peeled/Delaminated

    NO ADDITIONAL CUSTOMER INFO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22929143 ↗
  • MalfunctionAug 28, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE TISSUE PAD WAS PEELED OFF. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22916112 ↗
  • MalfunctionAug 18, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE TISSUE PAD IN THE GRASPING SECTION WAS WORN AWAY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22833884 ↗
  • MalfunctionApr 1, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC SURGICAL INSTRUMENT EXHIBITED PEELING OF THE TISSUE PAD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21742725 ↗
  • MalfunctionMar 28, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE TISSUE PAD WAS WORN OUT, AND IT WAS PARTIALLY PEELED OFF. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21721477 ↗
  • MalfunctionMar 3, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THE SUBJECT DEVICE EXHIBITED TORN TEFLON. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21504786 ↗
  • MalfunctionJan 7, 2025Melted; Melted

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21097014 ↗
  • MalfunctionJan 6, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE INVESTIGATION FROM AN IMAGE PROVIDED BY THE CUSTOMER, THE THUNDERBEAT EXHIBITED A PEELING PAD. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21085975 ↗
  • MalfunctionMar 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED, THE ULTRASONIC SURGICAL DEVICE TISSUE PAD FELL OFF AND IT IS UNKNOWN WHETHER IT IS OUTSIDE OR INSIDE THE BODY. THE ISSUE OCCURRED DURING HEPATOTOMY THERAPEUTIC PROCEDURE. THE CUSTOMER IMMEDIATELY COLLECTED IT AND REPLACED WITH ANOTHER ONE, AND THE INTENDED PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT INJURY OR HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18937590 ↗
  • InjurySep 5, 2023Peeled/Delaminated; Difficult to Remove; Detachment of Device or Device Component; Peeled/Delaminated; Difficult to Remove; Detachment of Device or Device Component

    DURING A LARYNGECTOMY WITH LARYNGEAL TUMOR AND NECK DISSECTION, THE PROBE TIP OF THE THUNDERBEAT OPEN FINE JAW TYPE X BROKE AND FELL OFF WHEN THE NECK DISSECTION WAS COMPLETED AND CONNECTED TO THE ESOPHAGUS DURING THE OPERATION. THE PROBE WAS IMMEDIATELY COLLECTED WITH FORCEPS AND THERE WERE NO REMNANTS. THE EVENT OCCURRED SIXTEEN HOURS AFTER THE START OF THE PROCEDURE WHICH TAKES ABOUT TWENTY-FOUR HOURS. ADDITIONALLY, THE TISSUE PAD WAS PEELING OFF AND WAS DAMAGED BY METAL-TO-METAL CONTACT. AT THE END OF THE PROCEDURE, THE THUNDERBEAT SCISSORS COULD NOT BE REMOVED FROM THE THUNDERBEAT TRANSDUCER AND THE USER INTENTIONALLY BROKE IT APART WITH AN ULTRASONIC SCALPEL AND REMOVED IT. THE PROCEDURE WAS COMPLETED WITH A MARIS BIPOLAR WHICH WAS ORIGINALLY USED IN THE SURGERY, AND WITHOUT CHANGING THE EQUIPMENT. THERE WAS NO PROCEDURE EXTENSION, NO DELAY IN SURGERY AND NO ADDITIONAL ANESTHESIA. THERE WERE NO REPORTS OF PATIENT HARM. THE DEVICE WAS INSPECTED BEFORE USE WITH NO ANOMALIES AND THE CONNECTION TO THE GENERATOR WAS VERIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17684935 ↗
  • MalfunctionAug 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING THE DEVICE EVALUATION, IT WAS CONFIRMED THAT THE BROKE WAS BROKEN. THIS REPORT IS BEING SUBMITTED FOR THE BROKEN PROBE. ADDITIONALLY, THE CUSTOMER PROVIDED ADDITIONAL INFORMATION WHICH CONFIRMED THAT THE DAMAGED DEVICE DID NOT FALL INTO THE PATIENT'S BODY OR ON THE GROUND. THE CUSTOMER HAD TROUBLE SEPARATING THE DEVICE FROM THE HANDPIECE. THE DAMAGED TRANSDUCER EVENT OCCURRED AFTER THE PROCEDURE DURING CLEAN UP. THERE WAS NO IMPACT TO THE PATIENT OR THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17510519 ↗
  • MalfunctionJun 2, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED TO OLYMPUS, THE TISSUE PAD OF THE THUNDERBEAT OPEN FINE JAW TYPE X WAS DAMAGED. THE ISSUE WAS FOUND ONE TO TWO HOURS AFTER STARTING A MALIGNANT TONGUE SURGERY. INSPECTION AND TESTING OF THE RETURNED DEVICE FOUND INSULATION PAD PEELING. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17054447 ↗
  • MalfunctionMar 27, 2023Material Separation; Material Separation

    AN OLYMPUS EMPLOYEE REPORTED ON BEHALF OF THE CUSTOMER, AT THE BEGINNING OF AN OPERATION AN ERROR OF ULTRASONIC AMPLITUDE ABNORMALITY (U509) OCCURRED WITH THE THUNDERBEAT. THE DEVICE BECAME UNUSABLE, AND THE THERAPEUTIC HEAD AND NECK SURGERY WAS COMPLETED USING A REPLACEMENT DEVICE. THERE WAS A DELAY FOR DEVICE REPLACEMENT, BUT NO ADDITIONAL ANESTHESIA WAS NECESSARY FOR THE PATIENT. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT. DURING INCOMING INSPECTION, THERE WAS SEVERE WEAR OF TISSUE PAD. THIS MEDWATCH IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16618922 ↗
  • MalfunctionMar 7, 2023Material Separation; Material Separation

    AS REPORTED FOR THIS EVENT BY THE CUSTOMER, 30 MINUTES INTO A THYROIDECTOMY IN LAPAROTOMY PROCEDURE, THE TISSUE PAD OF THE DEVICE BEGAN TO PEEL OFF. PROCEDURE WAS COMPLETED WITH A NEW DEVICE OF THE SAME MODEL NUMBER WITHOUT ANY DELAY. THE PATIENT DID NOT NEED ADDITIONAL ANESTHESIA. THERE IS NO HARM OR ADVERSE IMPACT TO THE PATIENT. TOTAL OPERATING TIME OF THE PROCEDURE IS WITHIN TWO HOURS. THE INTELLIGENT TISSUE MONITORING (ITM) SETTING WAS OFF DURING THE PROCEDURE. THE DEVICE WAS INSPECTED PRIOR TO USE WITH NO ANOMALY NOTED. THE CONNECTIONS TO THE GENERATOR WERE CHECKED. THE TRANSDUCER CORDS AND OTHER MEDICAL DEVICE CORDS WERE NOT BUNDLED TOGETHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16496467 ↗

Other Instrument, ultrasonic surgical devices