URETERO-RENO VIDEOSCOPE OLYMPUS URF-V3R (UDI 04953170403392)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of URETERO-RENO VIDEOSCOPE (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus ureteroreno videoscope is a flexible camera used inside the urinary tract. Many of the 180 reports describe parts coming loose, corrosion and breaks. The reports include 170 malfunctions, 10 injuries and no deaths. Reports from the last 12 months number 132, up from 26 in the 12 months before. One recall, started in September 2023 and still open, concerns scopes returned after repair with water left in their channels, which could lead to infection.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
182
Deaths reported
0
Injuries reported
10
Recalls naming it
1

Summary

The FDA's MAUDE database contains 182 adverse event reports that list this exact device (DI 04953170403392), received between Dec 27, 2019 and Sep 30, 2026 — classified as 10 injuries, 172 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (43 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 4 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 3 reportsSep 2025: 2 reportsOct 2025: 6 reportsNov 2025: 7 reportsDec 2025: 15 reportsJan 2026: 30 reportsFeb 2026: 9 reportsMar 2026: 10 reportsApr 2026: 13 reportsMay 2026: 20 reportsJun 2026: 10 reportsJul 2026: 9 reportsAug 2026: 5 reportsSep 2026: 6 reportsOct 2026: 0 reports20242025202630
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0348-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component43
  • Corroded22
  • Break16
  • Failure to Clean Adequately13
  • Degraded12
  • Material Split, Cut or Torn10
  • Mechanical Jam10
  • Device Reprocessing Problem9

Patient problems

  • No Clinical Signs, Symptoms or Conditions172
  • Foreign Body In Patient2
  • Unspecified Tissue Injury2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Insufficient Information1
  • Laceration(s)1
  • No Patient Involvement1
  • Physical Entrapment1

Recent adverse event reports

The 25 most recent of 182 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Corroded

    The customer returned the Uretero Reno Videoscope for a separate issue. No patient involvement. During the initial evaluation of the device, it was observed that the bending section, the rotation mechanism, the control body and the scope connector adhesive had corrosion.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941881 ↗
  • MalfunctionSep 28, 2026Corroded

    The customer returned the Uretero-Reno Videoscope for a separate issue. During inspection of the device, the service repair technician indicates that there was corrosion in the bending section. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915610 ↗
  • MalfunctionSep 25, 2026Positioning Failure; Physical Resistance/Sticking

    The customer reported that the Uretero-Reno Videoscope had a separate issue. During the initial evaluation of the device, it was observed that the control knob angulation lever was stuck at the neutral position. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50900337 ↗
  • MalfunctionSep 25, 2026Corroded; Positioning Problem

    The customer reported the Uretero-Reno Videoscope had a separate issue. During the initial evaluation of the device it was observed, that distal end plastic cover and the objective lens had corrosion, and the control knob was stuck, no movement. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50898382 ↗
  • MalfunctionSep 4, 2026Degraded

    No additional information received from the customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50621080 ↗
  • MalfunctionSep 4, 2026Residue After Decontamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50615525 ↗
  • MalfunctionAug 27, 2026Mechanical Jam

    The customer returned the Endoeye Flex Deflectable Videoscope for a separate issue. During the initial evaluation of the device it was observed that there was zero movement on the control knob. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50478473 ↗
  • MalfunctionAug 13, 2026Loss of or Failure to Bond

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50238821 ↗
  • MalfunctionAug 12, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50223304 ↗
  • MalfunctionAug 12, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50221673 ↗
  • MalfunctionAug 12, 2026Material Fragmentation

    The customer returned the Uretero-Reno Videoscope to the service center for a separate issue. During the device analysis, it was indicated that the bending section cover adhesive and C-cover was chipped. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50221007 ↗
  • MalfunctionJul 29, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50065045 ↗
  • MalfunctionJul 27, 2026Positioning Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50046630 ↗
  • MalfunctionJul 27, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50036892 ↗
  • MalfunctionJul 21, 2026Material Fragmentation

    THE CUSTOMER RETURNED THE URETERO-RENO VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATED THAT A CHIPPED ADHESIVE AT THE BENDING RUBBER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT. ¿

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25884718 ↗
  • MalfunctionJul 20, 2026Degraded

    THE CUSTOMER RETURNED THE URETERO-RENO VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, IT WAS INDICATED THAT A CHIPPED GLUE AT THE BENDING SECTION COVER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25882921 ↗
  • MalfunctionJul 15, 2026Loss of or Failure to Bond

    THE CUSTOMER RETURNED THE URETERO-RENO FIBERSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATED THE BENDING SECTION COVER/DISTAL SHEATH GLUE WAS LIFTED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25829728 ↗
  • MalfunctionJul 13, 2026Detachment of Device or Device Component

    THE CUSTOMER RETURNED THE URETERO-RENO VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT THERE IS MISSING BENDING SECTION COVER GLUE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25803648 ↗
  • MalfunctionJul 7, 2026Loss of or Failure to Bond

    THE CUSTOMER RETURNED THE URETERO-RENO VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, IT WAS INDICATED THAT A PEELED BENDING SECTION COVER ADHESIVE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25738885 ↗
  • MalfunctionJul 7, 2026Material Fragmentation

    THE CUSTOMER RETURNED THE URETERO-RENO VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATED THAT A CHIPPED ADHESIVE ON THE BENDING SECTION COVER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25737979 ↗
  • MalfunctionJun 30, 2026Detachment of Device or Device Component

    THE CUSTOMER RETURNED THE URETERO-RENO VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT A CHIP ON THE A-RUBBER GLUE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25668589 ↗
  • MalfunctionJun 25, 2026Degraded

    THE CUSTOMER RETURNED THE URETERO-RENO VIDEOSCOPE, WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, ADHESIVE WAS CHIPPED AND LIFTED AT THE BENDING SECTION. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE WAS DUE TO STRESS PROBLEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25631176 ↗
  • MalfunctionJun 25, 2026Peeled/Delaminated

    THE CUSTOMER RETURNED THE URETERO-RENO VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE ADHESIVE WAS DETACHED AT THE BENDING SECTION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25631139 ↗
  • MalfunctionJun 23, 2026Material Split, Cut or Torn

    THE CUSTOMER RETURNED THE URETERO-RENO VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATED THAT A CHIPPED A-RUBBER ADHESIVE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25602052 ↗
  • MalfunctionJun 22, 2026Break

    THE CUSTOMER RETURNED THE URETERO-RENO VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATED THAT A CHIPPED BENDING SECTION ADHESIVE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT. ¿

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25586478 ↗