CONNECTING TUBE MAJ-2111Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of CONNECTING TUBE (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus connecting tube is an accessory used for endoscope cleaning. All 164 of its adverse event reports are malfunctions, with no injuries and no deaths. Most describe the tube's clips or lock levers breaking, often during cleaning and disinfection, with some describing it coming apart. One open recall, started in March 2026, concerns lock levers on a newer version that may fail early. Reports fell from 5 in the prior 12 months to 0 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
164
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 164 adverse event reports that list this exact device (DI 04953170404061), received between Mar 15, 2022 and Oct 17, 2025 — classified as 164 malfunctions.

The most frequently reported specific device problem is “Break” (92 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K190969); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 4 reportsFeb 2024: 9 reportsMar 2024: 11 reportsApr 2024: 5 reportsMay 2024: 4 reportsJun 2024: 13 reportsJul 2024: 14 reportsAug 2024: 5 reportsSep 2024: 12 reportsOct 2024: 4 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 25, 2026Class IIOpen, Classified

    Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.

    Z-1899-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break92
  • Connection Problem32
  • Detachment of Device or Device Component18
  • Disconnection8
  • Fracture4
  • Crack3
  • Material Integrity Problem3
  • Material Separation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions164

Recent adverse event reports

The 25 most recent of 164 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 17, 2025Degraded; Degraded

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23327990 ↗
  • MalfunctionOct 17, 2025Degraded; Degraded

    IT WAS REPORTED THAT THE CONNECTING TUBE WAS DEFECTIVE AND BECAME DISCONNECTED DURING THE ALCOHOL FLUSH PHASE OF THE REPROCESSING CYCLE. THE ISSUE OCCURRED DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23325993 ↗
  • MalfunctionFeb 25, 2025Break; Break

    IT WAS REPORTED THAT THE CONNECTOR TUBE CLIPS WERE BROKEN DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21457438 ↗
  • MalfunctionFeb 24, 2025Break; Break

    IT WAS REPORTED THAT THE CONNECTOR TUBE CLIPS WERE BROKEN DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21445622 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT THE CONNECTING TUBE HAD A BROKEN PLASTIC CONNECTOR. THE EVENT OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21294094 ↗
  • MalfunctionOct 23, 2024Break; Break

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE CONNECTING TUBE HAD BOTH LOCK LEVERS WERE BROKEN OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20519291 ↗
  • MalfunctionOct 16, 2024Break; Connection Problem; Break; Connection Problem

    IT WAS REPORTED THAT THE CONNECTING TUBE (AN ENDOSCOPE REPROCESSOR ACCESSORY) COULD NOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN DUE TO BROKEN CLIPS. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20458853 ↗
  • MalfunctionOct 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING THE DEVICE EVALUATION, THE CONNECTING TUBE EXHIBITED BROKEN LOCK LEVERS ON THE REPROCESSOR SIDE CONNECTOR. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20457295 ↗
  • MalfunctionOct 14, 2024Crack; Material Integrity Problem; Crack; Material Integrity Problem

    IT WAS REPORTED THAT THE CONNECTOR OF THE ENDOSCOPE CONNECTING TUBE WAS FADING SHOWING A WEAKING OF THE PLASTIC AND SUPER SMALL CRACK. IT IS UNKNOWN WHEN THE REPORTED ISSUE OCCURRED. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20447889 ↗
  • MalfunctionSep 25, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE CONNECTING TUBE CONNECTORS WERE BREAKING AND POPPING OFF MID CYCLE ON BOTH ENDOSCOPE REPROCESSORS. THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20304389 ↗
  • MalfunctionSep 25, 2024Break; Disconnection; Break; Disconnection

    IT WAS REPORTED THE CONNECTING TUBE CONNECTORS WERE BREAKING AND POPPING OFF MID CYCLE. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20303641 ↗
  • MalfunctionSep 25, 2024Break; Break

    THE CUSTOMER CALLED AND REPORTED THAT THE CONNECTORS ON HER OLYMPUS CONNECTING TUBES WERE BREAKING AND POPPING OFF MID-CYCLE DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20302930 ↗
  • MalfunctionSep 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE CONNECTING TUBE CLIPS HAD CRACKED AND BROKEN. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20260328 ↗
  • MalfunctionSep 16, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT, THE CLEAR PLASTIC ON CONNECTING TUBE SEPARATED FROM THE METAL PORTION. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20226280 ↗
  • MalfunctionSep 12, 2024Break; Break

    IT WAS REPORTED THAT THE CONNECTING TUBE'S CONNECTOR WAS BROKEN. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20208428 ↗
  • MalfunctionSep 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED, THAT THE CONNECTING TUBE'S TAB BROKE OFF. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20194780 ↗
  • MalfunctionSep 11, 2024Break; Break

    IT WAS REPORTED THAT THE CONNECTING TUBE'S CONNECTOR WAS BROKEN. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20189253 ↗
  • MalfunctionSep 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE CONNECTING TUBE WAS DAMAGED WHEN DISCONNECTING AFTER REPROCESSING. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20172160 ↗
  • MalfunctionSep 9, 2024Connection Problem; Connection Problem

    IT WAS REPORTED, THE CONNECTING TUBE EXHIBITED CONNECTING TUBE COULD NOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF DELAY, PATIENT HARM, INJURY, OR DEATH. ADDITIONAL DETAILS RELATING TO THE PATIENT AND THE EVENT HAVE BEEN REQUESTED, BUT NO RESPONSE HAS BEEN RECEIVED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20169175 ↗
  • MalfunctionSep 9, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE CONNECTING TUBE BROKE AND CANNOT BE CONNECTED. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20168014 ↗
  • MalfunctionSep 9, 2024Break; Break

    IT WAS REPORTED, THE CONNECTING TUBE EXHIBITED THE CONNECTORS ON BOTH OER-ELITES ARE BREAKING AND POPPING OFF MID CYCLE ON BOTH OER-ELITES.. THE ISSUE IS UNKNOWN AS TO WHEN IT OCCURRED. THERE WERE NO REPORTS OF DELAY, PATIENT HARM, INJURY, OR DEATH. ADDITIONAL DETAILS RELATING TO THE PATIENT AND THE EVENT HAVE BEEN REQUESTED, BUT NO RESPONSE HAS BEEN RECEIVED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20167752 ↗
  • MalfunctionAug 21, 2024Break; Break

    THE REPORTED ISSUE WAS NOTICED AFTER REPROCESSING AND IT WAS CONFIRMED THAT THE DEVICE WAS INSPECTED BEFORE USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20014214 ↗
  • MalfunctionAug 18, 2024Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS REPORTED THAT THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19999096 ↗
  • MalfunctionAug 15, 2024Break; Break

    IT WAS REPORTED THAT THE CONNECTING TUBE WAS BROKEN. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19989347 ↗
  • MalfunctionAug 6, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS REPORTED, THE CONNECTING TUBE EXHIBITED CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR DUE TO POOR REPROCESSING. THE ISSUE IS UNKNOWN AS TO WHEN IT OCCURRED. THERE WERE NO REPORTS OF DELAY, PATIENT HARM, INJURY, OR DEATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19920489 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K190969. One submission often covers many models, so check the product and lot details in each record.

  • Z-0638-2022 · Dec 13, 2021 · Terminated
    OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessor

Other Accessories, Cleaning, For Endoscope devices

All Accessories, Cleaning, For Endoscope devices →