Skip to content

CONNECTING TUBE MAJ-2113Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of CONNECTING TUBE (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus connecting tube links an endoscope to its reprocessing equipment during cleaning. Reports mostly describe the tube breaking or coming apart during reprocessing, often at its lock levers. The 159 reports include no deaths and no injuries. One open recall, begun in March 2026, concerns the lock levers on this reprocessor connecting tube, which may break early.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
159
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 159 adverse event reports that list this exact device (DI 04953170404085), received between Mar 15, 2022 and Oct 17, 2025 — classified as 159 malfunctions.

The most frequently reported specific device problem is “Break” (78 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K190969); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 5 reportsFeb 2024: 6 reportsMar 2024: 10 reportsApr 2024: 7 reportsMay 2024: 4 reportsJun 2024: 2 reportsJul 2024: 7 reportsAug 2024: 4 reportsSep 2024: 9 reportsOct 2024: 10 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 25, 2026Class IIOpen, Classified

    Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.

    Z-1901-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break78
  • Connection Problem44
  • Detachment of Device or Device Component18
  • Incomplete or Inadequate Connection9
  • Fracture6
  • Disconnection4
  • Component Missing3
  • Crack3

Patient problems

  • No Clinical Signs, Symptoms or Conditions159

Recent adverse event reports

The 25 most recent of 159 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 17, 2025Degraded; Degraded

    IT WAS REPORTED THAT THE CONNECTING TUBE WAS DEFECTIVE AND BECAME DISCONNECTED DURING THE ALCOHOL FLUSH PHASE OF THE REPROCESSING CYCLE. THE ISSUE OCCURRED DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23327817 ↗
  • MalfunctionOct 17, 2025Degraded; Degraded

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23327714 ↗
  • MalfunctionFeb 21, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21437173 ↗
  • MalfunctionJan 10, 2025Break; Disconnection; Break; Disconnection

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21129979 ↗
  • MalfunctionJan 10, 2025Loose or Intermittent Connection; Loose or Intermittent Connection

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21123566 ↗
  • MalfunctionJan 10, 2025Disconnection; Disconnection

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21123327 ↗
  • MalfunctionOct 28, 2024Break; Break

    IT WAS REPORTED THE CONNECTING TUBE HAD BROKEN LOCK LEVERS. IT WAS NOT REPORTED DURING WHAT TYPE OF DEVICE USAGE THE REPORTED EVENT OCCURRED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20553527 ↗
  • MalfunctionOct 23, 2024Connection Problem; Connection Problem

    DURING THE DEVICE EVALUATION, THE CONNECTING TUBE EXHIBITED BROKEN LOCK LEVERS ON THE REPROCESSOR SIDE CONNECTOR. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20517935 ↗
  • MalfunctionOct 23, 2024Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE CONNECTING TUBE CHANNEL CONNECTOR WAS BROKEN. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20517599 ↗
  • MalfunctionOct 23, 2024Connection Problem; Connection Problem

    DURING THE DEVICE EVALUATION, THE CONNECTING TUBE EXHIBITED BROKEN LOCK LEVERS ON THE REPROCESSOR SIDE CONNECTOR. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20516442 ↗
  • MalfunctionOct 23, 2024Connection Problem; Connection Problem

    DURING THE DEVICE EVALUATION, THE CONNECTING TUBE EXHIBITED BROKEN LOCK LEVERS ON THE REPROCESSOR SIDE CONNECTOR. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20516006 ↗
  • MalfunctionOct 23, 2024Connection Problem; Connection Problem

    DURING THE DEVICE EVALUATION, THE CONNECTING TUBE EXHIBITED BROKEN LOCK LEVERS ON THE REPROCESSOR SIDE CONNECTOR. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20515955 ↗
  • MalfunctionOct 23, 2024Break; Break

    IT WAS REPORTED THAT THE SUCTION CONNECTOR AND THE AUXILIARY WATER CONNECTOR WERE BROKEN FOR NO REASON. CONSEQUENTLY, THERE WAS CANCELLATION OF ALL PROCEDURES ON BECAUSE THE FACILITY COULD NOT REPROCESS THE SCOPES. THE ISSUE OCCURRED DURING REPROCESSING WITH NO EVIDENCE OF DIRECT PATIENT INVOLVEMENT. THOUGH PROCEDURES WERE CANCELLED, THERE WAS NO INDICATION THAT THESE WERE BEING DONE URGENTLY OR EMERGENTLY FOR LIFE-THREATENING REASONS OR THAT THE CANCELLATION OF SAID PROCEDURES CAUSED ANY LIFE-THREATENING COMPLICATIONS OR HEALTH CONDITIONS. AT THIS TIME, THERE IS NO EVIDENCE THAT A LIFE-THREATENING EVENT OCCURRED, THAT A PATIENT HAS DEVELOPED A PERMANENT DISABILITY OR PERMANENT DAMAGE THAT CAUSED SUBSTANTIAL DISRUPTION IN THEIR ABILITY TO CONDUCT NORMAL LIFE FUNCTIONS, OR THAT ADDITIONAL MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED TO PRECLUDE PERMANENT IMPAIRMENT OF A BODY FUNCTION OR DAMAGE TO A BODY STRUCTURE THAT IS SUSPECTED TO BE DUE TO THE USE OF THE OLYMPUS DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20513413 ↗
  • MalfunctionOct 22, 2024Break; Break

    IT WAS REPORTED THE REPROCESSOR CONNECTING TUBE CLIPS WERE BREAKING OFF OFTEN. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20507463 ↗
  • MalfunctionOct 14, 2024Degraded; Degraded

    IT WAS REPORTED CONNECTING TUBE WAS FADING AND SHOWING WEAKENING OF THE PLASTIC AND SUPER SMALL CRACKS ALONG THE CONNECTORS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20436728 ↗
  • MalfunctionOct 9, 2024Break; Break

    IT WAS REPORTED THE CONNECTING TUBES FROM THE ENDOSCOPES TO THE REPROCESSOR WERE FRAGILE AND BREAKING OFTEN. THE COLORED ENDS HAVE TWO PLASTIC CLIPS WHICH HOLD THE CONNECTING TUBES TO THE REPROCESSOR, THESE CLIPS WERE BREAKING OFF. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20413795 ↗
  • MalfunctionSep 27, 2024Break; Break

    IT WAS REPORTED THAT THE CONNECTING TUBE HAD BROKEN CONNECTORS. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20322512 ↗
  • MalfunctionSep 25, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED, THE ENDOSCOPE REPROCESSOR ACCESSORIES CONNECTING TUBE ON THE REPROCESSER ARE BREAKING AND POPPING OFF MID CYCLE. THE ISSUE OCCURRED DURING [EVENT]. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20302818 ↗
  • MalfunctionSep 25, 2024Break; Break

    THE CUSTOMER CALLED AND REPORTED THAT THE CONNECTORS ON HER OLYMPUS CONNECTING TUBES WERE BREAKING AND POPPING OFF MID-CYCLE DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20302745 ↗
  • MalfunctionSep 25, 2024Break; Break

    THE CUSTOMER CALLED AND REPORTED THAT THE CONNECTORS ON HER OLYMPUS CONNECTING TUBES WERE BREAKING AND POPPING OFF MID-CYCLE DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20302163 ↗
  • MalfunctionSep 23, 2024Break; Break

    IT WAS REPORTED THAT THE CONNECTING TUBE HAD BROKEN CONNECTORS. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20280510 ↗
  • MalfunctionSep 20, 2024Break; Break

    IT WAS REPORTED THAT THE CONNECTING TUBE HAD BROKEN CONNECTORS. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20268943 ↗
  • MalfunctionSep 17, 2024Break; Break

    IT WAS REPORTED, THE CONNECTING TUBE'S CONNECTORS ON BOTH OER-ELITES WERE BREAKING AND POPPING OFF MID CYCLE ON BOTH OER-ELITES. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20242648 ↗
  • MalfunctionSep 16, 2024Break; Break

    IT WAS REPORTED THAT THE CONNECTING TUBE HAD BROKEN CONNECTORS. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20232398 ↗
  • MalfunctionSep 4, 2024Break; Break

    IN SPEAKING WITH THE OLYMPUS TECHNICAL ASSISTANCE CENTER (TAC), THE CUSTOMER REPORTED THAT THE TAB BROKE ON MAJ-2113, THE CUSTOMER WAS THEN LATER TRANSFERRED TO CUSTOMER CARE TO OBTAIN AN RMA. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20140464 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K190969. One submission often covers many models, so check the product and lot details in each record.

  • Z-0638-2022 · Dec 13, 2021 · Terminated
    OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessor

Other Accessories, Cleaning, For Endoscope devices

All Accessories, Cleaning, For Endoscope devices →