ENDOSCOPE REPROCESSOR OER-MINIRecalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of ENDOSCOPE REPROCESSOR (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04953170331619), received between Dec 22, 2022 and Apr 23, 2026 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Device Reprocessing Problem” (8 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 31, 2025Open, Classified

    Inform existing affected device customers about pertinent Warnings/Cautions, existing recommended maintenance schedules and that only properly trained personnel conduct repairs and maintenance.

    Z-0874-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. May 26, 2023Open, Classified

    Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

    Z-2343-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Reprocessing Problem8
  • Break1
  • Device Remains Activated1
  • Failure to Clean Adequately1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Unspecified Infection2

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2026Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ENDOSCOPE REPROCESSOR PRESENTED SCOPES THAT HAD WHITE SPECKS ON THEM AFTER REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24973703 ↗
  • MalfunctionJan 20, 2026Device Reprocessing Problem

    NO NEW INFORMATION RECEIVED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24116227 ↗
  • MalfunctionOct 30, 2024Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT THE ENDOSCOPE REPROCESSOR WAS CONSTANTLY FILLING UP WITH WATER WHILE OUT OF OPERATION AND POWERED OFF. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20564187 ↗
  • MalfunctionOct 22, 2024Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS REPORTED THAT THE ENDOSCOPE PREPROCESSOR EXHIBITED ISSUES RELATED TO WATER FLOW AND SLOW FILL TIMES. AFTER DETERMINING THAT THE PROBLEM WAS DUE TO SLOW WATER FILL TIMES, THE FIELD SERVICE ENGINEER (FSE) FACILITATED THE REPLACEMENT OF THE FILTERS. ALTHOUGH THE FILTERS WERE SLIGHTLY AHEAD OF THE REGULAR REPLACEMENT SCHEDULE, THE FSE NOTED THAT THE LAST FEW FILTER REPLACEMENTS HAD ALSO BEEN CARRIED OUT ABOUT A WEEK EARLIER THAN PLANNED. ONCE THE FILTERS WERE REPLACED, NO FURTHER ISSUES WITH THE UNIT'S DRAINING AND CHEMICAL LOADING WERE OBSERVED. IT IS UNKNOWN WHEN THE ISSUE WAS FOUND. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20509810 ↗
  • MalfunctionJul 10, 2024Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS REPORTED THAT THE ENDOSCOPE REPROCESSORS WATER FILTER CANNOT BE DISINFECTED. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19711687 ↗
  • MalfunctionJun 12, 2024Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS REPORTED THE ENDOSCOPE REPROCESSOR HAD FAILING TEST STRIPS. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19519042 ↗
  • MalfunctionMar 13, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS REPORTED THAT THERE WAS A BLACK SPECK WITHIN THE BASIN, DURING USE OF THE ENDOSCOPE REPROCESSOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18898253 ↗
  • MalfunctionOct 9, 2023Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS ADDITIONALLY REPORTED BY THE CUSTOMER THAT THE ISSUE WAS RESOLVED THE SAME DAY IT WAS REPORTED. THE FACILITY RAN ANOTHER CYCLE AND THE TEST STRIPS PASSED. NO PATIENTS WERE AFFECTED. IT WAS FURTHER CLARIFIED THAT NO SCOPES WERE PROCESSED UNTIL THE MATTER WAS FIXED, AND THERE WERE NO INFECTIONS/CONCERN FOR INFECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Infection

    MAUDE report 17896544 ↗
  • InjuryOct 9, 2023Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS ADDITIONALLY REPORTED BY THE CUSTOMER THAT THE ISSUE WAS RESOLVED THE SAME DAY IT WAS REPORTED. THE FACILITY RAN ANOTHER CYCLE AND THE TEST STRIPS PASSED. NO PATIENTS WERE AFFECTED. IT WAS FURTHER CLARIFIED THAT NO SCOPES WERE PROCESSED UNTIL THE MATTER WAS FIXED, AND THERE WERE NO INFECTIONS/CONCERN FOR INFECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Infection

    MAUDE report 17896477 ↗
  • MalfunctionJul 26, 2023Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THAT SHE HAD POTENTIALLY INADVERTENTLY USED INCORRECT REPROCESSING SOLUTION WHILE OPERATING HER OLYMPUS ENDOSCOPE REPROCESSOR. THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT. HOWEVER, ADDITIONAL DETAILS CONCERNING THE POTENTIAL CONNECTION TO PATIENTS HAVE BEEN REQUESTED. THUS FAR, NO RESPONSES HAVE BEEN RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17397173 ↗
  • MalfunctionDec 22, 2022Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, THE ENDOSCOPE REPROCESSOR HAD A DAMAGED YELLOW CONNECTOR. THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16035314 ↗

Other Accessories, Cleaning, For Endoscope devices

All Accessories, Cleaning, For Endoscope devices →