VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2R (REF N6005830)Recalls & reported problems
Olympus Medical Systems Corp. · VISERA CYSTO-NEPHRO VIDEOSCOPE · Class 2
All models of VISERA CYSTO-NEPHRO VIDEOSCOPE (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus Cysto-Nephro videoscope is a flexible scope used to look inside the urinary tract. The 60 reports mostly describe problems found during repair or cleaning: rust and corrosion on the bending section and lens, chipped adhesive, material splitting or tearing, and cleaning or reprocessing issues. There are no deaths and no injuries, and most reports involve no patient. No recalls name this device. Reports rose from 17 in the prior 12 months to 39 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 64
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 04953170411212), received between Oct 8, 2020 and Sep 24, 2026 — classified as 64 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (14 reports).
5 other recalls cover devices cleared under the same FDA submission (K221683); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component14
- Degraded7
- Device Reprocessing Problem6
- Corroded5
- Failure to Clean Adequately5
- Material Deformation5
- Material Split, Cut or Torn5
- Loss of or Failure to Bond4
Patient problems
- No Clinical Signs, Symptoms or Conditions61
- No Patient Involvement2
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Residue After Decontamination
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50892244 ↗MalfunctionSep 24, 2026Stretched; Material Deformation
The videoscope was returned to the service center with a report of a cracked sheath. This does not indicate an MDR reportable event. However, MDR reportable failure was found during testing at the service center. During inspection of the device, it was found that the bending section cover had stretched and bunched up. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885131 ↗MalfunctionSep 8, 2026Detachment of Device or Device Component
The customer returned the Cysto Nephro Videoscope for a separate issue. During the initial evaluation of returned device it was found the bending section cover/distal sheath is missing. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50668377 ↗MalfunctionSep 1, 2026Material Deformation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50548810 ↗MalfunctionAug 21, 2026Stretched
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50376103 ↗MalfunctionAug 12, 2026Material Deformation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50227786 ↗MalfunctionAug 10, 2026Degraded
The customer returned the Cysto-Nephro Videoscope to the service center for evaluation related to a separate issue. During testing, the service repair technician identified rust was present around the objective lens. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50197241 ↗MalfunctionJul 27, 2026Loss of or Failure to Bond
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50025521 ↗MalfunctionJul 10, 2026Material Fragmentation
THE CUSTOMER RETURNED THE CYSTONEPHROVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE BENDING SECTION ADHESIVE WAS FOUND TO BE CHIPPED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25788602 ↗MalfunctionJun 26, 2026Corroded
THE CUSTOMER RETURNED THE CYSTO-NEPHRO VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE BENDING SECTION WAS CORRODED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25637463 ↗MalfunctionJun 19, 2026Microbial Contamination of Device
THE CUSTOMER REPORTED THERE WAS FUZZ STUCK INSIDE OF AN OLYMPUS CYSTONEPHROVIDEOSCOPE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25563975 ↗MalfunctionMay 27, 2026Corroded
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CYSTO-NEPHRO VIDEOSCOPE OBJECTIVE LENS HAD RUST AROUND THE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25310860 ↗MalfunctionMay 27, 2026Mechanical Jam
THE CUSTOMER REPORTED STUCK WITH THE LOCKING MECHANISM ON DURING INSPECTION FOR USE INVOLVING AN OLYMPUS CYSTO-NEPHRO VIDEOSCOPE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25310719 ↗MalfunctionMay 27, 2026Material Split, Cut or Torn
THE VISERA CYSTO-NEPHRO VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A FAILED LEAK TEST. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, A CHIPPED CHIPPED ADHESIVE AT THE OBJECTIVE LENS AND LIGHT GUIDE LENS WERE FOUND TO BE MOST LIKELY DUE TO DEGRADATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25302835 ↗MalfunctionMay 27, 2026Detachment of Device or Device Component
THE VISERA CYSTO-NEPHRO VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A BLACK AND WHITE IMAGE WITH NO COLOR. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, A CHIPPED BENDING SECTION COVER GLUE WAS FOUND TO BE MOST LIKELY DUE TO STRESS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25295943 ↗MalfunctionMay 26, 2026Detachment of Device or Device Component
THE CYSTO NEPHRO VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF SCOPE HAS NO COLOR ON THE IMAGE. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, LENS GLUE IS MISSING WAS FOUND AT ESTIMATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25279043 ↗MalfunctionMay 22, 2026Contamination
THE VISERA CYSTO-NEPHRO VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A BLURRY IMAGE. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, BLACK RESIDUE INSIDE THE OBJECTIVE LENS WAS FOUND TO BE MOST LIKELY DUE TO DEGRADATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25257954 ↗MalfunctionMay 22, 2026Crack; Material Twisted/Bent
THE CYSTO-NEPHRO VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER FOR DAMAGE TO THE BENDING SECTION. WHICH IS NOT AN MDR REPORTABLE FAILURE. HOWEVER, MDR REPORTABLE FAILURES WERE FOUND DURING TESTING AT THE SERVICE CENTER. THE ADHESIVE WAS CHIPPED AT THE BENDING SECTION, ADHESIVE WAS CHIPPED AT THE OBJECTIVE LENS AND ADHESIVE WAS CHIPPED AT THE LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25251684 ↗MalfunctionMay 21, 2026Detachment of Device or Device Component
THE CYSTO-NEPHRO VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF FORGETTING TO PUT THE CAP ON AND THE END BLEW OUT. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, CHIPPED BENDING SECTION COVER GLUE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25250242 ↗MalfunctionMay 20, 2026Detachment of Device or Device Component
THE CYSTO-NEPHRO VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF A CRACK TOWARDS THE HANDLE. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT, HOWEVER, A REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, MISSING/DETACHED ADHESIVE AT LIGHT GUIDE LENS AND DETACHED CHIPPED/DETACHED BENDING SECTION SHEATH ADHESIVE WAS FOUND, LIKELY DUE TO STRESS. THERE WAS NO PATIENT INVOLVEMENT, AND NO HARM WAS REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25234744 ↗MalfunctionMay 20, 2026Material Split, Cut or Torn
THE CYSTO-NEPHRO VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF THE VIDEO PICTURE WILL NOT COME IN THE SCREEN, THEN IT WILL SWITCH TO COLOR BARS THEN IT WILL COMPLETELY SHUT OFF. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, CHIPPED ADHESIVE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25234551 ↗MalfunctionMay 13, 2026Loss of or Failure to Bond
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BENDING SECTION COVER ADHESIVE OF THE CYSTONEPHROVIDEOSCOPE LIFTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25161954 ↗MalfunctionMay 11, 2026Degraded
THE CYSTO-NEPHRO VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A FAILED LEAK TEST. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, AN MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, THE BENDING SECTION ADHESIVE WAS CHIPPED AND WAS FOUND TO BE MOST LIKELY DUE TO STRESS LED TO COMPONENT FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25136881 ↗MalfunctionMay 6, 2026Detachment of Device or Device Component
THE CYSTO-NEPHRO VIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF FAILED LEAK TEST AND PUNCTION MARK BY A NEEDLE IN THE WORKING CHANNEL. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, CHIPPED BENDING SECTION COVER GLUE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25107338 ↗MalfunctionApr 24, 2026Contamination /Decontamination Problem
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE CYSTO-NEPHRO VIDEOSCOPE EXHIBITED RESIDUE ON THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24987087 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K221683. One submission often covers many models, so check the product and lot details in each record.
- Z-0690-2024 · Dec 4, 2023 · Open, ClassifiedModel No. CYF-VHR, CYSTO-NEPHRO VIDEOSCOPE
- Z-0689-2024 · Dec 4, 2023 · Open, ClassifiedModel No. CYF-VH, CYSTO-NEPHRO VIDEOSCOPE
- Z-0691-2024 · Dec 4, 2023 · Open, ClassifiedModel No. CYF-V2, VISERA Cysto-Nephro Videoscope
- Z-0692-2024 · Dec 4, 2023 · Open, ClassifiedModel No. CYF-V2R, VISERA Cysto-Nephro Videoscope
- Z-0331-2024 · Sep 25, 2023 · Open, ClassifiedCystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
Other Cystoscope And Accessories, Flexible/Rigid devices
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHOlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls1 death1,380 reports
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHROlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls377 reports
- VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2Olympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid1 recall418 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall353 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall1 death332 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid2 recalls75 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid114 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall54 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗