OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5 (REF N3821744)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of OES CYSTONEPHROFIBERSCOPE (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04953170434532), received between May 11, 2023 and Jul 22, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Microbial Contamination of Device” (3 reports).
3 other recalls cover devices cleared under the same FDA submission (K221690); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Microbial Contamination of Device3
- Device Reprocessing Problem2
- Break1
- Detachment of Device or Device Component1
- Failure to Clean Adequately1
- Flaked1
- Loose or Intermittent Connection1
- No Display/Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS OBSERVED DURING OLYMPUS' DEVICE INSPECTION THAT THE CYSTONEPHROFIBERSCOPE'S BENDING SECTION COVER ADHESIVE WAS DETACHED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22566586 ↗MalfunctionApr 15, 2024Flaked; Flaked
IT WAS REPORTED THE CYSTONEPHROFIBERSCOPE HAD BLACK COATING FLAKED OFF. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19105885 ↗MalfunctionNov 30, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE FIELD SERVICE TECHNICIAN WAS ON-SITE AT A CUSTOMER LOCATION AND FOUND THE CYSTONEPHROFIBERSCOPE NEEDED AN INSPECTION AND REPAIR. THE INSPECTION FOUND THE DISTAL END AND BIOPSY CHANNEL HAD FOREIGN MATERIAL ATTACHED. THERE WERE NO REPORTS OF PATIENT OR USER HARM OR PROCEDURE ASSOCIATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18243739 ↗MalfunctionNov 10, 2023Break; No Display/Image; Break; No Display/Image
THE CUSTOMER REPORTED TO OLYMPUS, THE CYSTONEPHROFIBERSCOPE BENDING SECTION COVER WAS COMPLETELY BROKEN, THE SCOPE HAD A LEAK AND THE SCOPE HAD NO OPTIMAL VISUALIZATION. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18115749 ↗MalfunctionNov 9, 2023Loose or Intermittent Connection; Loose or Intermittent Connection
THE CUSTOMER REPORTED TO OLYMPUS, THE OES CYSTONEPHROFIBERSCOPE HAD AN IMAGE PROBLEM AND THE INSTRUMENT CHANNEL PORT WAS LOOSE. IT IS UNKNOWN WHEN THE ISSUE WAS FOUND. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18107181 ↗MalfunctionAug 15, 2023Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED TO OLYMPUS MEDICAL THAT THE OES CYSTONEPHROFIBERSCOPE WAS REPROCESSED WITHOUT THE ETO CAP ATTACHED AS PER DEVICE INSTRUCTIONS. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17553573 ↗MalfunctionAug 1, 2023Microbial Contamination of Device; Microbial Contamination of Device
AS REPORTED, AFTER A MICROBIOLOGICAL ROUTINE CONTROL TEST ON THE SUBJECT MEDICAL DEVICE AS REQUIRED BY FRENCH REGULATION, THE USER DETECTED AN UNEXPECTED CONTAMINATION. THE ISSUE FOUND DURING REPROCESSING. THE CUSTOMER THEN LEFT THE DEVICE TO THE OLYMPUS FRANCE (OFR) FRENCH OLYMPUS SUBSIDIARY FOR HYGIENE MICROBIOLOGICAL INVESTIGATION (HMI) FOR FURTHER INVESTIGATION. NO REPORT OF ANY CONTAMINATION OR ANY PATIENT INJURY OR PATIENT INFECTION TO WHICH THIS MEDICAL DEVICE COULD HAVE BEEN A CONTRIBUTORY CAUSE. NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17432992 ↗MalfunctionJul 28, 2023Device Reprocessing Problem; Device Reprocessing Problem
AN OLYMPUS MEDICAL ENDOSCOPY SUPPORT SPECIALIST PERFORMED A SCOPE REPROCESSING IN-SERVICE AT THE CUSTOMER SITE AS PART OF ANNUAL TRAINING. DURING THE ANNUAL TRAINING, A STAFF MEMBER DEMONSTRATED REPROCESSING AND THE TECHNICIAN DID NOT FLUSH THE DEVICE WITH DETERGENT SOLUTION, WATER AND AIR AFTER THE BRUSH-CLEANING WAS COMPLETED. THE FLUSHING STEPS ARE OUTLINED IN THE DEVICE REPROCESSING MANUAL. THE DEVICE DELIVERED A BIOPSY CHANNEL DETACHMENT ERROR WHEN IT WAS BEING REPROCESSED IN THE DEVICE REPROCESSOR. THE ENDOSCOPY SUPPORT SPECIALIST INSTRUCTED THE TECHNICIAN THAT IF THE SCOPE WERE BEING STERILIZED, THE FLUSHING STEPS HAD TO BE PERFORMED. THE TECHNICIAN REPEATED THE DEMONSTRATION OF REPROCESSING WITH ALL REQUIRED STEPS PERFORMED. THE ISSUE WAS RESOLVED DURING OBSERVATION. ONCE THE FACILITY INPUTTED THE CORRECT MODEL NUMBER, THE ERROR WENT AWAY. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17419869 ↗MalfunctionJul 6, 2023Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED TO OLYMPUS, THE OES CYSTONEPHROFIBERSCOPE TESTED POSITIVE FOR UNEXPECTED CONTAMINATION. THE SCOPE WAS SAMPLED ONCE. DURING THE SAMPLING, THE SCOPE TESTED POSITIVE FOR < 1 COLONY FORMING UNITS (CFUS) OF UNSPECIFIED AEROBIC MICROORGANISMS AND UNSPECIFIED ENTEROBACTERIA¿S. THE ISSUE WAS FOUND AT REPROCESSING DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17272628 ↗MalfunctionMay 11, 2023Microbial Contamination of Device; Microbial Contamination of Device
THE CUSTOMER REPORTED THAT THE OES CYSTONEPHROFIBERSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING REGULAR EXAMINATION IN THE HOSPITAL. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THIS MEDICAL DEVICE COULD HAVE BEEN A CONTRIBUTORY CAUSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16910383 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K221690. One submission often covers many models, so check the product and lot details in each record.
- Z-0684-2024 · Dec 4, 2023 · Open, ClassifiedModel No. CYF-5, Cystonephrofiberscope
- Z-0685-2024 · Dec 4, 2023 · Open, ClassifiedModel No. CYF-5R, Cystonephrofiberscope
- Z-0331-2024 · Sep 25, 2023 · Open, ClassifiedCystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
Other Cystoscope And Accessories, Flexible/Rigid devices
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHOlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls1 death1,380 reports
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHROlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls377 reports
- VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2Olympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid1 recall418 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall353 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall1 death332 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid2 recalls75 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid114 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall54 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page