CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHR (REF N3828633)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of CYSTO-NEPHRO VIDEOSCOPE (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04953170434730), received between May 28, 2024 and May 5, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Flaked” (1 report).
5 other recalls cover devices cleared under the same FDA submission (K221683); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Flaked1
- Loss of or Failure to Bond1
Patient problems
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2026Loss of or Failure to Bond
THE CUSTOMER DID NOT REPORT A MALFUNCTION. DURING AN INSPECTION OF THE ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE, THE ADHESIVE ON A-RUBBER BENDING SECTION HAD A CHIP FOUND. THE MOST LIKELY CAUSE OR PROBABLE CAUSE OF THE REPORTABLE MALFUNCTION IS DETERIORATION OR BREAKAGE OF THE ADHESIVE (CHEMICAL STRESS / PHYSICAL STRESS). THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25089772 ↗InjuryMay 28, 2024Flaked; Flaked
IT WAS REPORTED, DURING THE THERAPEUTIC CYSTOSCOPY, DEBRIS FROM THE CYSTO-NEPHRO VIDEOSCOPE FELL INTO THE PATIENT'S BLADDER. THERE WAS A PROCEDURAL DELAY WHILE THE DOCTOR TRIED TO BASKET THE DEBRIS OUT BUT COULD NOT CATCH THE DEBRIS IN THE NON-OLYMPUS BASKET. THE PATIENT URINATED THE DEBRIS OUT AND IS IN GOOD STANDING. ANOTHER CYSTOSCOPY WAS PERFORMED LAST WEEK. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE OUTCOME WAS NOT AFFECTED AS A RESULT OF THE ISSUE. THE DEVICE WAS INSPECTED BEFORE USE AND APPEARED TO BE NORMAL.
Patient problems: Foreign Body In Patient
MAUDE report 19403927 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K221683. One submission often covers many models, so check the product and lot details in each record.
- Z-0690-2024 · Dec 4, 2023 · Open, ClassifiedModel No. CYF-VHR, CYSTO-NEPHRO VIDEOSCOPE
- Z-0691-2024 · Dec 4, 2023 · Open, ClassifiedModel No. CYF-V2, VISERA Cysto-Nephro Videoscope
- Z-0692-2024 · Dec 4, 2023 · Open, ClassifiedModel No. CYF-V2R, VISERA Cysto-Nephro Videoscope
- Z-0689-2024 · Dec 4, 2023 · Open, ClassifiedModel No. CYF-VH, CYSTO-NEPHRO VIDEOSCOPE
- Z-0331-2024 · Sep 25, 2023 · Open, ClassifiedCystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
Other Cystoscope And Accessories, Flexible/Rigid devices
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHOlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls1 death1,380 reports
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHROlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls377 reports
- VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2Olympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid1 recall418 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall353 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall1 death332 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid2 recalls75 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid114 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall54 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page