CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHR (REF N3828633)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of CYSTO-NEPHRO VIDEOSCOPE (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04953170434730), received between May 28, 2024 and May 5, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Flaked” (1 report).

5 other recalls cover devices cleared under the same FDA submission (K221683); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Flaked1
  • Loss of or Failure to Bond1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2026Loss of or Failure to Bond

    THE CUSTOMER DID NOT REPORT A MALFUNCTION. DURING AN INSPECTION OF THE ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE, THE ADHESIVE ON A-RUBBER BENDING SECTION HAD A CHIP FOUND. THE MOST LIKELY CAUSE OR PROBABLE CAUSE OF THE REPORTABLE MALFUNCTION IS DETERIORATION OR BREAKAGE OF THE ADHESIVE (CHEMICAL STRESS / PHYSICAL STRESS). THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25089772 ↗
  • InjuryMay 28, 2024Flaked; Flaked

    IT WAS REPORTED, DURING THE THERAPEUTIC CYSTOSCOPY, DEBRIS FROM THE CYSTO-NEPHRO VIDEOSCOPE FELL INTO THE PATIENT'S BLADDER. THERE WAS A PROCEDURAL DELAY WHILE THE DOCTOR TRIED TO BASKET THE DEBRIS OUT BUT COULD NOT CATCH THE DEBRIS IN THE NON-OLYMPUS BASKET. THE PATIENT URINATED THE DEBRIS OUT AND IS IN GOOD STANDING. ANOTHER CYSTOSCOPY WAS PERFORMED LAST WEEK. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE OUTCOME WAS NOT AFFECTED AS A RESULT OF THE ISSUE. THE DEVICE WAS INSPECTED BEFORE USE AND APPEARED TO BE NORMAL.

    Patient problems: Foreign Body In Patient

    MAUDE report 19403927 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K221683. One submission often covers many models, so check the product and lot details in each record.

  • Z-0690-2024 · Dec 4, 2023 · Open, Classified
    Model No. CYF-VHR, CYSTO-NEPHRO VIDEOSCOPE
  • Z-0691-2024 · Dec 4, 2023 · Open, Classified
    Model No. CYF-V2, VISERA Cysto-Nephro Videoscope
  • Z-0692-2024 · Dec 4, 2023 · Open, Classified
    Model No. CYF-V2R, VISERA Cysto-Nephro Videoscope
  • Z-0689-2024 · Dec 4, 2023 · Open, Classified
    Model No. CYF-VH, CYSTO-NEPHRO VIDEOSCOPE
  • Z-0331-2024 · Sep 25, 2023 · Open, Classified
    Cystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.

Other Cystoscope And Accessories, Flexible/Rigid devices

All Cystoscope And Accessories, Flexible/Rigid devices →