URETERO-RENO VIDEOSCOPE OLYMPUS URF-V3R (REF N5782122)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of URETERO-RENO VIDEOSCOPE (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04953170435133), received between Dec 21, 2023 and Apr 14, 2026 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K181451); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Failure to Clean Adequately2
- Detachment of Device or Device Component1
- Leak/Splash1
- Material Split, Cut or Torn1
- Mechanical Problem1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 14, 2026Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETERO-RENO VIDEOSCOPE EXHIBITED FOREIGN MATERIAL STUCK INSIDE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24876531 ↗MalfunctionMar 6, 2026Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETERORENO VIDEOSCOPE EXHIBITED CHIPPED ADHESIVE BENDING SECTION COVER GLUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24539110 ↗MalfunctionOct 8, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THE VIDEOSCOPE UNDERWENT A SCOPE INSPECTION AND FAILED A LEAK TEST. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20394834 ↗MalfunctionSep 10, 2024Break; Failure to Clean Adequately; Break; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEOSCOPE EXHIBITED A BROKEN INSERTION TUBE AND A CLEAR FOREIGN SUBSTANCE WAS FOUND IN THE FORCEPS PASSAGE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20181619 ↗MalfunctionJul 12, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS OBSERVED DURING THE DEVICE MAINTENANCE THAT THE BENDING TUBE OF THE URETERO-RENO VIDEOSCOPE WAS DAMAGED DUE TO METAL PARTS STICKING OUT FROM THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19731416 ↗InjuryJan 4, 2024Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, AT THE END OF THE URETEROSCOPY PROCEDURE, PIECES OF THE OUTER LAYER (BENDING SECTION NEAR THE DISTAL TIP) ON THE URETERO-RENO VIDEOSCOPE WERE FOUND TO BE BROKEN OFF IN THE PATIENT'S BLADDER. THE PIECES WERE RETRIEVED WITH AN ALLIGATOR GRASPER. THERE WAS NO PROCEDURAL DELAY AND NO ADDITIONAL IMAGING WAS REQUIRED AS A RESULT OF THE ISSUE. THE DEFECT IN THE SCOPE MATCHED THE RETRIEVED PIECES. THE DEVICE WAS INSPECTED BEFORE USE WITH NO ANOMALIES. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18446608 ↗InjuryJan 4, 2024Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, AT THE END OF THE URETEROSCOPY PROCEDURE, PIECES OF THE OUTER LAYER (BENDING SECTION NEAR THE DISTAL TIP) ON THE URETERO-RENO VIDEOSCOPE WERE FOUND TO BE BROKEN OFF IN THE PATIENT'S BLADDER. THE PIECES WERE RETRIEVED WITH AN ALLIGATOR GRASPER. THERE WAS NO PROCEDURAL DELAY AND NO ADDITIONAL IMAGING WAS REQUIRED AS A RESULT OF THE ISSUE. THE DEFECT IN THE SCOPE MATCHED THE RETRIEVED PIECES. THE DEVICE WAS INSPECTED BEFORE USE WITH NO ANOMALIES. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18446596 ↗MalfunctionDec 28, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE HAD DEFECTS FROM BENDING RUBBER. THE ISSUE WAS DISCOVERED DURING THE INSPECTION OF THE DEVICE. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE IMAGE GUIDE TUBE COAT WAS FOUND PEELING. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18411790 ↗MalfunctionDec 21, 2023Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED TO OLYMPUS, THAT THE URETERO-RENO VIDEOSCOPE CANNOT ANGULATE SUFFICIENTLY. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT. THE DEVICE WAS RETURNED AND EVALUATED; THE DISTAL TIP DOES NOT FLEX ALL THE WAY. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18372664 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K181451. One submission often covers many models, so check the product and lot details in each record.
- Z-0348-2024 · Sep 25, 2023 · Open, ClassifiedUretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.
Other Ureteroscope And Accessories, Flexible/Rigid devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall1,125 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls397 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall305 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls190 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls170 reports
- ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-10-3DOlympus Medical Systems Corp. · Ureteroscope And Accessories, Flexible/Rigid186 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P7ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall121 reports
- n.a. 091279-01Karl Storz GmbH & Co. KG · Ureteroscope and accessories, flexible/rigid118 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page