URETERO-RENO VIDEOSCOPE OLYMPUS URF-V3R (REF N5782122)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of URETERO-RENO VIDEOSCOPE (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04953170435133), received between Dec 21, 2023 and Apr 14, 2026 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K181451); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Failure to Clean Adequately2
  • Detachment of Device or Device Component1
  • Leak/Splash1
  • Material Split, Cut or Torn1
  • Mechanical Problem1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETERO-RENO VIDEOSCOPE EXHIBITED FOREIGN MATERIAL STUCK INSIDE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24876531 ↗
  • MalfunctionMar 6, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE URETERORENO VIDEOSCOPE EXHIBITED CHIPPED ADHESIVE BENDING SECTION COVER GLUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24539110 ↗
  • MalfunctionOct 8, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED THE VIDEOSCOPE UNDERWENT A SCOPE INSPECTION AND FAILED A LEAK TEST. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20394834 ↗
  • MalfunctionSep 10, 2024Break; Failure to Clean Adequately; Break; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEOSCOPE EXHIBITED A BROKEN INSERTION TUBE AND A CLEAR FOREIGN SUBSTANCE WAS FOUND IN THE FORCEPS PASSAGE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20181619 ↗
  • MalfunctionJul 12, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS OBSERVED DURING THE DEVICE MAINTENANCE THAT THE BENDING TUBE OF THE URETERO-RENO VIDEOSCOPE WAS DAMAGED DUE TO METAL PARTS STICKING OUT FROM THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19731416 ↗
  • InjuryJan 4, 2024Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, AT THE END OF THE URETEROSCOPY PROCEDURE, PIECES OF THE OUTER LAYER (BENDING SECTION NEAR THE DISTAL TIP) ON THE URETERO-RENO VIDEOSCOPE WERE FOUND TO BE BROKEN OFF IN THE PATIENT'S BLADDER. THE PIECES WERE RETRIEVED WITH AN ALLIGATOR GRASPER. THERE WAS NO PROCEDURAL DELAY AND NO ADDITIONAL IMAGING WAS REQUIRED AS A RESULT OF THE ISSUE. THE DEFECT IN THE SCOPE MATCHED THE RETRIEVED PIECES. THE DEVICE WAS INSPECTED BEFORE USE WITH NO ANOMALIES. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18446608 ↗
  • InjuryJan 4, 2024Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS, AT THE END OF THE URETEROSCOPY PROCEDURE, PIECES OF THE OUTER LAYER (BENDING SECTION NEAR THE DISTAL TIP) ON THE URETERO-RENO VIDEOSCOPE WERE FOUND TO BE BROKEN OFF IN THE PATIENT'S BLADDER. THE PIECES WERE RETRIEVED WITH AN ALLIGATOR GRASPER. THERE WAS NO PROCEDURAL DELAY AND NO ADDITIONAL IMAGING WAS REQUIRED AS A RESULT OF THE ISSUE. THE DEFECT IN THE SCOPE MATCHED THE RETRIEVED PIECES. THE DEVICE WAS INSPECTED BEFORE USE WITH NO ANOMALIES. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18446596 ↗
  • MalfunctionDec 28, 2023Peeled/Delaminated; Peeled/Delaminated

    THE CUSTOMER REPORTED TO OLYMPUS, THE HAD DEFECTS FROM BENDING RUBBER. THE ISSUE WAS DISCOVERED DURING THE INSPECTION OF THE DEVICE. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE IMAGE GUIDE TUBE COAT WAS FOUND PEELING. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18411790 ↗
  • MalfunctionDec 21, 2023Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO OLYMPUS, THAT THE URETERO-RENO VIDEOSCOPE CANNOT ANGULATE SUFFICIENTLY. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT. THE DEVICE WAS RETURNED AND EVALUATED; THE DISTAL TIP DOES NOT FLEX ALL THE WAY. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18372664 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K181451. One submission often covers many models, so check the product and lot details in each record.

  • Z-0348-2024 · Sep 25, 2023 · Open, Classified
    Uretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.

Other Ureteroscope And Accessories, Flexible/Rigid devices

All Ureteroscope And Accessories, Flexible/Rigid devices →