Skip to content

Gelweave™ 731508Recalls & reported problems

Vascutek Ltd · Gelweave Gelatin Impregnated Woven Vascular Prosthesis STRAIGHT · Class 2 · Implant

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 05037881105079), received between Jun 26, 2017 and Mar 27, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem5
  • Leak/Splash4
  • Use of Device Problem2
  • Device Handling Problem1
  • Insufficient Device Problem Information1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • Blood Loss1
  • Insufficient Information1
  • No Code Available1
  • No Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 27, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    PHYSICIAN USED A 8MM GELWEAVE GRAFT ON HIS BENTALL CASE TODAY. THE GRAFT OOZED ALL OVER IT'S ENTIRE SURFACE AREA, FOR THE ENTIRE CASE.

    Patient problems: Insufficient Information

    MAUDE report 18985073 ↗
  • MalfunctionOct 18, 2018Leak/Splash; Leak/Splash

    THE REPORTED EVENT WAS REPORTED AS FOLLOWS: BLOOD BLED THROUGH DURING IMPLANT. HOSPITAL DOES NOT PRE-SOAK GRAFTS.

    Patient problems: Blood Loss

    MAUDE report 7976917 ↗
  • MalfunctionOct 5, 2017Leak/Splash; Use of Device Problem; Material Integrity Problem; Leak/Splash; Use of Device Problem; Material Integrity Problem

    INITIAL INFORMATION ON EVENT RECEIVED AT VASCUTEK ON 07-SEP-17 SIMPLY STATED: LEAK FROM FABRIC. FURTHER INFORMATION WAS RECEIVED ON 13-SEP-17 STATING BLOOD LEAKAGE FROM THE BODY OF THE GRAFT, THE GELWEAVE GRAFT WAS USED FOR BLOOD TRANSMISSION. AFTER GRAFTS WERE ANASTOMOSED END-TO-END, WHEN BLOOD FLOW WAS PASSED THROUGH, UNUSUAL LARGE AMOUNTS OF BLOOD LEAKAGE WAS IDENTIFIED FROM THE GRAFT BODY AND THE BLOOD ACCUMULATED IN THE OPERATIVE FIELD. ALTHOUGH THE LEAKING DID NOT STOP, THE PROCEDURE WAS CONTINUED AND SUCCESSFULLY COMPLETED, AFTER THE GRAFT WAS TAKEN OUT AND THE VESSEL WAS LIGATED. NO SERIOUS INJURY TO PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6916593 ↗
  • MalfunctionSep 11, 2017Leak/Splash; Use of Device Problem; Material Integrity Problem; Device Handling Problem; Leak/Splash; Use of Device Problem; Material Integrity Problem; Device Handling Problem

    THE EVENT WAS REPORTED TO VASCUTEK ON 14TH AUGUST AS FOLLOWS; AFTER A STRAIGHT CONNECTOR FOR ARTIFICIAL CARDIO PULMONARY BYPASS WAS INSERTED INTO THE GRAFT FOR BLOOD TRANSMISSION DURING ASCENDING AORTA REPLACEMENT SURGERY AND BLOOD FLOW WAS COMMENCED, LEAKAGE WAS IDENTIFIED IN PART OF THE GRAFT BODY. ADDITIONAL INFORMATION RECEIVED ON 16TH AUGUST; VASCUTEK WAS INFORMED THAT INTERVENTION BY USE OF HAEMOSTATIC AGENT (HYDROFIT) WAS REQUIRED TO STOP LEAKAGE AND THAT THE GRAFT HAD NOT BEEN PRE-SOAKED IN SALINE PRIOR TO USE AS RECOMMENDED IN IFU.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6854445 ↗
  • MalfunctionAug 29, 2017Leak/Splash; Material Integrity Problem; Leak/Splash; Material Integrity Problem

    WHEN BLOOD TRANSMISSION STARTED, OOZING WAS NOT OBSERVED, BUT AFTER A WHILE BLOOD SIGNIFICANTLY OOZED FROM THE ENTIRE SURFACE OF THE GRAFT. AMOUNT OF BLOOD LOSS IS UNKNOWN. THE GRAFT WAS NOT EXPLANTED. IT IS UNKNOWN HOW THE LEAKAGE WAS STOPPED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6826639 ↗
  • MalfunctionJul 13, 2017Material Integrity Problem; Material Integrity Problem

    UPDATED INFORMATION REGARDING THE CIRCUMSTANCES OF THE EVENT: THE GELWEAVE GRAFT WAS NOT IMPLANTED BY THE USER, BUT WAS USED FOR AXILLARY CANNULATION.

    Patient problems: No Information

    MAUDE report 6709814 ↗
  • MalfunctionJun 26, 2017Material Integrity Problem; Material Integrity Problem

    CLINICIAN HAD AN ISSUE WITH AN 8MM GELWEAVE GRAFT ON (B)(6) 2017 THAT WAS BEING USED AS PART OF AN ECMO (EXTRACORPOREAL MEMBRANE OXYGENATION) CASE. IT WAS REPORTED TO BE LEAKING THROUGHOUT BEING DESCRIBED AS A "SOAKER HOSE". THE REPORT ALSO ADVISED THAT THE GRAFT IS NOT AVAILABLE FOR EVALUATION.

    Patient problems: No Code Available

    MAUDE report 6666148 ↗

Other PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices

All PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices →