Skip to content

Gelweave™ 732410108/8S4Recalls & reported problems

Vascutek Ltd · Gelweave Gelatin Impregnated Woven Vascular Prosthesis COSELLI THORACOABDOMINAL GRAFT · Class 2 · Implant

All models of Gelweave (47)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05037881107233), received between Sep 25, 2018 and Dec 9, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

1 other recall cover devices cleared under the same FDA submission (K090987, K162794); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash1
  • Material Integrity Problem1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 9, 2024Leak/Splash; Leak/Splash

    (OPERATION TYPE: THORACOABDOMINAL AORTIC REPLACEMENT EVENT) EVENT REPORTED AS PARTIAL OOZING FROM THE MAIN BODY AND BRANCHES OF THE GRAFT AND POSTOPERATIVE INFECTION: WHEN THE CHEST WAS REOPENED AND THORACOABDOMINAL AORTIC REPLACEMENT WAS PERFORMED USING THE GELWEAVE COSELLI GRAFT, PARTIAL OOZING WAS NOTED FROM THE SURFACE OF THE MAIN BODY AND BRANCHES OF THE GRAFT. AS THE OOZING WAS FROM SOME SPOTS RATHER THAN THE ENTIRE GRAFT, THE SURGEON INDICATED THAT THE GELATINE MAY BE UNEVEN. HYDROFIT WAS THEN USED TO STOP OOZING. AFTER THE SURGERY, THE PATIENT BECAME INFECTED AND IS STILL HOSPITALIZED. THIS REPORT IS BEING SUBMITTED AS FOLLOW UP #1 FOR MANUFACTURING REPORT: 9612515-2024-00082 TO PROVIDE EVENT CLOSURE INFORMATION FOR TAG0105.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20877455 ↗
  • MalfunctionSep 25, 2018Material Integrity Problem; Material Integrity Problem

    ON (B)(6) 2018, THE PATIENT UNDERWENT A PROCEDURE TO HAVE A GELWEAVE COSELLI THORACOABDOMINAL GRAFT IMPLANTED. DURING THE PROCEDURE, THE CLINICIAN REPORTED THAT THE RIGHT RENAL BRANCH ROOT WAS LEAKING. PROTAMINE WAS APPLIED TO ARREST THE LEAK, WHICH WAS UNSUCCESSFUL. THE CLINICIAN THEN APPLIED A SUTURE TO THE AFFECTED AREA AND COMPLETED THE PROCEDURE WITHOUT ANY FURTHER INCIDENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7905082 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K090987, K162794. One submission often covers many models, so check the product and lot details in each record.

  • Z-2542-2021 · Jul 8, 2021 · Terminated
    The Instructions for Use (IFU) for: Gelsoft Plus Bifurcate, Gelsoft Plus Ax-Fem, GelsoftPlus ERS Ax-Fem, Gelweave Straight, Gelweave Trifurcate, Gelweave Ant-Flo, and Gelweave Val

Other PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices

All PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices →