Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05037881107639), received between Apr 22, 2020 and Mar 25, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Obstruction of Flow” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Obstruction of Flow2
  • Defective Device1
  • Material Frayed1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Thrombosis1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2025Material Frayed; Defective Device; Material Frayed; Defective Device

    EVENT WAS REPORTED AS; THE SURGEON HAD TO SUTURE THE GRAFT ON THE FRAYING AREA. THERE WAS NO INJURY TO THE PATIENT. THIS REPORT IS BEING SUBMITTED AS FOLLOW UP #1 FOR MANUFACTURING REPORT NUMBER 9612515-2025-00027 TO PROVIDE EVENT CLOSURE INFORMATION FOR TAG0181.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21685808 ↗
  • InjuryJun 29, 2023Obstruction of Flow; Obstruction of Flow

    HIS EVENT WAS REPORTED TO VASCUTEK LTD VIA THE PANTHER STUDY PROGRAMME AS FOLLOWS: ON (B)(6) 2021 A GELWEAVE BIFURCATE GRAFT WAS IMPLANTED TO TREAT AORTO-FEMORAL ANEURYSM. SHORTLY AFTER PROCEDURE POST-OPERATIVE ACUTE ISCHEMIA DUE TO EARLY OCCLUSION OF GELWEAVE DEVICE IDENTIFIED. TO RESOLVE THE EVENT MEDICATION (UNKNOWN TYPE) WAS GIVEN TO THE PATIENT. THROMBUS WAS REMOVED FROM THE DEVICE. 12X38MM BENTLEY STENT WAS IMPLANTED IN ADRENAL AORTA ABOVE GELWEAVE GRAFT TO AVOID STENOSIS OF ANASTOMOTIC SITE. PATIENT RECOVERED WITHOUT SEQUELAE. THE SITE HAS INDICATED THAT THIS EVENT IS POSSIBLY RELATED TO THE PROCEDURE, DEVICE AND PATIENT PRE-EXISTING CONDITIONS.

    Patient problems: Insufficient Information

    MAUDE report 17223645 ↗
  • InjuryApr 22, 2020Obstruction of Flow; Obstruction of Flow

    ORIGINAL INFORMATION RECEIVED: ON THE 26-MAR-20 VASCUTEK LTD. RECEIVED THE FOLLOWING INFORMATION FROM OUR DISTRIBUTOR (TERUMO FRANCE): THE EVENT WAS REPORTED AS FOLLOWS: A PATIENT UNDERWENT AN AORTO BI-FEMORAL BYPASS PROCEDURE AND A GELWEAVE EXTRA LONG BIFURCATE DEVICE WAS USED. THE PATIENT SUFFERED IMMEDIATE THROMBOSIS. INTERVENTION WAS PERFORMED (UNKNOWN TYPE). THE PATIENT WAS LATER TRANSFERRED TO INTENSIVE CARE. ADDITIONAL INFORMATION RECEIVED: FOLLOWING THE PROCEDURE, AND DURING THE NIGHT THE PATIENT PRESENTED WITH AN INCOMPLETE ACUTE ISCHEMIA OF THE RIGHT LOWER LIMB. DOPPLER IMAGING CONFIRMED AN OCCLUSION IN THE RIGHT LEG. INTERVENTION PERFORMED. SURGICAL APPROACH MADE AT THE LEVEL OF THE RIGHT SCARPA. THE FOGARTY BRINGS BACK A SIGNIFICANT QUANTITY OF FRESH THROMBUS. ARTERIOGRAPHY SHOWS THROMBUS WITHIN THE BODY OF THE PROSTHESIS AND IN THE LEFT LEG. THE LAPAROTOMY IS REDONE AS WELL AS THE LEFT SCARPA. AORTIC CLAMPING, UPON OPENING OF THE AORTIC ANASTOMOSIS; THE BODY OF THE PROSTHESIS IS THE SEAT OF AN ADHERENT THROMBUS WHICH IS REMOVED WITH FORCEPS. FOGARTY COMPLEMENTARY IN THE 2 LEGS AND THE DEEP FEMORALS. CLOSURE OF THE SURGICAL ACCESS WAS COMPLETED. FAVORABLE EVOLUTION THEN

    Patient problems: Thrombosis

    MAUDE report 9985167 ↗