Gelsoft Plus™ 631809PRecalls & reported problems
Vascutek Ltd · Gelsoft Plus Gelatin Impregnated Knitted Vascular Prosthesis BIFURCATE · Class 2 · Implant
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05037881115535), received between May 18, 2018 and Nov 14, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Material Puncture/Hole” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K162803, K955230); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Puncture/Hole2
- Hole In Material1
- Leak/Splash1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Blood Loss1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 14, 2024Material Puncture/Hole; Material Puncture/Hole
DURING SURGERY ON (B)(6) 2024, THE SURGEON UNCLAMPED THE RIGHT ILIAC AFTER COMPLETING ANASTAMOSIS AND STATED THAT THE GRAFT HAD 3 HOLES IN IT. THE SURGEON REPAIRED ALL 3 HOLES AND USED THE PRODUCT . NO PATIENT INJURY AND NO SURGICAL INTERVENTION AS REPORTED BY THE SURGEON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20681858 ↗MalfunctionSep 24, 2024Material Puncture/Hole; Material Puncture/Hole
EVENT DATE: 28 AUG 2024. IMPLANT DATE: ON (B)(6) 2024. DURING THE SURGERY, THE DOCTOR FINISHED SUTURE AND FOUND A HOLE IN THE PRODUCT. AND THEY FOUND BLOOD FLOWING IN THE PRODUCT. THERE WAS NO INJURY OR DEATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20291296 ↗MalfunctionJul 11, 2019Material Integrity Problem; Material Integrity Problem
AFTER THE GRAFT WAS ANASTOMOSED (PROXIMAL END) TWO (2) HOLES WERE IDENTIFIED (ONE IN THE BODY AND ONE IN THE RIGHT LEG). THE SURGEONS DECIDED TO KEEP THE GRAFT IN PLACE AND SUTURED THE HOLES.
Patient problems: Blood Loss
MAUDE report 8779542 ↗InjuryMay 18, 2018Hole In Material; Leak/Splash; Hole In Material; Leak/Splash
VASCUTEK LTD. WAS NOTIFIED OF AN EVENT THAT OCCURRED IN (B)(6), THE EVENT WAS DESCRIBED AS FOLLOWS: BIFURCATED GRAFT IMPLANTED (B)(6) 2017 FOR AAA. TYPICAL POSTOP SURVEILLANCE WAS NORMAL. BLOOD LOSS IN FIRST CASE DOCUMENTED AT 1500ML. PATIENT PRESENTED AT HOSPITAL ON (B)(6) 2018 WITH A FRANK INTRAPERITONEAL AAA RUPTURE DESPITE HIS PRIOR REPAIR. ON EXPLORATION, THERE WAS A 3-4MM HOLE IN THE LIMB WHICH HAD TO BE REPAIRED. CLEARLY FAR TOO BIG TO BE A MISSED HOLE IN THE GRAFT AT TIME OF IMPLANT (THE BLEEDING WAS PRETTY CATASTROPHIC). SURGEON THAT COMPLETED REPAIR NOTED "A TREMENDOUS AMOUNT OF BLOOD WAS EVACUATED FROM THE PERITONEAL CAVITY WITH LOTS OF CLOTS AND LOTS OF BLOOD, LOTS OF FLUID WAS REMOVED." THE ANAESTHETIST DOCUMENTED THAT BLOOD LOSS WAS CHALLENGING TO DISCERN DUE TO THE NUMBER OF BLOOD CLOTS. HOLE WAS NOT LOCATED WITHIN THE FIRST CENTIMETRE PAST THE CROTCH, SO NOT CONSISTENT WITH A CLAMP INJURY LOCATION. NO CALCIFICATIONS ON THE SURROUNDING AORTIC SAC THAT COULD HAVE ERODED THE GRAFT. GRAFT WAS STRAIGHT SO NOT AT A FLEXION POINT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7524966 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K162803, K955230. One submission often covers many models, so check the product and lot details in each record.
- Z-2542-2021 · Jul 8, 2021 · TerminatedThe Instructions for Use (IFU) for: Gelsoft Plus Bifurcate, Gelsoft Plus Ax-Fem, GelsoftPlus ERS Ax-Fem, Gelweave Straight, Gelweave Trifurcate, Gelweave Ant-Flo, and Gelweave Val
Other PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices
- Cardioroot HEWROOT0030Intervascular Sas · Prosthesis, Vascular Graft, Of 6mm And Greater Diameter1 recall2 deaths13 reports
- FLIXENE SLIDER GDS 25058Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall13 reports
- ADVANTA SLIDER GDS 22114Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall11 reports
- Cardioroot HEWROOT0032Intervascular Sas · Prosthesis, Vascular Graft, Of 6mm And Greater Diameter1 recall2 deaths10 reports
- FLIXENE SLIDER GDS 25129Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall9 reports
- FLIXENE SLIDER GDS 25137Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall8 reports
- ADVANTA SLIDER GDS 22222Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall8 reports
- FLIXENE SLIDER GDS 25138Atrium Medical Corporation · PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER1 recall7 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.