Skip to content

Gelweave Vascular Prostheses 731030ERecalls & reported problems

Vascutek Ltd · Gelweave Straight · Class 2 · Implant

All models of Gelweave Vascular Prostheses (299)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 18, 2025Class IIOpen, Classified

    Gelatin sealed woven polyester prostheses devices, designed for vascular repair, were incorrectly labeled with an additional month of shelf life. Device expiry should have been calculated from gel impregnation (date of manufacture) but it was incorrectly calculated from device packaging.

    Z-1492-2026 · Root cause: Error in labeling · Recall details · FDA record ↗

Other PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices

All PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page