Evidence MultiSTAT EV4115Recalls & reported problems

Randox Laboratories Limited · The Evidence MultiSTAT is a benchtop, fully automated, immunoassay analyser that has been designed and developed to be operated in a wide range of laboratory settings within different industries. These include clinical, hospital, academic, research and development, veterinary and food laboratories. · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Apr 5, 2024Open, Classified

    Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.

    Z-1786-2024 · Root cause: Device Design · FDA record ↗