Pad-Pak Pad-Pak-07Recalls & reported problems
Heartsine Technologies Ltd · Battery and Electrode Cartridge · Class 3
All models of Pad-Pak (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 05060167124488), received between Apr 17, 2024 and Jan 14, 2026 — classified as 16 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (14 reports).
7 other recalls cover devices cleared under the same FDA submission (P160008); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation14
- Failure to Power Up2
Patient problems
- No Clinical Signs, Symptoms or Conditions16
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 14, 2026Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24065425 ↗MalfunctionJan 14, 2026Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24065422 ↗MalfunctionJan 13, 2026Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24064416 ↗MalfunctionJan 7, 2026Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23990028 ↗MalfunctionDec 31, 2025Material Deformation; Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23931901 ↗MalfunctionDec 15, 2025Material Deformation; Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23800284 ↗MalfunctionDec 12, 2025Material Deformation; Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23787337 ↗MalfunctionDec 11, 2025Material Deformation; Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23778650 ↗MalfunctionDec 11, 2025Material Deformation; Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23778646 ↗MalfunctionDec 9, 2025Material Deformation; Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23761404 ↗MalfunctionDec 3, 2025Material Deformation; Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23704956 ↗MalfunctionNov 26, 2025Material Deformation; Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23657673 ↗MalfunctionNov 11, 2025Material Deformation; Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23524280 ↗MalfunctionNov 11, 2025Material Deformation; Material Deformation
IN RESPONSE TO A PRODUCT FIELD ACTION, THE CUSTOMER CONTACTED HEARTSINE TO REPORT THE PAD-PAK'S LOCATOR PIN IS BENT. THIS ISSUE CAN LEAD TO A FAILURE TO DETECT WHEN A PATIENT IS CONNECTED, OR COULD CAUSE THE DEVICE TO LOSE POWER DURING OPERATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23524186 ↗MalfunctionMay 7, 2024Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE COULD NOT BE POWERED ON WITH A SPECIFIC PAD-PAK. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19261198 ↗MalfunctionApr 17, 2024Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED HEARTSINE TO REPORT THAT THEIR DEVICE COULD NOT BE POWERED ON WITH A PAD-PAK. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19127646 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160008. One submission often covers many models, so check the product and lot details in each record.
- Z-3270-2026 · Jul 27, 2026 · Open, ClassifiedHeartSine samaritan PAD (Public Access Defibrillator); HeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI codes: 0506016712067 350P US Device, US English (Americ
- Z-3120-2026 · Jul 27, 2026 · Open, ClassifiedHeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI codes: 0506016712067 350P US Device, US English (American Airlines) - 2 Min CPR UDI-DI code: 0506016712207 350P U
- Z-0288-2026 · Sep 18, 2025 · Open, ClassifiedHeartSine Pad-Pak, REF: PAD-PAK-01, PAD-PAK-02, PAD-PAK-03, PAD-PAK-04, PAD-PAK-07, PAD-PAK-03J, PAD-PAK-04J
- Z-2174-2025 · Jun 30, 2025 · Open, ClassifiedHeartSine SAM 350P, HeartSine SAM 360P, and HeartSine SAM 450P The HeartSine samaritan PAD (Public Access Defibrillator)
- Z-2143-2023 · Apr 26, 2023 · Open, ClassifiedHeartSine, REF: PAD-PAK-01, PAD-PAK-03 AND PAD-PAK-03j, CE0123 sold separately and with the HeartSine Samaritan Public Access Defibrillator models: 350P/360P/450P/500P
- Z-1217-2023 · Feb 1, 2023 · Open, ClassifiedHeartsine Samaritan PAD: SAM 350P semi-automatic defibrillator, SAM 360P fully automatic defibrillator, SAM 500P semi-automatic defibrillator with CPR Advisor
- Z-0745-2022 · Jan 13, 2022 · TerminatedHeartSine Samaritan PAD, SAM 350P Semi-Automatic Defibrillator, used with Pad-Pak
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page