Quantum QVMRecalls & reported problems
Spectrum Medical Limited · Quantum Ventilation Module · Class 2
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 05060434420688), received between Aug 26, 2022 and Oct 9, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Gas Output Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Gas Output Problem5
- Suction Problem5
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Hypoxia1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 9, 2025Gas Output Problem; Gas Output Problem
USER CONTACTED SPECTRUM BECAUSE OF A SIGNIFICANT VARIATION IN FIO2 AND SWEEP FLOW OBSERVED DURING THE CASE, AND A GAS OUTLET ALARM WAS NOTED. DEVICE WAS PUT ON THE SIDE FOR REVIEW. NO PATIENT HARM REPORTED AND NO IMPACT TO PATIENT SAFETY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23252656 ↗MalfunctionAug 6, 2025Suction Problem; Suction Problem
USER REPORTED THAT WHILE USING THE VENTILATION MODULE DURING A CASE, THE VAVD WAS INACTIVE. THERE WAS NO VACUUM NOTED AT THE OUTLET. THERE WAS NO PATIENT HARM AS A RESULT OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22727831 ↗MalfunctionJul 22, 2025Gas Output Problem; Gas Output Problem
USER REPORTED THAT THE QVM WAS NOT SHOWING PCO2 MEASUREMENT. THE USER TRIED SWITCHING OUT THE CABLES BUT THIS DID NOT RESOLVE THE ISSUE. THIS HAPPENED DURING SETUP. THE SUSPECTED FAULT WAS IMMEADIATLY OBVIOUS PRIOR TO USE, AND THERE WAS NO PATIENT HARM OR DELAY TO OPERATION AS A RESULT OF THIS ISSUE. THIS ISSUE WAS REPORTED TO ANSM BY THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22563350 ↗MalfunctionJul 22, 2025Gas Output Problem; Gas Output Problem
USER REPORTED THAT DURING THE CASE SWEEP DROPPED TO 2.9 L AND FIO2 DROPPED TO 21%. USER SWITCHED OUT THE QVM AS THEY WERE UNABLE TO CONTROL THE GAS WITH THE QVM. FOLLOWING THE CASE THE QVM WAS SUCCESSFULLY CALIBRATED WITH OUT IT RESOLVING THE ISSUE. WATER TRAPS WERE CHANGED AS MOISTURE WAS FOUND TOWARDS THE QVM SIDE. THERE WERE NO REPORTS OF LONG-TERM HEALTH CONSEQUENCES TO THE PATIENT.
Patient problems: Hypoxia
MAUDE report 22563328 ↗MalfunctionApr 4, 2025Suction Problem; Suction Problem
USER REPORTED THAT THEY WERE UNABLE USE VACUUM-ASSISTED VENOUS DRAINAGE (VAVD) DESPITE HAVING VACUUM INPUT. USER REPORTED THAT THEY COULD NOT FEEL VACUUM FROM VAVD PORT. MALFUNCTION OCCURRED DURING A CASE. THERE WAS NO PATIENT HARM. NO INTERRUPTION OF GAS FLOW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21772775 ↗MalfunctionJan 31, 2025Suction Problem; Suction Problem
THIS VMSR REPORT SUMMARIZES MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT MODEL QVM (PRODUCT CODE: 51-000045-00) EXPERIENCED AN ISSUE REGULATING VACCUM (EITHER VACUUM WAS NOT AVAILABLE, FLUCTUATED OR WAS INCORRECT).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21282697 ↗MalfunctionDec 9, 2024Gas Output Problem; Gas Output Problem
USER NOTICED FLUCTUATION IN MEASURED VALUES OF FIO2 AND PAO2 THAT DID NOT ALIGN WITH THE SETPOINT OF THESE VALUES. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20878121 ↗MalfunctionDec 4, 2024Suction Problem; Suction Problem
PERFUSIONIST REPORTED THAT THEY HAD DIFFICULTY RESETTING THE VACUUM TO NORMAL AFTER REAPPLYING THE CLAMP TO THE TUBING. THE DEVICE WAS CONTINUED TO BE USED TO COMPLETE THE CASE SUCCESSFULLY. WE CONSIDER INABILITY TO REGULATE VACUUM TO BE REPORTABLE, DESPITE NO HARM TO THE PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20840881 ↗MalfunctionNov 7, 2024Gas Output Problem; Gas Output Problem
PERFUSIONIST UNABLE TO CONTROL GAS FLOW FOLLOWING AN ALARM ON QVM2. ANOTHER MEANS TO MANAGE THE GAS WAS USED. WE CONSIDER INABILITY TO CONTROL GAS FLOW TO BE REPORTABLE; HOWEVER, THE PATIENT WAS INVOLVED WITHOUT ANY CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20628554 ↗MalfunctionAug 26, 2022Suction Problem; Suction Problem
IT WAS REPORTED THE VENTILATION MODULE WAS HAVING PROBLEMS WITH THE APPLIED VACUUM ISSUES THROUGH OUT THE WEEK DURING CASES. THEIR WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15299759 ↗
Other Gas Control Unit, Cardiopulmonary Bypass devices
- Electronic Gas Blender 10 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass98 reports
- Quantum QVM2-D-HMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass36 reports
- Quantum QVM4-D-HMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass18 reports
- Quantum QVM2-N-HMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass11 reports
- Quantum QLS1-200Spectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass8 reports
- Electronic Gas Blender 5 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass5 reports
- Electronic Gas Blender 2 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass3 reports
- Quantum QLS1-120Spectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.