Quantum QVMRecalls & reported problems
Spectrum Medical Limited · Quantum Ventilation Module · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 05060434420688), received between Aug 26, 2022 and Oct 9, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Gas Output Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Gas Output Problem5
- Suction Problem5
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Hypoxia1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 9, 2025Gas Output Problem; Gas Output Problem
USER CONTACTED SPECTRUM BECAUSE OF A SIGNIFICANT VARIATION IN FIO2 AND SWEEP FLOW OBSERVED DURING THE CASE, AND A GAS OUTLET ALARM WAS NOTED. DEVICE WAS PUT ON THE SIDE FOR REVIEW. NO PATIENT HARM REPORTED AND NO IMPACT TO PATIENT SAFETY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23252656 ↗MalfunctionAug 6, 2025Suction Problem; Suction Problem
USER REPORTED THAT WHILE USING THE VENTILATION MODULE DURING A CASE, THE VAVD WAS INACTIVE. THERE WAS NO VACUUM NOTED AT THE OUTLET. THERE WAS NO PATIENT HARM AS A RESULT OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22727831 ↗MalfunctionJul 22, 2025Gas Output Problem; Gas Output Problem
USER REPORTED THAT THE QVM WAS NOT SHOWING PCO2 MEASUREMENT. THE USER TRIED SWITCHING OUT THE CABLES BUT THIS DID NOT RESOLVE THE ISSUE. THIS HAPPENED DURING SETUP. THE SUSPECTED FAULT WAS IMMEADIATLY OBVIOUS PRIOR TO USE, AND THERE WAS NO PATIENT HARM OR DELAY TO OPERATION AS A RESULT OF THIS ISSUE. THIS ISSUE WAS REPORTED TO ANSM BY THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22563350 ↗MalfunctionJul 22, 2025Gas Output Problem; Gas Output Problem
USER REPORTED THAT DURING THE CASE SWEEP DROPPED TO 2.9 L AND FIO2 DROPPED TO 21%. USER SWITCHED OUT THE QVM AS THEY WERE UNABLE TO CONTROL THE GAS WITH THE QVM. FOLLOWING THE CASE THE QVM WAS SUCCESSFULLY CALIBRATED WITH OUT IT RESOLVING THE ISSUE. WATER TRAPS WERE CHANGED AS MOISTURE WAS FOUND TOWARDS THE QVM SIDE. THERE WERE NO REPORTS OF LONG-TERM HEALTH CONSEQUENCES TO THE PATIENT.
Patient problems: Hypoxia
MAUDE report 22563328 ↗MalfunctionApr 4, 2025Suction Problem; Suction Problem
USER REPORTED THAT THEY WERE UNABLE USE VACUUM-ASSISTED VENOUS DRAINAGE (VAVD) DESPITE HAVING VACUUM INPUT. USER REPORTED THAT THEY COULD NOT FEEL VACUUM FROM VAVD PORT. MALFUNCTION OCCURRED DURING A CASE. THERE WAS NO PATIENT HARM. NO INTERRUPTION OF GAS FLOW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21772775 ↗MalfunctionJan 31, 2025Suction Problem; Suction Problem
THIS VMSR REPORT SUMMARIZES MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT MODEL QVM (PRODUCT CODE: 51-000045-00) EXPERIENCED AN ISSUE REGULATING VACCUM (EITHER VACUUM WAS NOT AVAILABLE, FLUCTUATED OR WAS INCORRECT).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21282697 ↗MalfunctionDec 9, 2024Gas Output Problem; Gas Output Problem
USER NOTICED FLUCTUATION IN MEASURED VALUES OF FIO2 AND PAO2 THAT DID NOT ALIGN WITH THE SETPOINT OF THESE VALUES. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20878121 ↗MalfunctionDec 4, 2024Suction Problem; Suction Problem
PERFUSIONIST REPORTED THAT THEY HAD DIFFICULTY RESETTING THE VACUUM TO NORMAL AFTER REAPPLYING THE CLAMP TO THE TUBING. THE DEVICE WAS CONTINUED TO BE USED TO COMPLETE THE CASE SUCCESSFULLY. WE CONSIDER INABILITY TO REGULATE VACUUM TO BE REPORTABLE, DESPITE NO HARM TO THE PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20840881 ↗MalfunctionNov 7, 2024Gas Output Problem; Gas Output Problem
PERFUSIONIST UNABLE TO CONTROL GAS FLOW FOLLOWING AN ALARM ON QVM2. ANOTHER MEANS TO MANAGE THE GAS WAS USED. WE CONSIDER INABILITY TO CONTROL GAS FLOW TO BE REPORTABLE; HOWEVER, THE PATIENT WAS INVOLVED WITHOUT ANY CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20628554 ↗MalfunctionAug 26, 2022Suction Problem; Suction Problem
IT WAS REPORTED THE VENTILATION MODULE WAS HAVING PROBLEMS WITH THE APPLIED VACUUM ISSUES THROUGH OUT THE WEEK DURING CASES. THEIR WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15299759 ↗
FDA clearance and classification
Cleared through 510(k) K181942 on Oct 18, 2018 (decision: Substantially Equivalent). Submitted by Spectrum Medical , Ltd. as “Quantum Ventilation Module”. The FDA review took 90 days. FDA record ↗
The FDA regulates product code DTX (gas control unit, cardiopulmonary bypass) under 21 CFR 870.4300 as a Class II device. 21 CFR 870.4300 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Quantum QVM
Has the Quantum QVM been recalled?
No FDA recall record in our data names this exact device identifier (DI 05060434420688), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Quantum QVM?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Aug 26, 2022 and Oct 9, 2025. In those reports, the most common specific device problems are “Gas Output Problem” (5) and “Suction Problem” (5); the most common patient problem is “Hypoxia” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Quantum QVM?
None of the reports are classified as deaths, 0 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Quantum QVM FDA approved?
It was cleared through 510(k) K181942 on Oct 18, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Quantum QVM, and is it still sold?
The FDA's device database lists Spectrum Medical Limited as the company responsible for it (brand Quantum, model QVM). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Gas Control Unit, Cardiopulmonary Bypass devices
- Electronic Gas Blender 10 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass95 reports
- Quantum QVM2-D-HMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass36 reports
- Quantum QVM4-D-HMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass18 reports
- Quantum QVM2-N-HMSpectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass12 reports
- Quantum QLS1-200Spectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass8 reports
- Electronic Gas Blender 5 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass5 reports
- Electronic Gas Blender 2 l/minLivaNova Deutschland GmbH · Gas control unit, cardiopulmonary bypass3 reports
- Quantum QLS1-120Spectrum Medical Limited · Gas Control Unit, Cardiopulmonary Bypass2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page